Manual Tecnico HL

March 30, 2018 | Author: EdisonGualpa | Category: Electrical Connector, Radio Frequency, Radio, Transmitter, Electromagnetic Compatibility


Comments



Description

HemoLight PlusTechnical Manual QL 38/08 Table of Contents 1 IMPORTANT INFORMATION ........................................................................................................4 1.1 1.2 1.2.1 1.3 1.4 2 GENERAL EQUIPMENT CHARACTERISTICS ............................................................................6 2.1 2.2 2.3 2.4 2.5 2.6 2.6.1 2.6.2 2.6.3 3 Electrical Characteristics ........................................................................................................6 Primary Battery .......................................................................................................................6 Dimensions .............................................................................................................................6 Classification...........................................................................................................................6 Environmental Operating Conditions......................................................................................6 Electromagnetic Compatibility ................................................................................................7 Manufacturer’s declaration – electromagnetic emissions.......................................................7 Manufacturer’s declaration – electromagnetic immunity ........................................................7 Recommended separation distances between portable and mobile RF communications equipment and the HemoLight Plus. ..................................................................................9 INITIAL STARTUP........................................................................................................................10 3.1 3.2 3.3 3.4 3.5 3.5.1 3.5.2 3.6 3.6.1 3.6.2 3.7 4 About the Technical Manual ...................................................................................................4 Use of the Technical Manual ..................................................................................................4 Complementary Documentation .............................................................................................4 Symbols used in this manual and on the equipment..............................................................5 Addresses ...............................................................................................................................5 The Equipment’s Packaging .................................................................................................10 Attach the Collection Tray.....................................................................................................10 Powering the Equipment On.................................................................................................11 Initial Tests............................................................................................................................12 HemoData NET Start Up ......................................................................................................13 System Requirements ..........................................................................................................13 Installing the HemoData NET ...............................................................................................13 Installing the HemoLight Plus data network .........................................................................14 Installation Requirements .....................................................................................................14 Operational Procedures for the Installation ..........................................................................14 Enabling the network on the HemoData NET ......................................................................16 REPAIRS AND ADJUSTMENTS .................................................................................................18 4.1 4.2 4.3 4.3.1 4.3.2 4.4 4.4.1 4.4.2 4.4.3 4.4.4 4.4.5 4.4.6 4.5 4.5.1 4.6 4.6.1 4.6.2 4.7 4.8 4.8.1 4.9 4.9.1 4.9.2 Fresenius Kabi Opening the Equipment ........................................................................................................18 Equipment’s Main Board.......................................................................................................20 Clamping System..................................................................................................................21 Removing and Installing the Clamp......................................................................................21 Operational Test of the Clamp..............................................................................................21 Weighting System.................................................................................................................22 Removing the Homogenization Unit .....................................................................................22 Installing and Adjusting the Homogenization Unit ................................................................23 Checking the Adjustments on the Homogenization Unit ......................................................23 Removing the Load Cell .......................................................................................................24 Installing the Load Cell .........................................................................................................24 Calibration of the Weighing System .....................................................................................27 Power Input Socket and Transformer...................................................................................28 Repairing the Power Input Socket and Transformer ............................................................28 Auto Sense Scanner.............................................................................................................31 Programming the Scanner....................................................................................................31 Testing the Scanner..............................................................................................................32 Simulating a Donation Process ............................................................................................33 The Memory Batteries on the Equipment .............................................................................34 How to change the memory batteries:..................................................................................35 The Equipments Keyboard ...................................................................................................37 Keyboard test........................................................................................................................37 How to replace the keypad ...................................................................................................37 HemoLight Plus QL-38/08 04/2011 Page 2 of 77 5 SOFTWARE UPGRADE ON THE HEMOLIGHT PLUS ..............................................................39 5.1 5.1.1 5.2 6 Necessary Materials: ............................................................................................................39 How to obtain the TIGER DOWNLOADER: .........................................................................39 Transferring a new software version to the HemoLight Plus................................................39 ELECTRICAL DIAGRAMS AND CIRCUIT DESCRIPTION OF THE HEMOLIGHT PLUS.........42 6.1 Block Diagram ......................................................................................................................42 6.2 Connector Positions..............................................................................................................43 6.3 PCB 01401V06 .....................................................................................................................44 6.3.1 Layout of the Printed Circuit Board PCB 01401V05.............................................................44 6.3.2 DC Power Supplies...............................................................................................................45 6.3.3 Primary Battery Circuit..........................................................................................................45 6.3.4 Standby Circuit .....................................................................................................................47 6.3.5 Micro-controlled CPU............................................................................................................48 6.3.5.1 Controlling the Keypad and Display Matrix ....................................................................48 6.3.5.2 Driving the Peripherals ...................................................................................................48 6.3.5.3 RS232 Serial Interface ...................................................................................................49 6.3.5.4 I2C Serial Interface ........................................................................................................49 6.3.5.5 Analogue Inputs .............................................................................................................50 6.3.6 The Scale Circuit ..................................................................................................................50 6.3.7 The Oscillating Tray Circuit...................................................................................................51 6.3.8 The Clamp Circuit .................................................................................................................51 6.3.9 Electronic Schematics PCB 01401V06 ................................................................................52 6.3.9.1 Electronic Schematics for the Power Supplies, Primary Battery Charger, Standby and Memory Batteries ..............................................................................52 6.3.9.2 Electronic CPU Circuit Schematics ................................................................................53 6.3.9.3 Electronic Circuit Schematics for the Tray, Clamp and Scale Circuit ............................54 6.4 PCB 01403V02 .....................................................................................................................55 6.4.1 Operator Interface.................................................................................................................55 6.4.1.1 Keyboard ........................................................................................................................55 6.4.1.2 Display............................................................................................................................55 6.4.2 RS232 Serial Communications Interface..............................................................................55 6.4.3 Printed Circuit Board Layout PCB 01403V02.......................................................................56 6.4.4 Electronic Schematics PCB 01403V02 and PCB 01410V01. ..............................................56 6.5 PCB 01410V01 .....................................................................................................................57 6.6 Indicators Interface ...............................................................................................................57 6.6.1 Printed Circuit Board Layout PCB 01410V01.......................................................................57 7 THE MEMORY CARD READER ..................................................................................................58 8 ELECTRICAL DIAGRAMS AND CIRCUIT DESCRIPTION OF THE MEMORY CARD READER .......................................................................................................................................59 8.1 8.1.1 8.1.2 8.1.3 PCB 01901V1 .......................................................................................................................59 Memory Card Reader Circuit ................................................................................................59 Printed Circuit Board Layout PCB 01901V01.......................................................................60 Electronic Schematics PCB 01901V01 ................................................................................61 9 HEMOLIGHT TRANSPORT CASE ..............................................................................................62 10 SAFETY END TESTS...................................................................................................................64 10.1 10.2 10.3 10.4 Leakage Currents Through the Protection Connector..........................................................64 Leakage currents through the cabinet (surface / ground) ....................................................65 Leakage currents through the cabinet (surface / surface)....................................................66 Ground Bound Test ..............................................................................................................67 11 MAINTENANCE CHECKLIST ......................................................................................................68 12 REPLACEMENT PARTS, TOOLS AND ACCESSORIES LISTS................................................70 12.1 Fresenius Kabi List of Replacement Parts only for serial numbers up to 01805108 E-code 000. ................73 HemoLight Plus QL-38/08 04/2011 Page 3 of 77 do not touch any parts that have power applied to it also protect the internal components as to contact with liquids.1 About the Technical Manual This manual technically describes the HemoLight Plus equipment.Chapter 1: Important Information 1 Important Information 1. then please observe the use of protection and measures against electrostatic discharges (ESD). The equipment’s usage instructions are to be found in the Users Manual. Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 4 of 77 . When doing maintenance and the equipment is open. indicating that it is in its 8 revision level. Some of the activities described in this technical manual require the use of technical measuring equipment and special tools. dated April 2011. If the need arises for the repair or replacement of electronic components. 1. This content may be changed without prior warning and without any legal implications to Fresenius Kabi. Always use the bracelets and/or other anti-static mechanisms.2. Its technical manual is complemented by the “Operating Instructions HemoData NET” as well as the “Operating Instructions HemoLight Plus”. modifications or repairs to the HemoLight Plus must only be done by trained personnel and that are recommended by the manufacturer.1 Complementary Documentation The HemoLight Plus is a part of a blood donation system manufactured by Fresenius Kabi. 1. Adjustments. Fresenius Kabi and its distributors will not be held responsible for any unforeseen circumstances if these recommendations are not complied with. and receives the following code QL-38/08 th 04/2011. The numbering and corresponding chapters are respectively marked on the headers and footers of the pages.2 Use of the Technical Manual This manual was developed to be used by service technicians who have been trained by the manufacturer and have experience in mechanical. electrical and electronic maintenance. 107/0001-84 European Representative: FRESENIUS KABI AG D-61346 Bad Homburg LOCAL SERVICE: Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 5 of 77 . Rua Roque Gonzáles. 1.3 Symbols used in this manual and on the equipment When printed in the manual this means important information.Brasil CNPJ 49. This label is attached internally on the equipment next to the main grounding protection. This equipment may be supplied with a sealed Lead-Acid battery. B Type equipment. 128.601. This symbol indicates the temperature limits for storage and transportation. This symbol indicates that the equipment is in conformity with the directive 93/92 of European Community. When attached to the equipment. without parts that are directly applicable to the patient. This symbol indicates areas on the equipment that may cause operator injury due to crushing. IPX0 Equipment unprotected as to the penetration of water.Chapter 1: Important Information 1. this means that the accompanying documents must be consulted. This symbol indicates that the operating instructions should be consulted. which must be sent in for recycling after its working life is over.4 Addresses Manufactured by: FRESENIUS HEMOCARE BRASIL LTDA. 06855-690 – Jardim Branca Flor Itapecerica da Serra – SP . 1 Electrical Characteristics Input Voltage Input Current 50/60Hz Input Fuses 110-127V~ 220-230V~ 140mA 70mA 2x200mAT Secondary Fuse (internal) 2.2 1A F Primary Battery Brand Panasonic¹ UNICOBA² Model PV-BP88 --------- Type Lead-Acid Lead-Acid Voltage 12V 12V Current Capacity 2.3Ah ¹ Panasonic is the registered trademark of the Matsushita Electric Co.5 Class I Equipment / Equipment Internally Powered TYPE B IPX0 (without protection) Continuous with an intermittent load Environmental Operating Conditions Operating Temperature Relative Humidity Sea level Fresenius Kabi HemoLight Plus QL-38/08 10ºC to 40ºC 30% to 70% Up to 2000m 04/2011 Page 6 of 77 .4 196 mm 270 mm 206 mm 3200g Classification Type of protection against electrical shock Degree of protection against electrical shock Degree of protection against water penetration Mode of Operation 2.Chapter 2: General Equipment Characteristics 2 General Equipment Characteristics 2.3Ah 2.3 Dimensions Maximum Values Height Width Depth Weight (without batteries) 2. Ltd. ² Provided exclusively by Fresenius 2. concrete or ceramic tile.Chapter 2: General Equipment Characteristics 2. Page 7 of 77 .6.2 Compliance Electromagnetic environment – guidance The HemoLight Plus uses RF energy only for its internal function.1 Manufacturer’s declaration – electromagnetic emissions The HemoLight Plus is intended for use in the electromagnetic environment specified below. If floors are covered with synthetic material. Therefore. If user requires operation in power lines frequently susceptible to surges. Mains power quality should be that of a typical commercial or hospital environment. The costumer or the user of the HemoLight Plus should assure that it is use ins such an environment Emissions test RF emissions CISPR 11 RF emissions CISPR 11 Harmonic emissions IEC 61000-3-2 Harmonic emissions IEC 61000-3-3 2.6 Electromagnetic Compatibility 2.5kV CM QL-38/08 04/2011 Immunity test Floors should be wood.5kV DM ±2 kV lines(s) to earth HemoLight Plus 0. Group 1 Class A The HemoLight Plus is suitable for use in all establishments other than domestic and those directly connected to the public low-voltage power supply network that supplies buildings use for domestic purposes. it is recommend that the HemoLight Plus be powered from a stabilized unit or its own battery. Not applicable Not applicable Manufacturer’s declaration – electromagnetic immunity The HemoLight Plus is intended for use in the electromagnetic environment specified below. The costumer or the user of the HemoLight Plus should assure that it is use ins such an environment Immunity test Immunity test Immunity test Electrostatic discharge (ESD) IEC61000-4-2 ±6 kV contact ±8 kV air ±6 kV contact ±8 kV air ±2 kV for power supply lines ±2 kV for power supply lines ±1 kV for input/output lines Not applicable Electrical fast transient/burst IEC 61000-4-4 Surge IEC61000-4-5 Fresenius Kabi ±1 kV lines(s) to line(s) 0. it RF emissions are very low and are no likely to cause any interference in nearby electronic equipment. the relative humidity should be at least 30% Mains power quality should be that of a typical commercial or hospital environment.6. 17√P (150kHz to 80MHz) d = 1.5 cycle <5% UT (>95% dip in T) for 0. b HemoLight Plus may be affected by interference if installed in the vicinity of equipments marked with the following symbol: Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 8 of 77 . Recommended separation distance (d) d = 1. Portable and mobile RF communications equipment should be used no closer to any part of HemoLight Plus. short interruptions and voltage variations on power supply input lines IEC 61000-4-11 Power frequency (50/60 Hz) magnetic field <5% UT (>95% dip in T) for 0.5 cycle 40% UT (60% dip in UT) for 5 cycles 40% UT (60% dip in UT) for 10 cycles 70% UT (30% dip in UT) for 25 cycles 70% UT (30% dip in UT) for 50 cycles <5% UT (>95% dip in T) for 5s <5% UT (>95% dip in T) for 5s 3 A/m 50Hz 3 A/m for 3m IEC 61000-4-8 Mains power quality should be that of a typical commercial or hospital environment.Chapter 2: General Equipment Characteristics Voltage dips.5GHz) 3 Vrms 150kHz to 80MHz 3V 3V/m 80Mhz to 2. it is recommended to keep HemoLight Plus to be supplied by its own external battery system.34√P (800MHz to 2. Power frequency magnetic fields should be at levels characteristic of a typical location commercial or hospital environment. and d is the recommended separation distance in meters (m).17√P (80MHz to 800MHz) Conducted RF IEC61000-4-6 Radiated RF IEC61000-4-3 d = 2. Field strengths from fixed RF transmitters. including cables. as determined by an electromagnetic site a survey . than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.5GHz 3V/m Where P is the maximum output power in watts (W) according to the transmitter manufacturer. If the user requires continued operation during power interruptions. should be less than the compliance level in each frequency range. 17 2.3 Recommended separation distances between portable and mobile RF communications equipment and the HemoLight Plus.17 1. an electromagnetic site survey should be considered.6.7 23.c.12 0. Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 9 of 77 . according to the maximum output power of the communications equipment.17√P d = 1. To asses the electromagnetic environment due to fixed RF transmitters. The HemoLight Plus is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The costumer or the user of the HemoLight Plus can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the HemoLight Plus as recommended below. the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter. NOTE 2 At 80MHz and 800MHz.70 7. the higher frequency range applies. such as re-orienting or relocating the HemoLight Plus b- Over the frequency range 150kHz to 80Mhz.5GHz (W) d = 1. AM and FM radio broadcast cannot be predicted theoretically with accuracy.34 10 3. NOTE 3 These guidelines may not apply in all situations. where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.65 1 1.34√P 0. If abnormal performance is observed. amateur radio.01 0. field strength should be less than 3V/m 2. Separation distance according to frequency of transmitter Rated maximum output (m) power of transmitter 150kHz to 80Mhz 80Mhz to 800Mhz 800Mhz to 2.1 0.4 For transmitters rated at a maximum output power not listed above.70 3. If the measured field strength in the location in witch the HemoLight Plus is used exceeds the applicable RF compliance level above. objects and people. mains voltage prior to application of the test level.40 100 11.24 0. objects and people.17√P d = 2. Electromagnetic propagation is affected by absorption and reflection from structures.7 11. Electromagnetic propagation is affected by absorption and reflection from structures.12 0. additional measures may be necessary. the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situations.Chapter 2: General Equipment Characteristics NOTE 1 UT is the a. such as base stations for radio (cellular/cordless) telephones and land mobile radios. a- Field strengths from fixed transmitters.37 0.37 0. NOTE 1 At 80MHz and 800MHz. the HemoLight Plus should be observed to verify normal operation. Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 10 of 77 . Keep this box for possible future transport. To guarantee that the weighing system operates correctly the HemoLight Plus must be setup on a horizontal surface on which all four of its feet rest. and it is for this reason that it must be handled with the appropriate care. Do not exert excessive force on the tray when attaching it. 3. The equipment’s weighing system is equipped with sensitive components. Do not try to move the oscillating unit by exerting force on the system. On removing the equipment from the package.Chapter 3: Initial Startup 3 Initial Startup The HemoLight Plus must be setup on an appropriate surface. This surface must be firm and free from vibrations. The guide pins on the bottom of the tray allow it to be attached in only one position. Do not expose the equipment to sunlight. packed in a protective box. Drops and knocks must be avoided.1 The Equipment’s Packaging The HemoLight Plus is delivered. check it to see if it was damaged during transport.2 Attach the Collection Tray The collection tray on the HemoLight Plus is detachable. 3. It must be attached to the oscillating base. press any key (with the exception of “RESET”) during the initial animation. Insert the three pole Input Power cable into the machine and then connect it to a mains power outlet socket.3 Powering the Equipment On At the rear of the equipment there is a power Input Socket. In this case. the equipment will then be supplied by the battery.Chapter 3: Initial Startup 3. indicating that the battery is in use. The green LED will light up. The yellow indicator LED (on) on the panel will light up. Check in item 4. rock the O/I switch. The HemoLight Plus will automatically enter into the “Stand By” mode after approximately 40 seconds of idle* time. The ON/OFF switch will not power off the equipment when it has a charged battery connected to it. Before powering on the HemoLight Plus. To power the equipment on again press the “RESET” key or reconnect it to the mains power supply. One may opt to use the primary battery to run the equipment. *The HemoLight Plus is considered idle only when the display is alternating between date and time. make sure that the voltage selector switch at the back of the equipment corresponds to the local mains voltage. an audible alarm will sound and the following initial animation will be displayed: FRESENIUS KABI To see the main processor’s software version and its compilation. To power the equipment on.5 on the procedure to select the correct voltage. With the battery being in the HemoLight Plus compartment and it being disconnected from the mains power supply. Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 11 of 77 . a charged battery must be placed into the available compartment at the bottom that is for this purpose and marked with the following label: If the battery is connected to the equipment and it is connected to the mains power with the ON/OFF switch in the “ON” position then the battery will be in a charging state. The charger on the HemoLight Plus takes around 10 hours to guarantee a satisfactory standalone charge. indicating that the equipment is powered on. When the audible alarm ceases and whilst the initial animation is still being displayed. Only recharge the primary battery on the HemoLight Plus or on the charges designed for this purpose. 3. The keys that are pressed will not be shown on the display. The following screen will be seen: Press ENTER twice and the equipment will sequentially carry out the following processes: • Activate the clamp system for three full consecutive cycles. The battery is an accessory and is not supplied with the HemoLight Plus.4 Initial Tests Press the “RESET” switch on the equipment. and this one must be consulted whenever necessary. 8 and 9 one-by-one sequentially and in this order. The system configuration procedure is available in the equipment user’s manual. and when red it means that the battery is discharged.Chapter 3: Initial Startup The second LED on the panel will indicate the battery’s current condition that is in the compartment. The following tray will be displayed: Press ENTER and then the system will start to alternate the display between Date and Time. When green the battery is being charged. Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 12 of 77 . press the keys 6. • Activate the alarm systems for approximately 3 seconds. • Move the oscillating tray for approximately 3 seconds. If the installation does not start automatically.5 HemoData NET Start Up 3.2 Installing the HemoData NET Depending on the operational system version that you are using. Operational System: Windows 9x/Me/NT/2000/XP 3. • • • • Insert the CD containing the HemoData NET installation in the PC’S CD-ROM drive. Confirm the installation steps until it is finalized Restart the PC The following directory structure will be generated: To execute the program.5. use the icon that was created in the desktop area or do the following Start/Programs/HemoData NET/HemoData NET Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 13 of 77 .Chapter 3: Initial Startup 3.1 System Requirements Standard IBM-PC with a Pentium 133Mhz processor or better 16 Mb of RAM memory SVGA Color Monitor Serial port with an available DB9 connector CD-ROM 2X Unit Minimum space on the hard drive of approximately 50 MB available to install the program. This space must be increased in accordance with the number of collections.EXE program.5. it may be necessary to consult a network administrator to allow the installation for the new software on the workstation. then access the CD unit via the Windows Explorer and execute the SETUP. 8.6 Installing the HemoLight Plus data network 3. The maximum distance between 2 mixers may not be more than 30 meters.Chapter 3: Initial Startup 3. The following figure shows the wire positions. Make sure that a certain wire color always occupies the position on the connectors.6. the following procedure must be adopted: • Press the RESET key whilst the equipment is powered on and when the buzzer stops and whilst the initial animation is still being displayed. Up to 32 pieces of equipment may be installed on one self same network. The connectors must be of the RJ11 6/4 type. All mixers used in the making up of the network must have their network module enabled. To use the Smart Card Memory. Incase a mixer is not enabled. including power supply and data cables. Y cables (One unit for each HemoLight on the network). 3 and 9.2 Operational Procedures for the Installation The necessary cables must be made from 4x26AWG flat cable (telephone standard). It is important that the parallelisms of the colors are complied with. 3. The correct code must be punched in to confirm the network module’s release (the code is furnished by Fresenius Kabi) All equipment belonging to the network must receive a control ID (network number). use the Memory Card Reader RS232 to RS485 Converter (part number 81053). Please follow the following instructions: Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 14 of 77 . sequentially punch in the following characters 2. A pair of clamping pliers specifically for this operation will be necessary. The HemoData NET network must be installed by technical personnel that are authorized by Fresenius Kabi.1 Installation Requirements • • • • RS485 Converter.6. • The system will request a release code. The total sum length of all cables on the network may not exceed 1000 meters. Network Cables (One unit for each HemoLight on the network). 4 are sequentially pressed (one-by-one) and in this order. • • • The input number will be used by the PC to acknowledge the mixer during communications. The RS485 data converter cable must be connected to an available serial port. Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 15 of 77 . that is. 3. When the mixer requests DATE / Time? Press NO. This number may not be repeated on one self same network. Input a number between 1 and 32 using the numeric keypad on the mixer. All equipment must be connected to the network via the data and “Y” cables. 2. The power supply must be connected to a power mains socket. To access the Configuration Routine on the System it is necessary after pressing the RESET key during the initial animation that the keys 1. an incorrect value may be corrected "n" number of times until the ENTER key is pressed. If a number is repeated by accident then the PC will not acknowledge either one of the pieces of equipment. An USB/Serial adapter may be used in case there are no available serial ports. Then the display will request the network number. The last keys pressed will be considered.Chapter 3: Initial Startup Access the configurations routine. To execute a program use the icon on the desktop or do the following Start/Programs/HemoData NET/HemoData NET. Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 16 of 77 .Chapter 3: Initial Startup 3. Access the Configuration/Groups menu Input a name for the group of mixers and confirm with the OK button. Login into the system by clicking on the MASTER button. Input access code 123 and confirm with the OK button.7 Enabling the network on the HemoData NET Start the HemoData NET. Select a group that ahs been setup beforehand in the selection box “Select Group”. Input the mixer’s serial number into the field and confirm with OK. The selected mixer must be one that ahs been numbered beforehand. Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 17 of 77 .Chapter 3: Initial Startup Access the Configuration/Mixer on Network menu. Repeat the last two aforementioned steps for all mixers on the network. Select a mixer in the available mixers column and click on the Add button. For help in operating the HemoData NET access the Help option on the system or consult the “Operating Instructions HemoData NET”. The number will indicate the mixer’s ID on the network. Remove the lower screws from (7) the housing (6). Remove the housing’s screws on the bottom. Loosen screw (2) and the fastening washer (3) from the oscillating base (5). Pull and remove the oscillating base (5). Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 18 of 77 . It is not necessary to remove the bearing (4).Chapter 4: Repairs and Adjustments 4 Repairs and Adjustments 4.1 Opening the Equipment Pull and remove the detachable tray (1). insert a flat screw driver into the recesses located on the sides of the housing. To do this. Then tilt the upper section of the housing as shown in the figure. Remove the flat cable (1) from its connector on the main board. Lift the upper part of the housing until it is no longer possible to see the tip of the clamp (2).Chapter 4: Repairs and Adjustments Release the clips that hold the upper and lower parts of the housing together. Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 19 of 77 . and carrying out the movement as shown in the picture on the side. confirm. • The following screen will be displayed. The new serial number will be recorded on the main board. and then the system will start to alternate with Date and Time on the display. 1. where “XXXX” is the default identification value. 6. the digits to be used would be: 9587 • Fresenius Kabi HemoLight Plus QL-38/08 Press ENTER. Whilst the initial animation is being displayed. 04/2011 Page 20 of 77 . Proceed in the following manner: • Press the “RESET” switch on the equipment. remembering that the keys pressed will not be shown on the display. sequentially press the following keys one-by-one 0.Chapter 4: Repairs and Adjustments 4. If the need arises for it to be replaced then this identity must be input into the board. This number must be obtained from the technical label on the equipment. being that only the 4 middle digits must be used. 7 and in this order.2 Equipment’s Main Board The equipment’s main board holds the equipment’s digital identification belonging to it. For the serial number shown on the image on the side . • After a few seconds the display will request: • Press ENTER and then the equipment will request the new serial number. Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 21 of 77 . This guarantees the correct positioning of the clamp. During replacement. The clamp should close when the display shows “PUNCTURE AND ENTER”.4 in this manual. 4.3. Closing Effectiveness.1 Removing and Installing the Clamp The clamp is mounted to the bottom of the equipment by means of 3 screws on its base.2 Operational Test of the Clamp Necessary equipment: • 1 x 20 ml disposable syringe. When this happens exert pressure on the syringe until the pressure gauge gives a reading of 1 BAR (kgf/cm²). Power the equipment on. be aware that the chamfered surface on the base of the clamp faces the transformer. Opening and Closing. Start a new donation procedure.Chapter 4: Repairs and Adjustments 4. • 1 pressure gauge with a scale of up to 2 Bar. Carry out the “INITIAL TEST” procedures in accordance with item 3. and at the moment that the syringe is pressed there must be no bubbles inside the container. To check if the clamp is “sealing” the pipe completely insert the end of the pipe into a container with water. Setup a system as shown in the figure on the side. The cables attached to the connector in the illustration are responsible for controlling the clamp’s activation.3 Clamping System 4. Place the clamp on a section of the tube immediately after the pressure gauge.3. 4 Weighting System The weighing system is made up of the homogenization unit and the load cell.4. The unit is then loose and may be removed. 4.Chapter 4: Repairs and Adjustments 4. Carefully cut the cable ties that hold the wires on the bottom of the equipment so as not to damage the insulation on the wires. Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 22 of 77 .1 Removing the Homogenization Unit Completely loosen the two Allen screws that fasten the oscillating system onto the load cell. Attach the homogenization unit to the load cell.4. Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 23 of 77 . 4. by tightening the screws with a force of 6. Keypad with a Display to adjust the HemoLight Plus (Part number 070187003). • Ruler. Access the equipment’s scale module by pressing the SCALE key. being that the maximum deviation may not be more than 3g on all 5 test points. So that the front face of the driving init remains parallel to the front face of the load cell.4. Wait for the scale to zero.Chapter 4: Repairs and Adjustments 4. first in the center and then on the 4 corners. • Torque wrench Make sure by visually checking that no part of the movement unit touches the protection of the load cell.4. this must correspond to the standard weight with a maximum deviation of 2g on all 5 test points. Place a calibrated weight piece of 500g on the tray. Repeat with the 1000g.5 Nm. Mount the oscillating base and collection tray onto the homogenizing unit.6). Make sure that the equipment is calibrated (please see item 4.3 Checking the Adjustments on the Homogenization Unit Necessary equipment: • • A set of standard weights that may be combined to make up loads of 500g and 1000g. If some anomaly is detected then the system must be adjusted again.2 Installing and Adjusting the Homogenization Unit Necessary equipment: • Keypad with a Display to adjust the HemoLight Plus (Part number 070187003). checking the display reading for each condition. Loosen the load cell from its base with a 4 mm Allen key.3mm and 0. Remove the base plate from the bottom of the equipment.1 mm.5 Installing the Load Cell Necessary materials: • • • • Fresenius Kabi HemoLight Plus QL-38/08 Feeler gauge of 0.4.6 mm.4 in this manual.4 Removing the Load Cell It is necessary for the homogenization unit to have already been removed before starting this procedure Remove the screws from the load cell supporting plate with an open ended spanner of 10 mm (or a socket spanner). Remove the load cell connector from the main board.4.Chapter 4: Repairs and Adjustments Still with the equipment open carry out the “INITIAL TEST” procedures according to item 3.45 mm and 1. Torque wrench 04/2011 Page 24 of 77 . Do not hold the unit by its load cell. Locking fluid for the screws. 4. Feeler gauge of 0. 4. To finalize remove the collection tray and oscillating base and then fasten the wires with plastic cable ties to the bottom of the equipment to the positions where there are the appropriate mounts. 45 mm must be able to slip between the cell and the plate. The feeler gauge of 0.45 and 1.3mm feeler gauge. Try to correctly position the load cell. but the 1. to do this use the 0.6 mm feeler gauge must be able to pass.1 mm must not be able to slip between the cell and the plate.Chapter 4: Repairs and Adjustments Mount the load cell as shown on the figure on the side. but the 0. The gap between the load cell sides and the protection sides must be checked before the screws are fully tightened. Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 25 of 77 . The arrow that is stamped on its body indicates the end that must remain free and the displacement direction. Between the load cell sides and the protection it must be possible to pass the 0.1mm feeler gauges. The gap at the bottom of the load cell must be checked. press the REST key and when the alarms stops. For this the offset value must be checked with only the load cell fixed on the base. The wires for the grounding system (green/yellow) coming from the transformer and the power mains input must be duly fastened to the screw as shown in the illustration. Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 26 of 77 . press the following keys: PAUSE. Install the load cell supporting base into the equipment. and check it again with protection kit mounted. CLEAR and STOP. The star washer is positioned between the metallic plate and the wire terminals. then the screws must be tightened with a force of 10 Nm. Do not forget to include the star washers when assembling. CHECK. It is possible to check if any part of the protection is touching the load cell.Chapter 4: Repairs and Adjustments After the gaps have been checked on the sides and bottom of the load cell. Do not hold the unit by the load cell. To check. For this purpose use the torque wrench. The read value can’t different. 6 Calibration of the Weighing System Necessary equipment: • • Set of weight that may be combined to furnish loads of 500g and 1000g. Again place the weight piece onto the tray and check if the adjustment was effective. If the net value on the display’s readout is greater than 2g than the real value. Keypad with Display to adjust the HemoLight Plus in case the option is taken to calibrate the equipment whilst it is open (Part number 070187003). When the value on the display converges to the standard weight value (with a variation of ± 2 g). check the system’s behavior with the weight piece placed at the 4 corners of the tray. checking the readout at the center and at all 4 corners. place a calibrated weight piece of 500g on the center of the tray. For this position on the scale the variation may not exceed 3 g. Repeat this step as many times as is necessary to obtain the ideal response. Wait for the zero process to finish. Power the equipment ON. Press the SCALE key. The calibration routine may be carried out with the equipment opened or closed. then the calibration potentiometer must be adjusted until the correct value is achieved. Replace the standard weight piece with one of 1000 g. When the display shows 0 g. If the option is taken to carry out the process with the equipment open. Remove the weight piece from the tray and press ENTER then wait for the display to show DATE and TIME. Wait for the zero process to finish. then it will be necessary to use the auxiliary display and the equipment must be checked after it is installed in its housing. Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 27 of 77 . Press the SCALE key.4.Chapter 4: Repairs and Adjustments 4. 5.25 VAC (RMS) when supplied with the rated voltage. The transformer has one no load output rated at 14 ± 1. (Please see items 4. remove the fuse holder cover (5) and the fuse holder (4). Turn the fuse holder so that the chosen voltage is shown in the fuse holder window (6) To switch between the voltages and/or replace fuses. remove the fuse holder cover (5) and the fuse holder (4).1 and 4. these must be removed from the equipment’s base. Turn the fuse holder so that the chosen voltage is shown in the fuse holder window (6). In the drawing the ON / OFF switch (1) is shown. To check this condition a Multimeter must be employed to meter between the output wires (white).1 Repairing the Power Input Socket and Transformer In case it becomes necessary to replace the Power Input Socket and/or Transformer.4. three pole cable socket (2) and fuses (3). 4. It is necessary to first remove the homogenization unit of each load cell.5 Power Input Socket and Transformer The standard AC input configuration is that of 230 V.4.Chapter 4: Repairs and Adjustments 4.4). Before powering on the HemoLight Plus make sure that the voltage selector switch at the rear of the equipment corresponds to the local mains power supply. The secondary wires from the transformer must be removed from the main board. To select between the voltages and/or replace fuses. Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 28 of 77 . The spade terminals must be soldered at the ends of the wires and the insulating covers then slipped over for protection. The secondary wires on the transformer (white) must be cut so that they will have 150 mm free. The caps on the insulated spade terminals must be removed and threaded through each wire. Protect the join with heat shrink sleeving. black and gray) must be threaded through the leatherette sleeving. The suppression capacitors must be soldered to the power supply socket according to the following instructions: Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 29 of 77 . Ensuring that the wire join faces the transformer. The solder joint must be made perpendicularly to the terminal and being threaded through its hole so that it is possible to fit the wire the wire in the groove at the bottom of the plastic on the equipment. and as close as possible.Chapter 4: Repairs and Adjustments The primary wires that are left over (red. (use the figure on the side as a reference). Secondary Wires with 160 mm free (considering the connector) Gray Black Red In case of replacement the transformer must be prepared by soldering the thermostat (outer most gray wire) in series with the zero tap primary coil on the transformer (outermost red wire). Leatherette Sleeving The grounding wire must be soldered to the incoming power terminal. which is a common to the housing as well. as follows. Note that the wires are placed in defined positions.Chapter 4: Repairs and Adjustments Through Hole Drill one 3 mm and1 hole of 1. It must also be a through hole going right through the socket. The legs must be fed through both holes to the soldered contacts on the socket. st Attach the 1 capacitor according to the illustration. The 1.3 mm in the power socket.3 mm hole must be made in such a manner that it becomes an extension of the hole in the metallic terminal. nd st Bring the 2 capacitor next to the 1 one using a plastic cable tie as shown here on the side. The legs of the capacitors must receive the “jumpers” and be soldered according to the illustration on the right. All solder joints are made with the wire going through orifice and wrapping around the metallic terminal. The leg on the left hand side must be extended so that the “jumper” itself makes a link between terminals 5 and 6 of the power supply socket. Wire Black Red Gray Power Input 4 5 7 Green-Yellow The Power / Transformer Input unit must be placed at the bottom of the equipment as shown in the illustration on the side. Pieces of heat shrink must protect the solder joints. Figure 2 The primary wires are soldered to the Power Input according to the table on the side. Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 30 of 77 . Figure 1 After the capacitors have been placed together they must be soldered to the terminals. Audible tones will be emitted after each reading that is successfully completed.6.6 Auto Sense Scanner 4. Use the scanner to read all of the codes in the table below (follow the numeric sequence). Power on the HemoLight Plus.1 Programming the Scanner Connect the Scanner to the equipment’s rear panel (DB9). Reset to Defaults Serial Output Mode Serial Baud Rate 2400 LED Power Enable Safe Mode Single Scan no Trigger Code 39 Enable Code 128 Enable Code 128 ISBT Enable Codabar Enable Interleaved 2of5 Enable CodaBar Star/Stop Chars Enable Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 31 of 77 . These configurations will only operate with the Auto Sense Scanner model as sold by HemoCare.Chapter 4: Repairs and Adjustments 4. The checksum is a digit on the barcode. E. Whilst the equipment is displaying DATE and TIME. CODE ISBT 128. Code 39 TEST8052 Code 128 Fresenius HemoCare Code 128 ISBT =B30010100019102 CodaBar A40156B Interleaved 2of5 12345670 To eliminate the probability of a reading error from the scanner it is recommended that barcodes that have a checksum be used. press the ENTER key.2 Testing the Scanner Power on the HemoLight Plus and whilst the equipment is displaying DATE and TIME. Repeat the same procedure for all other bar codes. The value displayed must coincide with the value show in the third column of the table. and if there is a failure in the reading then the scanner will not transmit the value to the mixer. read the 1st barcode from the table below and wait for the next field to be requested.Chapter 4: Repairs and Adjustments 4. press the ENTER key twice.6.g. When the equipment asks for the OPERATOR. Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 32 of 77 . Press RESET. CODABAR. this digit is a function of the value of the code. of barcodes that have a checksum: CODE 128. repeating the reading until there is a correct relationship. Thus there is a relationship between the value of the checksum and the barcode. 7 Simulating a Donation Process The purpose of the following test is to check the operation of the HemoLight Plus during a donation. Input random data as to the equipment’s requests. Wait for the end of the donation. Suspend the full bag at approximately 40 cm in relation to the oscillating tray. The clamp will open and the tray will start its movement. 04/2011 Page 33 of 77 . On reaching the volume the display will indicate “END OF COLLECTION” and the audible and visual alarms will be active. Procedure: • • • • • • Transfer all of the liquid to one of the bags and pinch the tube to prevent a new transfer. Please consult the operation manual to obtain further information. Press ENTER. Follow up on the donation progress until the display gives a reading of approximately 100ml. Release the flow. After a few moments (according to the configurations) the low flow alarm should sound and the display must warn on the “LOW FLOW”. Power the equipment on. The alarm should be silenced. Press RESET. Place the tube in the clamp. T he data requested at the beginning of the collection and the activation standards of the alarms depend directly on the equipment configuration. Place the empty bag on the tray.Chapter 4: Repairs and Adjustments 4. with the exception of the value of the volume that must be defined as 450ml. At this point block the flow between the bags. Necessary equipment: • Dual Bag System containing 500 ml of liquid. Press ENTER to release the donation. • • • • • • • • • • Fresenius Kabi HemoLight Plus QL-38/08 Once the volume has been confirmed the equipment will request the puncturing and a new ENTER. Length of time that the HemoLight Plus remains powered on. the batteries will not be used. some of these are: Battery type (alkaline or common). set in a compartment on the bottom of the HemoLight Plus. I The batteries life expectancy may vary and is influenced by a number of factors. When the memory batteries on the HemoLight Plus become weak the display on the equipment will display the following message: MEMO BATTERY LOW These must then be changed immediately with the equipment still powered on so as to retain its data. Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 34 of 77 . Therefore the longer that the HemoLight Plus remains powered on the longer the batteries will last. the batteries will only come into operation when the equipment is completely powered off.8 The Memory Batteries on the Equipment To retain all of the collection data during the time that the HemoLight Plus remains disconnected from mains power and without its primary battery connected. either to mains or to the primary battery. Temperature. It is possible that after a long period of inactivity. Whilst the HemoLight Plus remains powered on. Under these conditions the equipment must be reconfigured when it returns to normal usage. The memory power supply unit on the HemoLight Plus is made up of 3 AA alkaline batteries of 1. it is necessary for internal memory batteries.5V each.Chapter 4: Repairs and Adjustments 4. the HemoLight Plus will indicate that the batteries must be changed. when powered on. It is recommended that the batteries be removed from the equipment under long periods of inactivity. Chapter 4: Repairs and Adjustments 4. Never force the batteries into the compartment.1 How to change the memory batteries: Materials necessary to effect the change. Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 35 of 77 . then this must be inserted and if the equipment does not power on automatically the RESET key must be pressed. Before inserting the batteries carefully check to see that there are no signs of leakage. Never insert the batteries with its polarity +/. discoloring. Power the equipment on. If only one or two batteries are changed then there may be the risk of leakage or explosion and the batteries life expectancy will be shortened. start a fire or cause harm. never mix them.inverted. Do not use batteries without the insulating protection or with its protection damaged. deforming or any other anomaly. as it may leak liquid. The change of batteries must be done with the HemoLight Plus powered on. Do not insert anything else into the battery compartment other than the batteries. Three new batteries.8. Carefully insert the batteries as described in this manual. If its power is supplied via the primary battery. if the equipment is not powered on then the memory’s data and its configuration are lost. Even if they have been purchased in this condition do not use it. After the opening animation is displayed press the SCALE key. • • A Phillips screw driver. either with mains power or primary battery. When changing batteries always use new batteries of the same brand. The discarded batteries must be recycled. With a Phillips screwdriver remove the screws and open the battery compartment. neither drag or lift the equipment using the tray. the spring in the holder is where the negative terminal of the battery is placed. if its power is supplied by the primary battery. Close the battery compartment using the Phillips screws. On turning the HemoLight Plus do not support the equipment on its tray.Chapter 4: Repairs and Adjustments Turn the equipment to gain access to the battery compartment. only sufficiently to close the battery compartment without play. the position that the batteries must be inserted is shown in the holder. Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 36 of 77 . take care so as not to loose the screws. The screws that were removed will be used to close the compartment. paying careful attention to the batteries polarity. Look for information at the refuse collection services in your town to get to know the procedures. Do not excessively tighten the screws. Remove the exhausted batteries and insert the new ones. Place the equipment back in its original position and press the RESET key. If there is some question as to polarity. To do this it will be necessary to work all of the keys.1 Keyboard test Before proceeding to change the keyboard it is recommended that its operation be checked.Chapter 4: Repairs and Adjustments 4.9.2 How to replace the keypad Fresenius Kabi HemoLight Plus • Open the equipment and remove the flat cable that connects the CPU board to the display board. 4. • Release the flat cable from the keyboard’s socket.9. • Remove the keyboard by pulling on one of its ends. The keyboard test must be carried out whenever maintenance is done on the HemoLight Plus. • After removing the keyboard the recessed surface must be cleaned so that there are no adhesive residues left behind on the cover QL-38/08 04/2011 Page 37 of 77 .9 The Equipments Keyboard 4. one at a time. • The keyboard must be fitted into the cover’s recess. • Insert the flat cable into the display board’s socket.Chapter 4: Repairs and Adjustments Fresenius Kabi HemoLight Plus • Insert the new keyboard. passing the flat cable through the slot in the cover. QL-38/08 04/2011 Page 38 of 77 . within this select the “Downloader.com/. QL-38/08 04/2011 Page 39 of 77 . How to obtain the TIGER DOWNLOADER: • Access the site: http://www.1. • Reset the HemoLight Plus. • Click on “BASIC-Tiger™ Lite-Version/Downloader”. Transferring a new software version to the HemoLight Plus Fresenius Kabi HemoLight Plus • Open the HemoLight Plus housing gaining access to its main board. this software has the following extension “.2 Necessary Materials: • Serial crossover DB-9F/DB-9F cable • Personal computer • New version of the HemoLight Plus Software.1 5.wilketechnology.1 5. self-install-exe”. which is at the top of the screen.TGU”. and is free.Chapter 5: Software Upgrade on the HemoLight Plus 5 Software Upgrade on the HemoLight Plus The HemoLight Plus allows for its software to be updated without the need to change its microcontroller. JP2 and JP3 on the Main Board to the PC Mode side. To carry out this transfer it will be necessary to use a program called “TIGER DOWNLOADER”. http://www. This process consists of transferring a new version of the software from the PC to the HemoLight Plus. On the initial page access the “downloads” tab ( ). • Place jumpers JP1. this is the software that must be copied and installed on the PC. 5. • Software to make the transfer TIGER DOWNLOADER.wilketechnology. this program is available at the following site.com/. • Start the tiger downloader software.Chapter 5: Software Upgrade on the HemoLight Plus • Connect the Scanner’s input serial cable to the PC’s serial port. • To select the software that is to be recorded onto the HemoLight Plus click on the • Fresenius Kabi HemoLight Plus QL-38/08 button. Select the software and click on Open 04/2011 Page 40 of 77 . Chapter 5: Software Upgrade on the HemoLight Plus Fresenius Kabi HemoLight Plus • The download will start automatically. • When the recording progress bar disappears then the recording has been completed. • Return the jumpers to the initial position. • Remove the crossover cable and close the housing. The mixer must be reset. QL-38/08 04/2011 Page 41 of 77 Chapter 6: Electrical Diagrams and Circuit Description of the HemoLight Plus 6 Electrical Diagrams and Circuit Description of the HemoLight Plus 6.1 Block Diagram Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 42 of 77 Chapter 6: Electrical Diagrams and Circuit Description of the HemoLight Plus 6.2 Connector Positions Indicator Board: CN1 – Display board connector auxiliary/display board. Display Board; CN1 – Connector RJ11 (network). CN2 – Connector DB9 (Scanner). CN3 – Connector board CPU/ display board CN4 – Connector board display/auxiliary display board. CN5 – Audible alarm connector CN6 – Display. CN7 – Membrane keypad connector. CPU Board: (FS1 E FS2) – Power input AC ± 12V. CN1 – Connector board CPU/display board. CN2 – Primary Battery (BAT1). CN3 – Connector board CPU/ display board (Version same as or before 01303V02). CN4 – Memory batteries (Bat 2). CN5 – Homogenization motor unit. CN6 – Clamp. TP1 – Test point 1 (output of the 7812, U9). TP2 – Test point 2 (output of the 7805, U10). TP3 – Test point 3 (output of the 7809, U11). TP4 – Test point 4 (output of the 7805, U12). TP5 – Test point 5 (output of the 7805, U13). TP6 – Test point 6 (output of the 7809, 4). TP7 – Test point 7 (output of the REF 02, U8). TP8 – Test point 8 (V BAT). TP9 – Test point 9 (display contrast). TP10 – Test point 10 (A/D CONVERTER INPUT). TP11 – Test point 11 (output 7660 - 9VDC). TP12 – Test point 12 (power supply battery). TP13 – Test point 13 (power supply V+). JP1 – Jumper (disables the network when it is set between the Mode and PC pins). JP2 – Jumper (Enables the equipment to receive a new software version when it is set between the Mode and PC pins). JP3 – Jumper (disables the buzzer when it is set between the Mode and PC pins). JP4 – Jumper (board compatibility, when removed from the CPU board compatible with the previous display board model). JP5 – Jumper (prevents the equipment going into standby when it is set between the Mode and PC pins). Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 43 of 77 3.3 PCB 01401V06 6.Chapter 6: Electrical Diagrams and Circuit Description of the HemoLight Plus 6.1 Layout of the Printed Circuit Board PCB 01401V05 Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 44 of 77 . The rectifying of the incoming voltage is done by bridge rectifier BR1. Diodes D11. The battery compartment is accessible from the side of the equipment. If a battery is connected to the equipment and the equipment itself is connected to a mains outlet with the ON / OFF switch in the “ON” position.Chapter 6: Electrical Diagrams and Circuit Description of the HemoLight Plus 6.2 DC Power Supplies The HemoLight Plus has stabilized DC power supplies assembled with linear regulators of the 78XX family (with the exception of the primary battery charging regulator and the A/D reference converters).3. The charger on the HemoLight Plus takes around 10 hours to fully charge. D87 and D8 prevent the battery supplying current to source or even to its charging circuit. Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 45 of 77 . then the battery will be charged.3.3 Regulator Supplied Circuit U14 (7809) U10 (7805) U12 (7805) U11 (7809) U13 (7805) U9 (7812) U17 (L200) U8 (REF 02 SMD) Scale Digital Circuits Scanner Motors Motor Relays General Relay Battery Charger A/D Reference Converters Primary Battery Circuit The HemoLight Plus is able to operate with power coming from a rechargeable lead-acid battery. The primary battery is connected in parallel to bridge rectifier BR1. 6. There is only one way to connect and it is impossible to invert the polarity. When fully charged it is able to work under standalone for approximately 3 hours of effective collecting (approximately 15 complete procedures). The HemoLight Plus functions will be compromised. If there is a current circulating. The regulator limits the maximum charging voltage to 14. The discarded batteries must be recycled. guaranteeing that the yellow LED lights up. The voltage across R40 represents 1/3 of the battery’s voltage. The comparator built with U15 monitors the current circulating through the charging circuit. Remove it from the compartment whenever the equipment is to remain inactive for a long period of time. The batteries have a limit to their charge/discharge cycles. In this instance it must be replaced with a new one. LED LD1 on the indicators board is a bicolor LED which turns green when the battery is being charged and red when the battery is discharged. If the battery is removed or if the mains power is removed then the output of U15 (pin 6) will be low thus preventing LD1 turning green. The battery’s charging condition is digitally checked from an A/D converter.5 Volts and limits the circulating current to 200mA. A comparator circuit is built with an operational amplifier (U16 port D). Whenever the equipment is being supplied by the battery. the yellow indicator LED 4 will light up. (The equipment on standby continues to consume battery power).7 V) value the processor will notify the Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 46 of 77 . The charging circuit is built with a L200C (U17) regulator mounted so as to supply a controlled voltage and current to the battery when it is being charged. Do not incinerate. pin 6 of U15 will be high. If they do not fully recover their capacity.Chapter 6: Electrical Diagrams and Circuit Description of the HemoLight Plus Charge the battery before using it for the first time and then immediately charge it again after use. Under this condition the bicolor LED will be green indicating that the battery is being charged. Never fully discharge the battery. Do not continue to use the equipment if it indicates that battery charge level is low. In the absence of an external power source the supply of the circuits will be taken over by the battery. Look for information with the local refuse collection services in your town to know how to proceed. The switching between the battery and mains power is done automatically. The input to the A/D converter (pin 39 U2) is connected to resistor R40. In the absence of an external power source the comparator’s output is set to zero volt. Whenever the battery’s voltage is less than the critical (11. The sulfuric acid may cause injury to the skin and other parts of the body. then their working life is at an end. Recharge the battery only with the HemoLight Plus charger or charges specifically for this purpose. Do not try to disassemble the battery. This signal is transferred via operational amplifier U16 PIN 8. When the SET coil receives a voltage pulse the relay closes a contact that will be opened again when the RESET coil receives a pulse. polarizing transistor Q4 activating the RESET coil on the relay. Pin 10 on U2 will go high causing LD1 to turn red. Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 47 of 77 . Each of its coils must be energized individually when necessary. it will automatically enter the “Standby” mode after being idle* for 20 seconds. Tiny Tiger. which will switch the general relay through transistor Q3. via reconnection to mains power. R41 and R36. When this happens and if the equipment is idle then the controlling software triggers a timer. it will be necessary to remove jumper JP4. but to be able to use it. The CPU detects the absence of an external power source through the resistor bridge made up of R30. If the trigger reaches the programmed time. There are two possibilities for powering the equipment on again. This process is controlled by the CPU that carries out the powering off process through a relay (K4) with reversible contacts and a double holding coil. The current version of the CPU board is compatible with the previous display board version. The first.Chapter 6: Electrical Diagrams and Circuit Description of the HemoLight Plus operator that the battery must be changed. The HemoLight Plus is then powered off at this point. NOTE: The ON / OFF switch does not power the equipment off when there is a battery connected.3. then processor U2. will signal the relay with a high from its pin 11 (Standby). The second possibility is by pressing the RESET switch on the equipment. which will switch the general relay through transistor Q2.4 Standby Circuit With the intuition of increasing the standalone time of the HemoLight Plus. This relay does not need to remain energized after switching. *The HemoLight Plus is considered idle only when the display is alternating between date and time. the equipment’s general relay. and use the auxiliary display Board Cable (234985018). Look for information from your local refuse collection services to see how to proceed 6. by presenting a message on the display. The disposed batteries must be recycled. clamp and alarms. Pins 16 and 17 when high respectively enable the red LED alarm and buzzer. Its operation is managed through software. This configuration allows for a Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 48 of 77 . D Octal TTL Flip Flop 74LS373 (U1) and Octal TTL Transceiver 74LS245 (U7).5. 3.2 Driving the Peripherals The following are considered as peripherals.3.Chapter 6: Electrical Diagrams and Circuit Description of the HemoLight Plus 6. and 33 of the Tiny Tiger micro-controller are used to control the display and keyboard matrix of the HemoLight Plus. Pin 15 when high opens the clamp. and analogically. 20. The keypad sound is generated by pin 20 of the Tiny Tiger. 31. 6.5 Micro-controlled CPU The HemoLight Plus is a micro-controlled homogenizer. Whenever a key is pressed a line is shorted out with a column of the keypad matrix generating a binary code interpreted by the Tiny Tiger data bus.1 Keypad and Display Matrix Driving Peripherals Serial Interface Analogue Inputs Controlling the Keypad and Display Matrix Pins 1. 5.3. causing the transistor to enter into the cut off region and activating the buzzer. 6. Pin 14 when high (5Vcc) starts the tray movement. 8. This pin is always high (5Vcc) keeping transistor Q1 constantly saturated. The display’s contrast is adjusted by potentiometer R5. The red LED and buzzer are connected to the logical inverter 74LS04 (U4) output ports. oscillating base circuit.5. The CPU is made up of 4 different circuits: • • • • 6. 2. All of the CPU’s integrated circuits use TTL technology. the same pin when low closes the clamp. Tiny Tiger Pin Outs 1 to 8 20 31 32 33 14 15 16 17 Function Data bus Keypad sound Enabling the Keypad Enabling the Display Reading or Writing Operation on the Display Driving the oscillating tray Driving the Clamp Alarm LED Buzzer To make up the keypad matrix three integrated circuits are used: Demux TTL Decoder 74LS138 (U5).3. 4. 32. 7. Power Supply = 5Vcc (U10). When the keypad sound is enabled and a key is pressed then pin 20 goes low 0 Volt. 3.5. The network protocol is controlled via software. Pin 18 of the Tiny Tiger is the SDA data pin and pin 13 is the SCL clock. the addressed HemoLight Plus enters into a transmission condition (Pin 28 high). The HemoData NET module is made up of a PC (MASTER) and up to 32 HemoLight Plus (SLAVES) all of them connected to the same differential pair (network).4 I2C Serial Interface The HemoLight Plus records the data on the Smart Card Memory through the socket located on the CPU board. (Pin 28 low). The other is used to exchange data with a PC (RS485) when the HemoData NET network module is installed. in which the HemoLight Plus is the master and the Smart Card Memory is the slave. Pins 26 and 27 of the Tiny Tiger are connected to a DC-DC converter Max232 (U3).Chapter 6: Electrical Diagrams and Circuit Description of the HemoLight Plus slightly higher current level than which is supplied by the microcontroller’s pin. 24 and 28 of the Tiny Tiger module are connected to a RS485 Transceiver 75176 (U6).3. In configurations of this type it is possible to build a network where all of the equipment that is powered on is identified. Pins 23. The Oscillating Tray and Clamp circuits will be detailed later 6. This configuration allows for data with different logic levels other than 5Vcc to be read without the serial port being damaged. The initial information is sent as 9 Bits and always comes from the PC. Tiny Tiger Pin Outs 23 24 26 27 28 Function RS485 TX RS485 RX Scanner TX Scanner LX RS485 Control Pin The scanner communicates with the HemoLight Plus at a transmission rate of 2400 Baud. Communications between the two is done through the I2C protocol. (A direct connection to a PC is possible). When the network is inactive all of the HemoLight Plus will be in a listening condition. Connector RJ11 – 2 active routes – RS485 Interface Transmission rate 38400 Baud 6.3 RS232 Serial Interface The HemoLight Plus has two serial ports.5. Once the PC sends a piece of information. Pin 28 of the Tiny Tiger defines the communications direction: High (5Vcc) – HemoLight Plus transmitting Low (0Vcc) – HemoLight Plus receiving data For it to work on the network it is necessary to implement a HemoData NET module. Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 49 of 77 . The output of the 75176 is made up of a differential pair. One of them is used by the Scanner to obtain the data from reading a barcode. 5. This signal is used to check the charge condition of the memory batteries. This configuration affords impedance matching thus preventing that the memory’s batteries discharge across the A/D converter when the equipment is powered off. The load cell is of the “single point” type. This circuit is assembled in the following manner: The voltage signal for the batteries goes through a voltage (buffer) made up of a LM324 (U18). Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 50 of 77 . Analogue channel 0 is pin 37 of the Tiny Tiger. amplification circuit. U26). U25. Pin 19 of the Tiny Tiger. and then a new operation may be carried out. 350 Ω for up to 5Kg. This filter is made up of an Op Amp LM324 (U18) a capacitor (C38) and a resistor (R52). It is used to limit the output of the amplification stage to this value. After amplification the signal passes through a high pass filter. Whenever a piece of data is recorded or read the slave confirms the operation. This potentiometer is used to calibrate the scale and is accessible to the equipment user.9 V). The voltage value across this pin will be proportional to the weight placed over the load cell.3. 6.9 V so that is able to obtain a symmetrical power supply for amplification (+ 9 V and . It is a 12 bit A/D converter used to obtain the amplified signal from the load cell. The entire scale circuit is supplied with 9Vcc (U21). and A/D pre-converter filter. The converter 7660 (U24) generates a signal of . Analogue channel 1 is pin 40 of the Tiny Tiger (U1). This channel monitors the primary battery voltage. The amplifier circuit is an instrumentation amplifier built up of 3 op.3. To increase its sensitivity this capacity is reduced by the by the amplifier circuit to 1kg. when low will delete the data from the Smart Memory Card Memory. 6. The input of the third logic channel is pin 39 of the Tiny Tiger (U1).3. and if the data is recorded or read. please consult item 6.Chapter 6: Electrical Diagrams and Circuit Description of the HemoLight Plus For a piece of data to be recorded it will be necessary for the master to send a start signal. Please consult item 6. The amplifier’s gain is changed through potentiometer R53. At the input to the filter there is a 5V6 Zener diode.5 Analogue Inputs The HemoLight Plus uses three analogue inputs to obtain external data.3. amps OP07 (U23.6 The Scale Circuit The Scale’s Circuit is made up of the load cell. the data will follow in sequence. It is also a 12 bit analogue converter used to obtain the voltage on the batteries of the HemoLight Plus. So that the data is valid it will be necessary for the SCL signal to be high.3. after this signal has been sent the address will follow.6 for information on this circuit. The motor will make the closing movement when pin 15 on the Tiny Tiger is at 0V. When the clamp reaches its maximum opening. Whenever one of them is activated a 0V signal is transmitted to one of the “AND” gates.8 The Clamp Circuit The clamp circuit basically constitutes of a 9Vcc motor and 3 logic ports. A high signal from pin 15 (Clamp) from the Tiny Tiger causes pin 3 of the 74LS08 (U19) to go high causing the optocoupler U20 to close the relay. Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 51 of 77 . the optocoupler U21 will cause relay K2 to close thus placing the motor into operation.Chapter 6: Electrical Diagrams and Circuit Description of the HemoLight Plus 6.3.3. the 9Vcc voltage across the motor is inverted causing the motor to take up both rotation directions. The optocoupler is used to isolate logical circuits (control) of the motor circuit. that for each condition.7 The Oscillating Tray Circuit The tray movement is carried out by a 9Vcc motor. This signal undergoes an inversion at the “NOR” 74LS04 (U4) gate causing the same effect as previously mentioned on the gate. with two of them being “AND” 74LS08 gates and one "NOR” 74LS04. Whenever instructed by the Tiny Tiger. The operating logic is as follows: The clamp has two end of travel switches. then the end of travel switch will be activated causing the logic of the “AND” gate to return to 0V at the output and interrupting the motor’s operation in this direction. thus starting the clamp opening movement. 6. Note that on the circuit. Standby and Memory Batteries HemoLight Plus QL-38/08 04/2011 Page 52 of 77 .9 Electronic Schematics PCB 01401V06 6.9.Chapter 6: Electrical Diagrams and Circuit Description of the HemoLight Plus 6.3.3. Primary Battery Charger.1 Fresenius Kabi Electronic Schematics for the Power Supplies. 3.9.2 Fresenius Kabi Electronic CPU Circuit Schematics HemoLight Plus QL-38/08 04/2011 Page 53 of 77 .Chapter 6: Electrical Diagrams and Circuit Description of the HemoLight Plus 6. 3.9. Clamp and Scale Circuit Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 54 of 77 .3 Electronic Circuit Schematics for the Tray.Chapter 6: Electrical Diagrams and Circuit Description of the HemoLight Plus 6. 4 PCB 01403V02 6.1 Operator Interface 6.4. The keypad matrix is made up of 3 columns and 8 lines.4.Chapter 6: Electrical Diagrams and Circuit Description of the HemoLight Plus 6. being that two of the lines are placed to one side generating an effect of a single line with 16 positions.1.4.1. 6. The equipment has a matrix of 24 keys being that only 17 of them are being used.2 Function GND 5Vcc Contrast control Display reading or writing operation GND Display Enable Data bus N/C RS232 Serial Communications Interface RS232 Scanner Input: DB9 Connector Pin on the DB9 Connector 3 Function RX 2 TX 5 GND 1 5Vcc RS485 Communications Input: RJ11 Connector 2 active paths Transmission Rate 38400 Baud Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 55 of 77 .4.2 Display The display is of the LCD type with 2 lines/8 columns. Display Pins 1 2 3 4 5 6 7 to 14 15 and 16 6.1 Keyboard The keypad on the HemoLight Plus is of the Membrane type. 4.4 Electronic Schematics PCB 01403V02 and PCB 01410V01. Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 56 of 77 .Chapter 6: Electrical Diagrams and Circuit Description of the HemoLight Plus 6.4. 6.3 Printed Circuit Board Layout PCB 01403V02. LED2. Please consult items 6. LED3.1 Printed Circuit Board Layout PCB 01410V01 Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 57 of 77 .4. red.6. The bicolor LED takes on the green color when the battery is being charged and red when it is discharged. 6. and is always active when the I/O switch on the equipment is in the ON position.9.3. then the yellow indicator will light up. 6.9.3. is the visual alarm for the equipment. Please consult item 6.Chapter 6: Electrical Diagrams and Circuit Description of the HemoLight Plus 6.3. green. Whenever the battery is supplying power to the equipment.5 PCB 01410V01 6.2.6 Indicators Interface The visual indicators are mounted with LED’s. There is also a yellow LED4 and one bicolor LED1 that alternates between the colors green and red. indicates that the equipment is powered on.3.3.9 and 6.2 and item 6.4. Green indicates that the Memory Card Reader is powered on. 3. indicates that the Smart Card Memory has been inserted into the Memory Card Reader. Input for the Smart Card Memory. RS485 serial output. 3. Red indicates that the data from the Smart Card Memory is being transferred to a PC.Chapter 7: The Memory Card Reader 7 The Memory Card Reader The Memory Card Reader is made up of: 1. 9 V DC 50 mA. RS232 serial input. Please check the insertion position on the Smart Card Memory label. Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 58 of 77 . 4. To open the Memory Card Reader. 2. 2. Blue. A power supply input. It has the following Indicator LEDS: 1. on the bottom must be loosened. The conversion from RS485 to RS232 takes place through Transceiver 751769 (U4). Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 59 of 77 .Chapter 8: Electrical Diagrams and a Circuit Description of the Memory Card Reader 8 Electrical Diagrams and Circuit Description of the Memory Card Reader 8. The data is read using a PIC 16F628 (U6) microprocessor.1 PCB 01901V1 8. pin 7 on the PIC is connected to pin 1 of the 75176 and pin 6 of the PIC is connected to pin 4 of the 75176.1 Memory Card Reader Circuit It may be used as a converter to use the HemoLight Plus in a network or as a Memory Smart Card reader. through pins 8 (SCL) clock signal and pin 9 (SDA) data. The Memory Card Reader has a RS 485 to RS 232 converter and a RS 485 to I2C converter. For the data on a card to reach the PC it must first be read using the I2C protocol.1. Pin 8 on the PIC is connected to pin 3 of the Smart Card Memory socket and pin 9 on the PIC is connected to pin 6 of the socket. The converters are bidirectional. And so they are connected in parallel to the RS 485 network. The PIC is connected to a RS 485 Transceiver 751769 (U5). transferring the data from the card to the PC. which is connected to a DC-DC Max232(U3) converter. pin 1 of the 75176 (RX) is connected to pin 11 of the MAX232 and pin 4 of the 75176 (TX) is connected to pin 9 of the MAX232. 1.Chapter 8: Electrical Diagrams and a Circuit Description of the Memory Card Reader 8.2 Printed Circuit Board Layout PCB 01901V01 Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 60 of 77 . 1.3 5 .Chapter 8: Electrical Diagrams and a Circuit Description of the Memory Card Reader Electronic Schematics PCB 01901V01 CN1 1 2 3 1 6 2 7 3 8 4 9 5 R X + C1 100nF C5 C10 10uF/50V + 10uF/50V U1 LM7805 VO +V C4 100nF C1+ C1C2+ VCC GND C2 100nF 1 3 4 5 VCC + + C3 470uF/25V C6 C11 10uF/50V + 10uF/50V VCC R1 2K2 C7 100nF U4 RD R/E/ RX TX 8 VCC SN75176 + - 6 7 VCC R2 4K7 BUS485 R8 4K7 6 7 C8 VCC U5 100nF + GND GND 1 2 3 4 SN75176 - VCC 4 2 3 1 LD1 GREEN 3mm C12 10nF - RX RD R/E/ TX 1 2 3 4 VCC + R3 4K7 D T R C13 47uF/25V 17 18 1 2 3 4 6 C9 100nF U6 RA0 RA1 RA2 RA3 RA4 RX INT VCC 16 13 12 11 10 9 8 RA6 15 RA7 PGD PBC RB5 PGM CCP1 TX PIC16F628 VSS 5 01901V01 7 14 VDD RJ11 CN3 5 + - + 8 11 10 12 9 10 VCC T1IN C2- T1OUT T2IN R2OUT R1OUT T2OUT V- V+ U3 VI 2 6 14 7 16 T X D T R MAX232 R2IN 13 R1IN 8 9 15 GND VCC X1 20MHz R7 330R VCC R4 2K2 R10 4K7 LD2 YELLOW 3mm Q1 BC337 4K7 R9 VCC LD3 BLUE 3mm R5 47K S A D S L C K Y E A C R D VCC R6 10K CN4 1 2 3 4 5 6 7 8 9 10 SMART CARD Page 61 of 77 04/2011 DC POWER CN2 DB9-F 8 7 6 5 1 2 3 4 RST CL NC NC QL-38/08 HemoLight Plus Fresenius Kabi SCL SDA GND 8. A movable locking piece guarantees that the accessories remain in their place even if there is sudden rough movement. which are subject to damage if poorly handled. But. The bed is supported on a “shock absorber”. To clean it. it is important to consider that even transported in this manner the HemoLight Plus is still a piece of equipment with very sensitive measuring devices. Drops and knocks must be avoided. do not transport any foreign object inside the case. The transport case does not need preventive maintenance. The mixer must also be handled with the appropriate care even when housed in the transport case. • To open the case the four catches on the sides must be unlocked. If the Smart Card Memory is being used. which absorbs part of the vertical impacts. battery and cables. • • Fresenius Kabi HemoLight Plus QL-38/08 04/2011 The lower bed of the case has niches to hold the Scanner. The transport case was designed to transport the HemoLight Plus and its accessories (scanner and battery). do not close the case with the card inserted in the HemoLight Plus. Page 62 of 77 . use a soft cloth dampened with 70% alcohol.Chapter 9: HemoLight Transport Case 9 HemoLight Transport Case The transport case was designed to protect the equipment from eventual shocks during transport. when closed this backing holds the HemoLight Plus to the lower bed. The HemoLight Plus when placed in the transport case will sit between the four corner pieces on the lower bed. The front part of the HemoLight Plus must face the scanner’s “head”. To do this the cover must be placed with the opening facing upwards and the lower bed must then be placed on top of the cover. preventing the mixer from shifting position during its transportation. HemoLight Plus QL-38/08 04/2011 Page 63 of 77 .Chapter 9: HemoLight Transport Case • • Fresenius Kabi • The case’s cover has an internal backing. • The transport case may also be used as a collection table. Measurement Situations S1 closed and S2 normal For other conditions S1 closed and S2 inverted I ≤ 1000µA S1 open and S2 normal S1 open and S2 inverted HemoLight Plus QL-38/08 04/2011 Page 64 of 77 . A condition is considered a normal one when S1 and S2 not inverted.1 Leakage Currents Through the Protection Connector NORM OF REFERENCE IEC60601-1 Fresenius Kabi STANDARDS For a normal condition I ≤ 500µA METHOD OF ANALYSIS The equipment must be tested according to the figure below.Chapter 10: Safety End Tests 10 Safety End Tests 10. DM is the measurement device and will indicate the value of the leakage current. All the combinations of the keys S1 and S2 must be tested. DM is the measurement device and will indicate the value of the leakage current. All combinations of the keys S1. S2 and S3 must be tested. The measurement point is shown at the detail For a normal condition I ≤ 100µA For other conditions I ≤ 500µA Fresenius Kabi HemoLight Plus METHOD OF ANALYSIS QL-38/08 Measurement Situations S1 closed S2 normal S1 closed S2 normal S1 closed S2 inverted S1 closed S2 inverted S1 open S2 normal S1 open S2 normal S1 open S2 inverted S1 open S2 inverted 04/2011 S3 closed S3 open S3 closed S3 open S3 closed S3 open S3 closed S3 open Page 65 of 77 . A condition is considered a normal one when S1 and S3 are closed and S2 not inverted.2 Leakage currents through the cabinet (surface / ground) NORM OF REFERENCE IEC60601-1 STANDARDS The equipment must be tested according to the figure below.Chapter 10: Safety End Tests 10. The measurement point is shown at the detail below. A condition is considered a normal one when S1 and S3 are closed and S2 not inverted. S2 and S3 must be tested. IEC60601-1 For a normal condition I ≤ 100µA For other conditions I ≤ 500µA Fresenius Kabi HemoLight Plus Measurement Situations S1 closed S2 normal S1 closed S2 normal S1 closed S2 inverted S1 closed S2 inverted S1 open S2 normal S1 open S2 normal S1 open S2 inverted S1 open S2 inverted QL-38/08 04/2011 S3 closed S3 open S3 closed S3 open S3 closed S3 open S3 closed S3 open Page 66 of 77 . All combinations of the keys S1.3 Leakage currents through the cabinet (surface / surface) NORM OF REFERENCE STANDARDS METHOD OF ANALYSIS The equipment must be tested according to the figure below.Chapter 10: Safety End Tests 10. DM is the measurement device and will indicate the value of the leakage current. To run this test. and the impedance is calculated with basis on the current and voltage. The difference of voltage between the described parts is measured.Chapter 10: Safety End Tests 10.4 Ground Bound Test NORM OF REFERENCE STANDARDS IEC60601-1 Z ≤ 0. QL-38/08 04/2011 Page 67 of 77 . the On / Off switch of the equipment must be at the On position. The circuit and the measurement points are shown in figure below.1Ω Fresenius Kabi HemoLight Plus METHOD OF ANALYSIS A 25 A current from a current supply with 60Hz frequency and with 6V open-circuit voltage must be put through the protection ground terminal on each metallic part available. 3 2.2 2.3 04/2011 Method Repaired Component / Evaluated Characteristic Nº Checked Maintenance Checklist Item in this Manual 11 Use a soft sponge or clothe dampened with a neutral detergent.4 3 3. breakage or loss of insulation Internal cable Connections in the 6.1 user manual       Visually Inspect   Visually Inspect  Visually Inspect   Visually Inspect   Visually Inspect   Connect the primary battery to the equipment.1 Subject the equipment to conditions where all of the keys are employed. being completely powered off.3 4.3.5 Fuses 1.9 Inside of the housing free from dust and other residues Internal wires and cables Condition of the wires without any signs of and cables. The equipment should enter into STANDBY mode. There may be no signs of breakage or loss of insulation Check the switch operation According to the specifications mentioned in item 2.1 2. without blood or other fluid residues Mounting pins intact / adequate fitting Power cable and connection ON / OFF Switch 3.1 4.1 General Cleanliness 1.   Keep the equipment supplied with power from the battery and wait for approximately 20 seconds. and if the battery in use (yellow LED) remains OFF. Keep the equipment supplied with mains power and check if the Power-ON (Green LED) indicator lights up.5 6.2 2 4.2 Tray 1.3 Electrical Power Input Housing and Tray clean. Example – Simulate a donation According to item 7.5 Labels Legible and without signs of coming loose 2 Inside of the Housing 2.   Page 68 of 77 .   Rock the I/O switch to the OFF position and see if the yellow LED lights up. Operation Via AC Adapter and Battery General Cleanliness Operation of the power supply circuits Fresenius Kabi Equipment operating efficiently when supplied with power from a battery or AC mains HemoLight Plus QL-38/08 3.2 connections correct positions Oscillating and clamping Fastening of the internal systems as well as components boards in the correct positions. To disinfect use an alcohol solution of a 70%.  Visually Inspect     Visually Inspect the cable.4 Membrane Keypad All keys responding / without locking up 1.Chapter 11: Maintenance Checklist Description 1 Outside of the Housing 1. 3 in this manual Other Conditions (the greatest value)  I ____µA Impedance of the Protection Ground QL-38/08  I ____µA 10 Normal Condition HemoLight Plus  According to item 10.4 Leakage Currents Through the Protection Connector Leakage Currents Through the cabinet (surface / ground) Leakage Currents Through the cabinet (surface / surface Ground Bound Fresenius Kabi Normal Condition I ____µA According to item 10. clamp 4.1 9 9.3 4.1 Calibration 6 Scanner (Optional) Operational reading of Correct reading 4.1 Homogenization Unit 4.         According to the definition in item 4.2 in this manual   According to item 4. then rock the mains (I/O) power switch to the ON position and see if it comes on again.Chapter 11: Maintenance Checklist 4 Peripherals 4.1 8 8.1 in this manual Other Conditions (the greatest value) I ____µA Normal Condition Other Conditions (the greatest value) I ____µA According to item 10.9 5.2 test.3 9.1 7 7.4 in this manual.4 Carry out the initial test routines as described in no item 3. Safety End Tests 6.2 Deviation not more than 2g when calibrated at the 4. 6.4.4 5 Clamp Occlusion (closing effectiveness) Weighing System Tray movement is unhindered and without any sections catching against the housing LED’s and Buzzer operational Opening and closing movement Occlusion at 1BAR Press the RESET key and check if the equipment powers on again.9.2 in this manual I ____µA According to item 10.7 According to item 4.6 software carry out a standard programming and Operation 3.1 in this manual the bar codes Data Communications (Only on equipment with an active network module) 3.4 6. 4.7 in this manual during a donation and alarm systems.6.6 According to item 4.2 Operational Test Operability of the Equipment behavior homogenization.3 Clamping System 4.3.6 (500 g) 6.6 in this manual         4.3.   Wait again for 20 seconds for the equipment to power off.3.2 9.4.4 in this manual 04/2011 Z____Ω  Page 69 of 77 .3.3.2 Audible and Visual Alarm 4.5 Using the HemoData NET management Response to 3.1 9.7 configuration on the equipment under Configuration via PC 6.4 center of the scale 6. Tools and Accessories Lists 12 Replacement Parts. Tools and Accessories Lists Position Code 01 234985003 02 03 04 234985036 234985048 234985006 Fresenius Kabi Description HemoLight Plus Assembled Tray Tray + Support + Oscillating unit base HemoLight Plus Assembled Cover on Indicator Board HemoLight Plus Mounted Display Board HemoLight Plus QL-38/08 04/2011 Page 70 of 77 .Chapter 12: Replacement Parts. Chapter 12: Replacement Parts. Tools and Accessories Lists Position 01 02 03 04 05 06 07 08 09 09 Code 234985013 234985155 234985014 234985189 234985064 234985187 234985060 234985191 234985192 234985193 Fresenius Kabi Description HemoLight Plus RJ11 Connector HemoLight Plus DB9 Connector HemoLight Plus Buzzer Connector Upper Label Membrane Keyboard Cut Hand Label (crushing) Primary Battery Indication Label Power Input Label HemoLight Plus Technical Label/ Injected Housing Europe HemoLight Plus Technical Label/ Injected Housing USA (81050) HemoLight Plus QL-38/08 04/2011 Page 71 of 77 . Chapter 12: Replacement Parts. Tools and Accessories Lists Position 01 02 03 04 Code 234985088 234985010 234985040 234985089 Description “Groove Free” Driving Unit Load Cell with Connector Load Cell “U” Protection System Inverted Logic Clamp Unit Position 01 02 03 04 05 Code 234985045 234985016 234985061 234985037 234985103 Description HemoLight Plus Main Board HemoLight Plus Power Supply Input (w/ transformer) Battery Compartment Label HemoLight Plus Mounted Bottom Rubber Feet Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 72 of 77 . 1 List of Replacement Parts only for serial numbers up to 01805108 E-code 000.Chapter 12: Replacement Parts. Position 01 02 Code 234985044 234985023 Fresenius Kabi Description Auxiliary Panel Board HemoLight Plus Assembled Cover (UL94 V0) HemoLight Plus QL-38/08 04/2011 Page 73 of 77 . Tools and Accessories Lists 12. Tools and Accessories Lists Position 01 02 Code 234985067 234985063 Description Upper Label Membrane Keyboard Position 01 Code 234985024 Description HemoLight Plus Assembled Base Fresenius Kabi HemoLight Plus QL-38/08 04/2011 Page 74 of 77 .Chapter 12: Replacement Parts. Chapter 12: Replacement Parts. HemoLight Plus Assembled Scale Board 234985021 Only for serial numbers up to 01805108 E-code 000. 234985038 Display board cable for indicators. 234985017 Flat Cable 40 Way HemoLight Plus Assembled Network Board 234985020 Only for serial numbers up to 01805108 E-code 000. 234985049 Miniature Fuse 200 mAT 5 x 20 mm 234985050 Miniature Fuse 1AF 5x20 mm 234985152 Self Adhesive Fastener REKLIP RK1 234985153 Self Adhesive Fastener REI01SP 234988003 Fresenius Kabi HemoLight Plus QL-38/08 04/2011 RS 485 Network Cable Page 75 of 77 . Tools and Accessories Lists Illustrative Drawing Code Description 234985186 Primary Battery Cable 234985018 Auxiliary display board cable Only for serial numbers up to 01805108 E-code 000. Chapter 12: Replacement Parts. RS485 cable and a RS232 cable) Page 76 of 77 . Tools and Accessories Lists Illustrative Drawing Code 234985147 Medical Grade Standard Power Cable 234985150 Continental Europe Power Cable 234985151 Power Cable 070187003 Keypad with Display for adjustments 81053 Fresenius Kabi HemoLight Plus QL-38/08 Description 04/2011 Memory Card Reader + RS232 To RS 485 Converter (Made up of a converter. power supply. 070188006 Memory SmartCard For HemoLight Plus 070188005 HemoLight Transport Case 04/2011 Page 77 of 77 .Chapter 12: Replacement Parts. Tools and Accessories Lists 070188004 T Distributor 070187007 81043 Serial crossover cable to upgrade software AKKU-Battery Clamp Unit 234985086 Fresenius Kabi HemoLight Plus QL-38/08 Only for serial numbers up to 01805108 E-code 000.
Copyright © 2024 DOKUMEN.SITE Inc.