AESTIVA - Service Manual

March 26, 2018 | Author: Vinicius Belchior da Silva | Category: Sterilization (Microbiology), Nature


Comments



Description

Software Revision 1.X Operation Manual Setup Part 2 Cleaning and Sterilization Maintenance and Troubleshooting User Responsibility This Product will perform in conformity with the description thereof contained in this operating manual and accompanying labels and/or inserts, when assembled, operated, maintained, and repaired in accordance with the instructions provided. This Product must be checked periodically. A defective Product should not be used. Parts that are broken, missing, plainly worn, distorted, or contaminated should be replaced immediately. Should repair or replacement become necessary, Datex-Ohmeda recommends that a telephonic or written request for service advice be made to the nearest Datex-Ohmeda Field Service Support center. This Product or any of its parts should not be repaired other than in accordance with written instructions provided by Datex-Ohmeda and by Datex-Ohmeda trained personnel. The Product must not be altered without the prior written approval of Datex-Ohmeda’s Quality Assurance Department. The user of this Product shall have the sole responsibility for any malfunction which results from improper use, faulty maintenance, improper repair, damage, or alteration by anyone other than Datex-Ohmeda. w CAUTION U.S. Federal and Canadian law restrict this device to sale by or on the order of a licensed medical practitioner. Outside the U.S.A. and Canada, check local laws for any restriction that may apply. Datex-Ohmeda products have unit serial numbers with coded logic which indicates a product group code, the year of manufacture and a sequential unit number for identification. AAA A 12345 This alpha character indicates the year of product manufacture and when the serial number was assigned; “C” = 1999, “D” = 2000, “E” = 2001, etc. “I” and “O” are not used. 1006-0802-000 07/01 Table of Contents 1 Introduction How to use this manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2 Symbols used in the manual or on the equipment . . . . . . . . . . . . . . . . . . . . 1-3 Maintenance summary and schedule . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6 Operator maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6 Datex-Ohmeda approved service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7 2 Cleaning and Sterilization Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2 Patient path . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2 Scavenging path . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3 Clean and sterilize . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4 To wash (by hand or machine) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4 Autoclave . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5 Special requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6 Assemble . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6 Disassemble the patient path . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7 Canister disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-10 Disassemble the scavenging path . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-12 How to clean and disinfect the flow sensors . . . . . . . . . . . . . . . . . . . . . . . .2-13 CIDEX sterilization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-13 Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-13 How to clean and sterilize the optional CO2 Bypass assembly . . . . . . . . .2-15 1006-0802-000 07/01 i . . . . . . . . . . . . . . . . . . 3-12 DIN connections . . . . . . . . . . . . . . . . . . . .3-6 Scavenging from the external manual breathing circuit . . . . . . . . . . . . . . . . . 4-18 100% O2 sensor calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-17 4 User Maintenance Repair policy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12 Cylinder yokes . . . . . . . . . . . . . . . . . . . . . 3-14 How to attach equipment to the top shelves . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-18 21% O2 sensor calibration . . . . . . . . . . . . . . . . . . . . . . . . 4-23 Flow sensor calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-3 Exhalation valve maintenance . . . . . . . . . . . . .4-6 Receiver maintenance (active gas scavenging only) . . . . 4-10 Bellows assembly maintenance . . . . . . 3-15 Top of machine . . . .4-2 Manifold maintenance . . . . . . . . . . . . . . . . .Aestiva/5 7100 3 Setup and Connections Breathing system setup . . . . . .3-6 Canister Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-5 Scavenging the ACGO sample flow . . . . . . . . . . . . 4-15 O2 sensor calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-9 Breathing circuit maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-9 How to install gas cylinders (high pressure leak test) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-7 Pneumatic and electrical connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-5 Auxiliary Common Gas Outlet (ACGO) (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-26 ii 07/01 1006-0802-000 . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-8 Flow sensor maintenance . . . . . . . . . . . . . . . . . . . . 3-15 Accessory shelf . . . . . . . . . . . . . . . . . . .3-5 Common Gas Outlet (CGO) . . . . . . . . . . . . 4-25 How to prevent water build-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2 External manual breathing circuit setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-20 Paw sensor calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12 Bellows assembly tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16 Installation notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-8 Power cord .8-11 Environmental requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-16 Pneumatic problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-15 Electrical problems (power failure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2 Main manifold . . . . . . . . . . . . 8-9 Electrical fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-5 Pneumatic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-8 Test tools and system parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3 Breathing system problems (no alarm) . 6-7 Bellows . . . . . . . . . . . . .8-10 Physical specifications . . . . . . . . 8-2 System specifications . . 7-2 Protocol description (Datex-Ohmeda Com 1) . . . . . . . . . . . . . . . .5-18 6 Illustrated Parts Breathing system parts . . . . . . . . . . . . . . . . . . . . . . . . 6-9 7 External Communications External communications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-5 Flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-9 Electrical block diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-12 1006-0802-000 07/01 iii . . . . . . . . . 5-2 Alphabetical list of alarm messages . . . . . . . . . . . . . . . . . . 7-3 DEVICE COMMANDS Sent By External Device . . . . . . . . . . . . . . . 7-3 Serial Communication Parameters . . . . . . . . . . . . . . . . . . . . . . 8-9 Electro-magnetic compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-6 Electrical power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Table of Contents 5 Alarms and Troubleshooting About alarms . . . . . . . . . . . . . . . . . . . 8-2 Pneumatic circuit diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . etc. . . . . . . . . . . . . . . . . . . . 8-9 Battery information . . . . . . . . . . 6-4 Exhalation valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4 8 Specifications and Theory of Operation System pneumatic circuits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6 Breathing Circuit Modules . . . . . . . . . . . . . . . . . . .) . . . . . . . . . . . . . . . . . . . . . . 8-23 O2 flowmeter (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-14 Breathing system schematic . . . . . . . . . . . . . . . 8-23 Continuous Suction Regulator . . . . . 8-21 Monitoring accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-16 Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-16 General . . . . .Aestiva/5 7100 Breathing system specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-23 Venturi Suction Regulator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-13 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-19 Ventilator accuracy data . . . . . . . . . . . . . . . . . . . . . . . . 8-21 Delivery accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-22 Suction regulators (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-15 Ventilator Theory . . . . . . . . . . . . . . . . 8-13 Gas scavenging . . . . . . . . . . . . . . . . . 8-17 Ventilation operating specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-23 iv 07/01 1006-0802-000 . . . . . . . . . . 8-21 Heliox Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2 Symbols used in the manual or on the equipment . . . 1-7 1006-0802-000 07/01 1-1 . . . . . . . . . . . . . . . . . . . . . . 1 Introduction In this section How to use this manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3 Maintenance summary and schedule . . . . . . . . . . . . . . . . . . . . . . 1-6 Operator maintenance. . . . . . . . . . . . . . . . . . . . . . . . 1-6 Datex-Ohmeda approved service . . 1-2 07/01 1006-0802-000 . References made in this manual to Pressure Mode. w WARNING If an alarm occurs. • Identify and replace worn or damaged parts. • Set up the system. tells you what causes each alarm and what you can do about it. safeguard the patient first.Aestiva/5 7100 How to use this manual This is Part 2 of the Aestiva/5 7100 Operation manual. and Pressure Waveform Display are only applicable to systems equipped with these functions. It tells you how to: • Remove and clean parts. Volume Monitoring. • Calibrate the O2 sensor. before troubleshooting or doing repair procedures. • The last section. Use this manual (Part 2) with Part 1 which includes the operating instructions and preoperative checkout. The second half of this chapter (1/Introduction) is a maintenance schedule. Volume Compensation. Note: The Aestiva/5 7100 can be equipped with several optional ventilation functions. O 2 Monitoring. Troubleshooting. p “ON” only for part of the equipment. ≈ Input. N N This way up. WARNINGS tell about a condition that can cause injury to the operator or the patient. π Dangerous Voltage. These symbols include: ø On (power). Other symbols replace words on the equipment or in Datex-Ohmeda manuals. å Alarm silence button. O Off (power). the equipment. A Note provides additional information to clarify a statement in text. µ Type BF equipment. q Standby or preparatory state for part of H Type CF equipment. r Frame or chassis ground. 1 Introduction Symbols used in the manual or on the equipment wWARNINGS and wCAUTIONS tell you about dangerous conditions that can occur if you do not follow all instructions in this manual. 1006-0802-000 07/01 1-3 . t Variability. R Bag position/ manual ventilation. REF Stock Number. An Important statement is similar to a Note. Output. wW Attention. CAUTIONS tell about a condition that can cause damage to the equipment. refer to product instructions. w Caution. but provides a comment of greater emphasis. Read top of float. ISO 7000-0434. œ “OFF” only for part of the equipment. IEC 601-1. Í Not autoclavable. y Earth ground. o Standby. x Protective earth ground. † Direct current. SN Serial Number. ∏ Alternating current. No one device or manual uses all of the symbols. m Type B equipment. Y Equipotential. Read and follow all warnings and cautions. positive polarity. U Close drain. q Inspiratory flow. lighting. r Mechanical ventilation. Minus.Aestiva/5 7100 T Variability in steps. O2+ O2 Flush button. Suction bottle outlet. Vacuum inlet. Isolation transformer. Q t O2 sensor connection. u Open drain (remove liquid). Alarm silence touch key. . Low pressure leak test. Z Unlock. 1-4 07/01 1006-0802-000 . Menu touch key. Expiratory flow. Volume alarms On/Off touch key. z Lock. + Plus. Cylinder. negative polarity. illumination. N Movement in one direction. 134°C C Autoclavable. Linkage system. Risk of Explosion. P Lamp. ˆ Movement in two directions. 179 (Tec 6).23.179 the European Council Directive XXXX (93/42/EEC) for Medical Devices when they are used as specified in their Operation and Maintenance Manuals. The primary regulator is set to pressure less than 414 kPa AB43045 < 414 kPa (60 psi). 1 Introduction End case touch key The primary regulator is set to < 345 kPa pressure less than 345 kPa (50 psi). The xxxx is the certification number of the Notified Body used by Datex- Ohmeda’s Quality Systems. Alarm silence touch key OT.tif AB. 1006-0802-000 07/01 1-5 . Absorber off (CO2 Bypass active). Absorber on.23.01. European Union Representative. Systems with this mark agree with AB.205. Two weeks • Drain the vaporizers and discard the agent. • Drain the breathing circuit module. Annually • Replace the external o-rings on the vaporizer ports. (This is included in the Preoperative Test Procedure.) • Replace the autoclavable flow sensors (metal). 1-6 07/01 1006-0802-000 . Replace or repair as and setup necessary. (Under typical use the sensor meets specifications for 1 year.) • Drain the overflow trap on the optional suction regula- tor. • Put Krytox (or a lubricant approved for use with 100% O2) on all tee handle threads. • Drain the breathing system. Weekly • Flow sensor calibration. Operator maintenance. • 21% O2 calibration (circuit O2 sensor).) • Replace the receiver filter (active gas scavenging only). This is not necessary for Tec 6 vaporizers.) • Replace the disposable flow sensors (plastic). Monthly • 100% O2 calibration (cIrcuit O2 sensor). • Replace the circuit O2 sensor.Aestiva/5 7100 Maintenance summary and schedule These schedules are the minimum frequency based on typical usage of 2000 hours per year. (Under typical use the sensor meets specifications for 1 year. (Under typical use the sensor meets specifications for 3 months. • Replace the absorbent in the absorber. Minimum Frequency Maintenance Daily • Clean the external surfaces. As necessary • Install new cylinder gaskets on cylinder yokes. During cleaning • Inspect the parts for damage. You should service the equipment more frequently if you use it more than the typical yearly usage. 1006-0802-000 07/01 1-7 . 1 Introduction Datex-Ohmeda approved service Minimum Frequency Maintenance 6 months Have a trained service person do the service tests and scheduled service maintenance. Aestiva/5 7100 1-8 07/01 1006-0802-000 . . . . . . . . . . 2-15 1006-0802-000 07/01 2-1 . . . . . . . . . . . . . . . . . . . . . . . . . • Wear gloves and safety glasses. . . . . . . . . . . . . . . . . . . . glass cleaners. . . . or petroleum based solvents. . or other harsh cleaning agents. . . 2 Cleaning and Sterilization w WARNING Obey applicable safety precautions: • Read the material data sheet for each cleaning agent. . . . . . . . . . 2-13 How to clean and sterilize the optional CO2 Bypass assembly . . . . . In this section Summary . . . . . • Do not breathe the fumes. . • Read the operation and maintenance manual for all sterilization equipment. . . . . • Keep all liquids away from electronic parts. . . . 2-10 Disassemble the scavenging path . • Do not use organic. . . Swelling or faster aging can occur. . . . . . . . . . • Do not permit liquid to go into the equipment housings. w CAUTION To help prevent damage: • Refer to the manufacturer’s data if you have questions about a cleaning agent. . . . . . . . . . . . . . . . 2-12 How to clean and disinfect the flow sensors . . . . . . . . . . . . . . . . . . . halogenated. . . . . . . . . . . . . 2-2 Clean and sterilize . . . . . . . . • Do not soak synthetic rubber parts for more than 15 minutes. . . . . . . . . . . . acetone. . . . . . . . . . . . . . . . anesthetic agents. . . . . . . . . . . . . . . . . . . . . . 2-4 Disassemble the patient path . . . . A damaged O2 sensor can leak and cause burns (contains potassium hydroxide). . . . . . . . 2-7 Canister disassembly . . . . . . • Do not use abrasive cleaning agents (such as steel wool. . . . . . . . • Only autoclave parts that are marked 134 °C. . . . . . . . . . . . silver polish or cleaner). . . . . . . . refer to cleaning/disinfection * procedure.23. Autoclave (134 °C) or AB.23. autoclave (134 °C) Wipe with a damp cloth.003 wash (mild detergent pH <10.5) If plastic.079 * Figure 2-1 • These parts return exhaled gas to the patient 2-2 07/01 1006-0802-000 . They may require more frequent cleaning/sterilization than parts in Figure 2-2. If metal.Aestiva/5 7100 Summary Patient path The shaded parts in Figure 2-1 send exhaled gas back to the patient. Refer to your hospital’s infection control policy. AB. 003 Autoclave (134 °C) or wash (mild detergent pH <10.23.5) AB.23. 2 Cleaning and Sterilization Scavenging path AB.080 Figure 2-2 • Shaded parts do not send gas back to the patient 1006-0802-000 07/01 2-3 . Aestiva/5 7100 Clean and sterilize The Disassembly part of this section tells you how to remove parts for cleaning. To wash (by hand or machine) Patient path AB. User maintenance tells you how to disassemble parts and clean inside them if necessary.083 Use a mild detergent (pH <10.5).085 Scavenging path AB. rinse and dry completely. Then. and flow sensors can be washed. 2-4 07/01 1006-0802-000 .23. All parts except the O2 sensor.23. 083 Autoclave at 134 °C. Inspect the parts for deterioration.23.23.086 Scavenging path AB. 1006-0802-000 07/01 2-5 . 2 Cleaning and Sterilization Autoclave Patient path Metal only AB. The user Maintenance section tells you how to do this.087 Upside down AB.23. wipe it with a damp cloth. These materials can go into the patient’s lungs and airways and cause irritation or injury. The Preoperative tests in part 1 (the first half of this manual). the convolutions can stick together. 2-6 07/01 1006-0802-000 . • Disassemble the bellows assembly before you wash it. Do not get the connectors wet. tell you how to test the system for correct operation.088 * * Hang the bellows upside down (extended) to dry. calcium carbonate. corn starch or equivalent materials to prevent tackiness. • To clean the circuit O2 sensor. w CAUTION Do not put the circuit O2 sensor or flow sensor connector in liquid. • Assemble the bellows assembly before you autoclave. • To clean/disinfect metal or plastic flow sensors. w Do not autoclave the circuit O2 sensor or the plastic flow sensors. Inspect all parts for deterioration. w Do not clean the interior surfaces of the flow sensors.23. use the flow sensor cleaning procedures.Aestiva/5 7100 Special requirements AB. Hang the bellows upside down to dry. Autoclave the bellows assembly upside down. Do not put the sensor in liquid. Assemble The Set-up section tells you how to assemble the breathing system. w WARNING Do not use talc. it will take a very long time to dry. Replace them if necessary. If not. zinc stearate. If not. Use a damp cloth on external surfaces only. 011 Step 4 Disconnect the cable and remove the O2 sensor.010 Step 3 Remove the flow sensor module. AA. AA.009 Step 2 Push up on the latch located under the flow sensor module.96p.96p. AA.96p013. 2 Cleaning and Sterilization Disassemble the patient path Step 1 Open the access panel. You will feel some resistance. Continue to pull. AA. 014 1006-0802-000 07/01 2-7 .96p. AA. continue to pull. AA.96p016 Step 6 Using the thumbscrew.96p020 2-8 07/01 1006-0802-000 .96p017 Step 7 Push the metal button to the first stop. Remove the breathing circuit module. Pull out the bag arm.96p018 Step 8 Push the metal button to the second stop and open the top panel. AA. AA.Aestiva/5 7100 Step 5 Completely loosen the thumbscrew on the breathing circuit module. You will feel some resistance. AA.96p024 Autoclave assemblies marked 134 °C. 2 Cleaning and Sterilization Step 9 Push the release button and remove the bellows assembly (pull up). 1006-0802-000 07/01 2-9 . Refer to the Section “Clean and sterilize” for complete instructions. To assemble the circuit refer to the Setup Section. 23 Step 10 Remove the main manifold.96p021. AA. AA. pull the release handle forward. wZ AA.96p028 Step 3 Pull down and out on the canisters. AA. Then turn the handle.96p029 2-10 07/01 1006-0802-000 .96p026 Step 2 Turn the release handle clockwise.Aestiva/5 7100 Canister disassembly w CAUTION To prevent damage. Step 1 Pull the release handle forward. AA. AA. To assemble. refer to the Setup Section. 2 Cleaning and Sterilization Step 4 Push in the buttons on each side of the top dish.96p054 Continue with the next section or go to Cleaning and Sterilization. Remove the top dish and drain dish.96p053 Step 5 Lift up the drain dish. 1006-0802-000 07/01 2-11 . Pull the reservoir up and to the rear.090 2-12 07/01 1006-0802-000 . Twist and pull down to remove the tube. AB.23. not the patient.23. Push in the latches and pull out the exhalation valve.Aestiva/5 7100 Disassemble the scavenging path These shaded parts send exhaled gas to the disposal system.089 Scavenging reservoir Loosen the knob and remove the rear cover. Exhalation valve Complete the basic disassembly procedure. AB. (Refer to “Disassemble the patient path” in this chapter. 1006-0802-000 07/01 2-13 .) 2.23. Remove the flow sensors. with activator vial (REF REORDER # 2245). Remove the flow sensor module from the absorber. c. b. a. b. c. CIDEX Plus). Remove the flow sensors from the module.151 a. • One liter of this solution cleans four (4) flow sensors.g. • CIDEX must be 14 day mixture. w Do not use cleaning solvents that are not approved for use with polycarbonates (e. AB. 2 Cleaning and Sterilization How to clean and disinfect the flow sensors wCAUTION Do not autoclave plastic flow sensors. CIDEX sterilization Both Datex-Ohmeda and the manufacturer of CIDEX (Johnson & Johnson) have tested this procedure. Pull off the cover. Push in the latch. Procedure 1. w Do not use high pressure gas or brushes to clean the flow sensors. Submerge the flow sensor and tubes in activated CIDEX solution. Rinse as indicated in CIDEX instructions. AB. Do steps 5 and 6 again to remove all CIDEX. Keep the solution in the tubes for the sterilization period. tubes. Again.Aestiva/5 7100 3.48. 8. Use a dry syringe or connect vacuum or pressure to remove all liquid from the sensor (sensor. AB.48.047 6.068 4.48.048 precautions: • Maximum vacuum 30 in Hg • Maximum flow 10 l/min • Maximum pressure 345 kPa (50 psi) 2-14 07/01 1006-0802-000 . Submerge the flow sensor and tubes in distilled water. do not get the connector wet. 7. 5. and connector): w Dry for > 1 min with these AB. COMPLETELY dry the flow sensor and the tubes before you use the sensor. Keep the connector dry. Pull the canister release handle forward.) a.96p053 Step 3 Pull down on the CO2 Bypass assembly (a) to remove it and lift the drain dish (b) up. b. clean and lubricate the pins in the drain dish and sockets in the base. c. AA. 2 Cleaning and Sterilization How to clean and sterilize the optional CO2 Bypass assembly These instructions assume that the system has the optional CO 2 Bypass feature installed. Step 2 Release the CO2 Bypass assembly by pushing both buttons (one on each side of the assembly). If they do not. (Also described in “Canister disassembly” in this section. w CAUTIONTo prevent damage. Step 1 Remove the absorber canisters. pull the canister release handle forward before turning the handle. a b 1006-0802-000 07/01 2-15 . Note: The canisters should drop under their own weight when released. Turn the release handle clockwise. Lift out the canisters. reattach the hose to the bottom part of the assembly. Note: Optionally. Rinse the assembly by immers- ing it in clean water and agitat- ing it again with the plate actuated and released. d. When reassembling. Cool about 40 minutes at room temperature. Agitate the assembly in the solu- tion while repeatedly actuating and releasing the plate. c. Pull down on the assembly to make sure it is locked into place.Aestiva/5 7100 Step 4 Clean and sterilize the bypass assembly: a. c. e. d. Insert the bypass assembly. Push back and adjust to locate the access holes for the top b ports. Autoclave the assembly at 134 a °C. Step 5 Reinstall the bypass assembly: d a. 2-16 07/01 1006-0802-000 . Immerse the assembly into a so- lution of mild detergent and wa- ter. both buttons should snap into place. Reinstall the drain dish. e. Push the buttons on each side of a the assembly and push the as- sembly up. Secure the hose with the clamp and pinch the clamp one notch past finger-tight with pliers. the Bypass assembly can be divided into two smaller sections for easier cleaning by releasing the top clamp (a) and pulling the hose from the assembly. b. c b. Operation and Checkout. Refer to the ‘Appendix . complete the Preoperative Test procedure.Preoperative Tests’ section of the Operation Manual Part 1. Ensure seals and rim are free of soda lime dust. b. Step 8 Before you use the system. 2 Cleaning and Sterilization Step 6 Reinstall the canisters: a. Turn the canister release handle counterclockwise to its locked position while gently applying pressure on the canisters to hold them against the canister guides. System Controls. Step 7 Test the Bag circuit for leaks according to the Breathing System tests in the Appendix section of Part 1 of the Operation manual. • Do the test in both the Absorber mode and the Bypass mode. 1006-0802-000 07/01 2-17 . See “CO2 Bypass mode operation” in section 3 of Part 1 of the Operation manual. c. Place the canisters in the drain dish. • Ensure the alarm message “No CO2 Absorption” activates when in Bypass mode. Aestiva/5 7100 2-18 07/01 1006-0802-000 . . . 3-12 How to attach equipment to the top shelves . . . Higher levels can cause nuisance alarms that may stop mechanical ventilation. . . . . use only Datex-Ohmeda cables. . . . . . . . w To protect the patient when electrosurgical equipment is used: • Monitor the correct operation of all life support and monitoring equipment. . w This system operates correctly at the electrical interference levels of IEC 601-1-2. hoses and tubing. . w WARNINGS Always make sure that the pipeline supply hoses and the breathing circuit components are not toxic and will not: • Cause an allergic reaction in the patient. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . • Keep backup manual ventilation available in case the electrosurgical equipment prevents safe use of the ventilator. . . . . . . . . . . 3-9 How to install gas cylinders (high pressure leak test) . . . . . 3-15 Installation notes . . . . . . . . . . . . w To prevent incorrect values or equipment malfunction. . . . . . . 3-7 Pneumatic and electrical connections . . . . . . . . . . . . . . 3 Setup and Connections Important Datex-Ohmeda strongly recommends that you use O2 monitoring with this equipment. . . 3-2 External manual breathing circuit setup . • Do not put the electrosurgical leads on any part of the anesthesia system. . 3-5 Canister Setup . . . . . . . . . . . . . . . . . . . . . . . Important European Standard EN 740 requires CO2 monitoring during ventilation. 3-17 1006-0802-000 07/01 3-1 . . . . . . . . . . . . . Important European Standard EN 740 requires anesthetic agent monitoring when anesthetic vaporiz- ers are in use. . . . . . . . . Refer to local standards for mandatory monitoring. . . . . . . . . • React with the anesthetic gases or agent to produce dangerous by-products. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . w To help prevent false alarms from devices with high-intensity electrical fields: • Keep the electrosurgical leads away from the breathing system and the flow and oxygen sensors. . . . . . • Do not use conductive masks or hoses. . In this section Breathing system setup . . . . . . . . . . . . . . • Make sure the release button aligns correctly.Aestiva/5 7100 Breathing system setup To assemble the patient circuit. AA. • Install a rebreathing bag on the bag arm.96p012 Step 4 While pressing down firmly on the control panel.96p019 • Ensure the control panel is latched into position. install the bag arm. AA. 3-2 07/01 1006-0802-000 . gently close the control panel.96p022 Step 3 While pressing and holding the metal button. start at step 5. AA.96p024 Step 2 Install the bellows assembly. Step 1 Install the main manifold. • Cycle the Bag/Vent switch to ensure locking. AA. • Completely tighten the AA. 3 Setup and Connections Step 5 Identify the breathing circuit Circle module (Circle or Mapleson. 014 sensor.96p017 resistance.067 Look at the tabs on the rear or the label on the front.) AB.96p016 Step 8 Install the O2 sensor or O2 port plug. Step 7 Completely tighten the thumbscrew AA. 1006-0802-000 07/01 3-3 .23. Mapleson Step 6 Install the breathing circuit module. Ensure that the o-ring is not showing. You will feel some AA.96p013. Continue to push. • Connect the cable (O2 sensor only). 96p011 position. AB. • You will hear a “click” when it locks into AA.181 3-4 07/01 1006-0802-000 .96p009 Step 11 Connect the patient circuit.182 Circle: Insp Exp Mapleson D / Bain Module: To-fro Insp Exp AB. “Flow sensor maintenance.23.Aestiva/5 7100 Step 9 Install the flow sensor module.23.” Step 10 Close the front panel. AA. • Refer to section 4/User Maintenance. Common Gas Outlet 1. The sample flow to the O2 sensor is dependent on the pressure in the external circuit. The sample flow reduces the fresh gas flowrate to the external breathing circuit. Auxiliary Common O2 monitoring of fresh gas is available automatically when the ACGO is selected. APL valve and bag arm are not part of the external circuit. • Volume and pressure monitoring are not available. 3. install the cell into the adapter. 1006-0802-000 07/01 3-5 . 3 Setup and Connections External manual breathing circuit setup When you operate a breathing apparatus with fresh gas from the CGO or ACGO: • Mechanical ventilation is not available • The pressure gauge. A sample of the fresh gas is diverted to the O2 cell in its standard location in the breathing system. Oxygen Monitoring of the fresh gas is recommended and available for use with external manual breathing circuits. 2. and the cell with adapter into the Tee piece. Bag/Vent switch. Remove the O2 cell from the breathing circuit module. Gas Outlet (ACGO) (Optional) Fresh gas oxygen concentration is shown on the ventilation display. Note: The fresh gas oxygen concentration may not reflect fiO 2 in rebreathing circuits such as the Mapleson series. Set alarm limits appropriately. • Use of etCO2 monitoring is recommended. Attach a Tee piece to the CGO and make the fresh gas connection from the Tee piece (CGO) to the external circuit. Note that fresh gas oxygen concentration may not reflect fiO 2 in rebreathing circuits such as the Mapleson series. Set alarm limits appropriately. Fresh gas oxygen concentration is shown on the ventilation display. Scavenging from the If the external manual breathing circuit is to be used with N2O or volatile anesthetics. or bag. It provides a 30 mm male swivel connection into the air brake of active AGSS units under the Aestiva breathing system. Occlude the patient circuit of the Aestiva breathing system on the leak test port on the breathing system handle. • The optional auxiliary inlet is a convenience inlet to the air brake of active AGSS units. will fill slowly with the fresh gas sample flow and then spill to the AGSS. Set the Bag to Vent switch to the ventilator mode. 1. Its effectiveness is limited by the extract flow of the particular active AGSS device.Aestiva/5 7100 Scavenging the If the external manual breathing circuit is to be used with N2O or volatile anesthetics. Check for clinically correct settings. 3. breathing circuit An optional auxiliary inlet is available for active AGSS units. a separate exhaust hose is necessary from the external manual circuit to the disposal point. Alternatively set the Bag to Vent switch to the bag mode and set the APL to MIN and attach a bag. For passive AGSS. ACGO sample flow the sample flow should be scavenged. The bellows. external manual the exhaust should be scavenged. There is no reservoir to capture exhaust flows higher than the extract flow. 3-6 07/01 1006-0802-000 . 2. During non-use. AB. Step 1 Pull the release handle forward.23. absorbent can go back to the original appearance. replace the absorbent. Inspect absorbent color at the end of a case. it will give off CO (carbon monoxide) when exposed to anesthetic agents.35 kg of loose absorbent or 1.061 1006-0802-000 07/01 3-7 . Datex-Ohmeda recommends calcium. 3 Setup and Connections Canister setup Each canister holds 1.23. sodium or barium hydroxide based absorbent. For safety. If the absorbent completely dries out. w WARNING Change absorbent often to prevent the build up of non-metabolic gases when the system is not in use. Refer to the absorbent labeling for more information about color changes.13 kg of commercially packaged (pre-pack) absorbent. AB. Always perform a breathing system leak test in the Bag Mode after opening the absorber.061 Step 2 Turn the release handle clockwise. Step 5 Push back on the canisters and turn the release handle counter- clockwise. AA.96p029 Push down on the drain dish if necessary. AA.96p026 3-8 07/01 1006-0802-000 .Aestiva/5 7100 Step 3 Push in the buttons and install the top dish and drain dish.96p028 Step 6 Push in the release handle. AA. Step 4 Fill the canisters with absorbent and install them.96p054 Pull down to make sure the top dish is locked in place. AA. 96. oil.116 Tec 6 power Arrow shows the Tec 6 power connection (not on all AA. AA. An optional isolation transformer kit is available to reduce total leakage current. or other contaminants.095 models).96. The Aestiva gas supplies also supply these devices through internal connections: • The venturi suction regulator (optional) • The external O2 flowmeter (optional) • Ventilator drive gas • Pneumatic outlet (optional) 1006-0802-000 07/01 3-9 . Other types of gas supplies may contain water. Regularly test the leakage current. 3 Setup and Connections Pneumatic and electrical connections wWARNING Equipment connected to the electrical outlets can increase the leakage current.116 Outlets Labels show outlet voltage ratings and circuit breaker amp ratings. Mains inlet Arrow shows the mains inlet.96. AA. wCAUTION Use only medical grade gas supplies. Aestiva/5 7100 Pipeline Inlets Label identifies the gas. AA.96.96.116 O2 Pneumatic outlet AA.116 Scavenging Adapters may be necessary. Refer section 6/Illustrated Parts.137 3-10 07/01 1006-0802-000 .96. AA. 172 vacuum supply.116 1006-0802-000 07/01 3-11 . Note: Venturi regulators use the system air or O2 supply.227 vacuum (non-venturi) 2. AA.2188 Serial port Refer to the communications appendix for command and data formats. 3 Setup and Connections Suction regulator Optional item. Standard regulators must be connected to an external AA.96.96. Filter 1 2 3. 1.96. Collection bottle 4 4.96. AA. Patient connection 3 External O2 flowmeter Optional item. External AA. Open the cylinder yoke.124 4. Fully loosen the tee handle. Find the cylinder wrench.96.122 3.96.123 2. AA.96.126 3-12 07/01 1006-0802-000 . Close the cylinder valve on the cylinder to be replaced. AA.125 5. AA.Aestiva/5 7100 How to install gas cylinders (high pressure leak test) Cylinder yokes 1. AA. AA. Remove the used cylinder and the used gasket.96.96. Align the cylinder post with the index pins. w CAUTION No gasket or more than one gasket can cause a leak. 11. 7.Close the yoke gate and tighten the tee handle. 10. This removes dirt from the cylinder outlet. Record the cylinder pressure. Disconnect pipeline supplies. e. w WARNING Do not leave gas cylinder valves open if the pipeline supply is in use. If the leak continues. b. do not use the system. Close the cylinder. Cylinder supplies could be depleted. Disconnect all equipment from the pneumatic outlet and turn off the auxiliary flowmeter. 9. For all other cylinder types. Set the system switch to Standby.Do a high pressure leak test: a. g. 12. To repair a leak: • Install a new cylinder gasket and tighten the tee handle. Open the cylinder. 8. Quickly open and close the cylinder valve. • Do this step again. If the air or oxygen cylinder pressure decreases more than 5000 kPa (725 psi) in one minute.Install a cylinder plug and gasket in all empty cylinder yokes. Install a new gasket. Point the cylinder outlet away from all items that can be damaged by a release of high pressure gas. 13. there is a significant leak. 3 Setup and Connections 6. c. a pressure decrease of 690 kPa (100 psi) in one minute represents a significant leak. d. f. h. 1006-0802-000 07/01 3-13 . leaving an insufficient reserve supply in case of pipeline failure. Remove the cap from the cylinder valve on the new cylinder. Quickly open and close the cylinder valve. leaving an insufficient reserve supply in case of pipeline failure. Remove the cap from the cylinder valve on the new cylinder. Loosen the adapter and remove the cylinder. c. Close the cylinder valve on the cylinder to be replaced. To repair a leak: • Install a new cylinder gasket and tighten the tee handle.96. w WARNING Do not leave gas cylinder valves open if the pipeline supply is in use. Do a high pressure leak test: a. For all other cylinder types. do not use the system. a pressure decrease of 690 kPa (100 psi) in one minute represents a significant leak. e. Record the cylinder pressure. Open the cylinder.127 2. AA. b.96. Disconnect pipeline supplies.129 3. 3-14 07/01 1006-0802-000 . Cylinder supplies could be depleted. 6. there is a significant leak. Point the cylinder outlet away from all items that can be damaged by a release of high pressure gas. g. If the cylinder pressure decreases more than 5000 kPa (725 psi) for O 2 and air in one minute. 4. Disconnect all equipment from the pneumatic outlet and turn OFF the auxiliary flowmeter. If the leak continues. AA. This removes dirt from the cylinder outlet. f. Set the system switch to Standby. 5. • Do this step again.Aestiva/5 7100 DIN connections 1. h. Close the cylinder. d. 3 Setup and Connections How to attach equipment to accessory shelf or machine top w Top accessory shelves have a weight limit of 46 kg (100 lbs). Fully tighten the straps.50. D) D A B C A B C 3. Accessory shelf 1. w WARNING If you do not fully tighten the strap. Make sure that the straps hold the equipment in position. C) B A C A B C AA. B. 4. Install the clips in the two slots nearest to the equipment. Loop two times (A. equipment can fall off the shelf. AA71. A folding side shelf has a weight limit of 23 kg (50 lbs). Install the straps. B. Large equipment.015 Small equipment.145 2. Loop one time (A. 1006-0802-000 07/01 3-15 . C. w Systems without accessory shelves have a weight limit of 46 kg (100 lbs). D C D C or: AA. 4.135.72. equipment can fall off the shelf. AB.96.012 2. or: AA. Locate the clips or slots. Make sure that the straps hold the equipment in position. w WARNING If you do not fully tighten the strap. Fully tighten the straps. 3-16 07/01 1006-0802-000 .72. AB. Install the straps.96.013 B A B A 3.Aestiva/5 7100 Top of machine 1.130. 1006-0802-000 07/01 3-17 . the Datex-Ohmeda representative will check the following settings and change them if necessary. • Language. wW WARNING These settings can only be changed by qualified personnel. 3 Setup and Connections Installation notes When the system is installed. activation and deactivation of monitoring and ventilation features. • Heliox mode availability. • Altitude. • Automatic calculation of VE alarm limits during mechanical ventilation. • Upgrade. • Ventilator drive gas. Aestiva/5 7100 3-18 07/01 1006-0802-000 . 4-15 O2 sensor calibration . . . . . . . . . . . . • All covers used on the system must be made from antistatic (conductive) materials. . . . . . . . . . . . . . . . . . . . . . . . . . such as Krytox® . . . . . . . . . . . . In this section Repair policy . 4-2 Manifold maintenance . . . . . . 4-9 Breathing circuit maintenance . . • Do not use lubricants that contain oil or grease. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-23 Flow sensor calibration . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12 Bellows assembly tests . . . . . . . . . . . . . 4-26 1. . . . 4-3 Exhalation valve maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-18 Paw sensor calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .I. . . . . . . . . . . . . . . They burn or explode in high O2 concentrations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Used equipment may contain blood and body fluids. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Static electricity can cause fires. . . . . . . . . . . . . . . . . . . . . 4-25 How to prevent water build-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8 Flow sensor maintenance . . . . . . & Company Inc 1006-0802-000 07/01 4-1 . . . . . . . . . . . . . . . ® Krytox is a registered trademark of Dupont de Nemours E. . . . . . . . . . . . . . . . . . . 4-6 Receiver maintenance (active gas scavenging only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . w WARNING Obey infection control and safety procedures. . . . . . . . 4-10 Bellows assembly maintenance . . . . . . . . . . . . . . . 4 User Maintenance w WARNING TO HELP PREVENT FIRES: 1 • Use lubricants approved for anesthesia or O2 equipment. . . . . . . . . . . . . . . . . . . . . . After repair. 4-2 07/01 1006-0802-000 . If this cannot be done. replacement and maintenance of those parts listed in this manual may be undertaken by a competent. To ensure full reliability. Replace damaged parts with components manufactured or sold by Datex- Ohmeda. Then test the unit to ascertain that it complies with the manufacturer’s published specifications. w CAUTION No repair should ever be attempted by anyone not having experience in the repair of devices of this nature. repairs will be made at Datex-Ohmeda’s current list price for the replacement part(s) plus a reasonable labor charge. have all repairs and service done by an authorized Datex-Ohmeda service representative. In all cases. trained individual having experience in the repair of devices of this nature. Make all necessary repairs or have the equipment serviced by an authorized Datex-Ohmeda service representative. other than where Datex-Ohmeda’s warranty is applicable. test the equipment to ensure that it is functioning properly.Aestiva/5 7100 Repair policy Do not use malfunctioning equipment. Contact your local Datex-Ohmeda Field Service Representative for service assistance. in accordance with the manufacturer’s published specifications. 070 • Replace the seal if damaged or worn. Step 3 (Seals) Loosen the thumbscrew and remove the plate. AB.23.23. or diaphragm is damaged.) AB.069 • The diaphragm looks worn. Replace it if: • The disk. 1006-0802-000 07/01 4-3 . 4 User Maintenance Manifold maintenance Step 1 Complete the Basic disassembly to remove the main manifold. AB.094 Step 2 (APL valve) Remove the diaphragm. (Refer to the 2/Cleaning and Sterilization section.23. cage. • The diaphragm fits over the rim. Aestiva/5 7100 Step 4 (Seals con.073 Step 5 (Seals con. and bottom parts). AB.23.) Replace the parts if they are damaged or worn.23. AB.) Open the manifold (top middle.) Loosen the thumbscrews.23.074 Step 6 (Seals con. Side AB.071 Top Bottom 4-4 07/01 1006-0802-000 . Step 8 Put the parts together: • Make sure all gaskets are fully seated.23. Step 9 Before you use the system.23. Refer to the ‘Appendix . complete the Preoperative Test procedure. 4 User Maintenance Step 7 (Bag/Vent) AB. 1006-0802-000 07/01 4-5 .Preoperative Tests’ section of the Operation Manual Part 1. System Controls.075 • Fully tighten the OK OK thumbscrews.077 Replace the parts if they are damaged or worn. Operation and Checkout. AB. 23. These conditions indicate a possible leak in the exhalation valve: • Alarms for low tidal volume. • The ventilator circuit has a leak. AB.23. Step 1 Remove the exhalation valve.089 Step 2 Remove the top cover. (2/Cleaning and Sterilization section. but the bellows pass the bellows tests in this section.093 4-6 07/01 1006-0802-000 .) AB.Aestiva/5 7100 Exhalation valve maintenance Disassemble the manifold and examine the seals after every 12 autoclave cycles. Refer to the ‘Appendix . System Controls.23. 4 User Maintenance Step 3 Unscrew the lock ring (one quarter turn) and remove the cap. 1 AB. align it with the base and push down. 1006-0802-000 07/01 4-7 . • To install the top. Operation and Checkout.072 Step 4 Put the parts together : • Align the dimples (1) in the cap with the valve seat.Preoperative Tests’ section of the Operation Manual Part 1. AB.021 Step 5 Before you use the system. complete the Preoperative Test procedure.23. In extreme cases this can open the high pressure relief valve and vent to room air.23. System Controls. Note: Passive gas scavenging does not require maintenance. Higher levels can cause nuisance alarms that may stop mechanical ventilation. Step 1 b Remove the receiver: • Loosen the thumbscrew. complete the Preoperative Test procedure. w WARNING This system operates correctly at the electrical interference levels of IEC 601-1-2. c AB.091 a Step 2 Replace the filter (a) (friction fit with gasket. Refer to the ‘Appendix .23. Operation and Checkout.Preoperative Tests’ section of the Operation Manual Part 1.092 a Step 3 Reassemble by: doing above steps 1a through 1c in opposite order. 4-8 07/01 1006-0802-000 . • Always make sure that the rear cover is installed.) B. • Completely tighten the thumb screw. Step 4 Before you use the system. • Remove the rear cover • Pull out and up on the receiver.Aestiva/5 7100 Receiver maintenance (active gas scavenging only) A blocked filter decreases flow through the scavenging system. 96p. 011 • Same steps as 1 .039 • Keep the tubes straight.038 • Make sure the tab clicks to lock the cover. • Do not pinch the tubes. 010. Step7 Re-install the flow sensor module. Refer to the ‘Appendix . 4 User Maintenance Flow sensor maintenance Steps 1 . complete the Preoperative Test procedure.Preoperative Tests’ section of the Operation Manual Part 1. Step 8 Before you use the system.96p.96p. Step 6 Slide on the cover.3 Remove the flow sensor module. System Controls. AA. AA. 1006-0802-000 07/01 4-9 .037 Step 5 Remove the sensor and install the new sensor with the arrows up.009.3 of “Disassemble the patient path” in 2/Cleaning and Sterilization. AA.96p. Step 4 Push in the tab (a) and a remove the cover. Operation and Checkout. AA. Step 7 Hold in the tab.Aestiva/5 7100 Breathing circuit maintenance Disassemble the breathing circuit module to replace check valves or seals.041 Step 8 Remove the rear cover. AA.043 4-10 07/01 1006-0802-000 .96p.96p. AA. Note the rear seal is keyed to fit a specific breathing circuit module. AA.6 • Same steps as 1 . Step 1 .96p.6 of “Disassemble the patient path” in 2/Cleaning and Sterilization.042 Step 9 Push in the tabs and remove the top cover. Step 11 Do the steps in the opposite order to assemble the module.96p. complete the Preoperative Test procedure. Check movement of drain button on circle-type AA. Turn and pull to remove the check valves.96p. Operation and Checkout. AA. System Controls.045 Step 12 Before you use the system.044 breathing circuit module. Refer to the ‘Appendix . 4 User Maintenance Step 10 Remove the seals.Preoperative Tests’ section of the Operation Manual Part 1. 1006-0802-000 07/01 4-11 . 96p. AA.051 Step 3 Remove the bottom edge of the bellows from the rim. AA. AA.96p.052 4-12 07/01 1006-0802-000 .Aestiva/5 7100 Bellows assembly maintenance Step 1 Remove the bellows assembly from the breathing system.023 Step 2 Turn the housing counter- clockwise and lift.96p. AA. AA.96p. 4 User Maintenance Step 4 Remove the disk from the bellows.031 Step 5 Remove the ring from inside the top of the bellows.96p.032 Step 6 Push the latch toward the center and remove the rim. AA. This can damage the seat or diaphragm and cause injury to the patient.96p. 1006-0802-000 07/01 4-13 .96p.034 w WARNING Do not disassemble the pressure relief valve.033 Step 7 Remove the pressure-relief valve. AA. The rim is locked in position. apply a thin layer of KRYTOX lubricant to the ribs of the bellows housing.96p. c. The arrow on the seal points up. f. 4-14 07/01 1006-0802-000 . The inner ring is correctly installed inside the top of the bellows. d. Make sure the lubricant is applied smoothly and there are no lumps. Step 11 Perform the bellows assembly tests (which follow below) before reinstalling the bellows assembly on the system. If you see a dust-like powder on the bellows housing or on the bellows. Make sure that: a. b.96p. You hear a double click when you install the rim. The housing is locked in position.035 Step 9 Remove the seal.Aestiva/5 7100 Step 8 Push the latch toward the center and remove the locking tabs. AA. e. AA. You cannot lift it off. Only the bottom ring of the bellows fits over the rim.036 Step 10 Do these steps in the opposite order to assemble the bellows assembly. install it in the system. AA. It is not an alternative to a complete system checkout. 4 User Maintenance Bellows assembly tests w WARNING Objects in the breathing system can stop gas flow to the patient. This can cause injury or death: • Do not use a test plug that is small enough to fall into the breathing system. If there is a problem. Always complete the tests in the section Preoperative tests before you use the system with a patient. Step 1 Hold the bellows assembly vertical and close the ports (A and B).96p. If the bellows assembly operates correctly. Look for and replace damaged parts.125 A B 1006-0802-000 07/01 4-15 . This test makes sure that all components are correctly assembled. • Make sure that there are no test plugs or other objects caught in the breath- ing system. disassemble the bellows assembly. w WARNING The bellows assembly test does not replace the preoperative tests. Aestiva/5 7100 Step 2 Invert the bellows assembly. Step 3 Remove the plugs from the ports.96p. If it does: • The ports are not tightly B sealed.048 Step 4 Close port C.96p.123 • The bellows is incorrectly installed. Permit the bellows to fully extend.124 4-16 07/01 1006-0802-000 . • The seal inside the bellows is not correctly installed (with its groove pointed up). The bellows must not fall more than 100 ml/min.96p. • Parts are damaged. AA. A AA. C AA. Preoperative Tests’ section of the Operation Manual Part 1. 4 User Maintenance Step 5 Hold the bellows assembly vertical. If it does: AA. The bellows must not fall more than 100 mL/min. 1006-0802-000 07/01 4-17 . • Parts are damaged. complete the Preoperative Test procedure. Operation and Checkout.” install it in the system.050 • The port is not tightly sealed.96p. • The bellows or the pressure relief valve is not correctly installed. System Controls. Refer to the ‘Appendix . Step 7 Before you use the system. Step 6 If the result for all the bellows assembly tests was “passed. Step 3 Push the knob to show the next screen. AB43030 4-18 07/01 1006-0802-000 . Step 2 AA.115 Step 4 Push the knob to show the calibration steps.-. AB43069 Turn the knob to select the Calibration screen. During O2 calibration the screen replaces O2 data with .69.124 the main menu. AB. You must do the 21% O2 calibration before the 100% O2 calibration.Aestiva/5 7100 O2 sensor calibration w WARNING Do not perform calibration while unit is connected to a patient.350.43. 21% O2 sensor calibration This procedure takes three minutes or less. AB. Step 1 Push the Menu key to see AB43046.43. 43.068 Step 6 Push the knob to start calibration. • After repeated failures.23. replace the O2 sensor and re-calibrate at 21%. 4 User Maintenance Step 5 Complete the steps shown on the screen. AB. If the calibration fails: • Do the calibration again. calibrate at 21% again.081 After a successful calibration. • If it still fails. If 100% O2 sensor calibration passes. An unsuccessful calibration shows “FFailure”. AB43024 • Remove the O2 sensor from the circuit. do a 100% O2 sensor calibration (step 11). AB43025 • The screen shows Calibrating” during the “C procedure. • Make sure the cable is connected. After installing the sensor. AB. 1006-0802-000 07/01 4-19 . the screen shows “Complete” and a flashing “Reinstall Sensor”. • Do not twist or stress the cable. you can continue to 100% calibration (step 11) or exit to the normal screen (step 7). 100% O2 sensor calibration This procedure takes three minutes or less. select “Go to O2 AB. You can also press the menu key to exit to the normal screen Step 8 Select “Go to Main Menu” and push the knob. AB43027 4-20 07/01 1006-0802-000 . Go to Main Menu Step 9 Select “Exit to Normal Screen” and push the knob.087 Cal Menu” and push the knob.43.Aestiva/5 7100 Step 7 To exit. Step 11 Push the knob to show the next screen. You must complete the 21% calibration before you can select the 100% calibration. w WARNING Do not perform calibration while unit is connected to a patient. Exit to Normal Screen Step 10 If you are not doing the 100% calibration which follows. do a breathing circuit leak test before using the system. 4 User Maintenance Step 12 Complete the steps shown on the screen. Step 13 Push the knob to start calibration. Step 14 To exit. After repeated failures replace the O2 sensor and re-calibrate at 21%. • Then flow 100% O2 at 5 AA. The screen shows “C If the calibration fails. • do it again or • decrease the airway pressure..96. Continue with steps 14 through 16 to exit to a normal screen.113 L/min. select “Go to O2 Cal Menu” and push the Go to O2 Cal Menu knob. AB43028 Calibrating . ”.. AB43027 With the O2 sensor in the circuit. followed by the result: “Complete” or “FFailure”. Then try again. You can also press the menu key to exit to the normal screen 1006-0802-000 07/01 4-21 . fill the circuit with 100% O2: • Push the flush button. Go to Main Menu Step 16 Select “Exit to Normal Screen” and push the knob.Aestiva/5 7100 Step 15 Select “Go to Main Menu” and push the knob. 4-22 07/01 1006-0802-000 . Exit to Normal Screen Step 17 Do a breathing circuit leak test before using the system. You must stop mechanical ventilation before you calibrate the Paw sensor. AB43069 Turn the knob to select the Calibration screen. Perform this procedure on systems without volume monitoring.43.116 Step 4 Push the knob to show the calibration steps. 4 User Maintenance Paw sensor calibration w WARNING Do not perform calibration while unit is connected to a patient.69. Step 2 AA.43. Step 1 Push the Menu key to see AB43046.117 1006-0802-000 07/01 4-23 . Systems with volume monitoring automatically calibrate the Paw sensor when the flow sensor module is removed.43. Step 3 Push the knob to show the next screen.124 the main menu. AB.350. AB. AB. 43.119 • The screen shows Calibrating” during the “C procedure. select “Go to Cal AB. 4-24 07/01 1006-0802-000 . • After a successful calibration. Go to Main Menu Step 9 Select “Exit to Normal Screen” and push the Exit to Normal Screen knob.43.Aestiva/5 7100 Step 5 Complete the steps shown on the screen. Step 8 Select “Go to Main Menu” and push the knob.118 Menu” and push the knob.117 Step 6 Push the knob to start calibration. AB.43.” Step 7 To exit. the screen shows “Complete” and a flashing “Reconnect AB.118 patient hoses.43. You can also press the menu key to exit to the normal screen. AB. 4 User Maintenance Flow sensor calibration w WARNING Do not perform calibration while unit is connected to a patient. You will feel some resistance.96p010 Step 3 When calibration is complete. The system automatically corrects for zero offset when you unplug the flow sensor connectors with power on. AA.96p.011 module. No Insp Flow Sensor the screen shows.” Step 4 Install the flow sensor AA. You must stop mechanical ventilation before you calibrate the flow sensors. “No Insp No Exp Flow Sensor flow sensor” and “No Exp flow sensor.011 Step 2 Remove the flow sensor module. Step 5 Do a breathing circuit leak test before using the system. Continue to pull. 6p. Step 1 Push up on the latch under the flow sensor module. 1006-0802-000 07/01 4-25 . How much water is too much? A thin layer of water or a foggy look in the flow sensors is OK. • Ensure that water condensing in the breathing circuit tubes is kept lower than the flow sensors and is not allowed to drain into the flow sensors. • If Check Flow Sensors alarms occur during a VERY LONG case. replace the flow sensor module. • Drain all hoses as necessary. push the drain button before every case (listed in Pre-operative procedures). 4-26 07/01 1006-0802-000 . • Water condensation in the breathing circuit tubing can be eased using a Heat & Moisture Exchange (HME) filter at the airway connection. Solutions: • Drain the absorber each morning. • More moist. • With a circle breathing circuit. exhaled gas stays in the absorber. At lower fresh gas flows more water builds up because less gas is scavenged and: • More CO2 stays in the absorber to react and produce water. Allow the original flow sensors to dry before you use them again. Where does the water Water comes from exhaled gas and a chemical reaction between CO 2 and the come from? absorbent in the absorber.Aestiva/5 7100 How to prevent water build-up Why is water buildup a problem? Pooled water in the sensor or water in the sensing lines causes false alarms. . . 5-15 Electrical problems (power failure. . . In this section About alarms . . . . 5 Alarms and Troubleshooting w CAUTION No repair should ever be attempted by anyone not having experience in the repair of devices of this nature. . . . . . . . . . . . . . . . .) . . 5-3 Breathing system problems (no alarm) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . etc. . . . . . . . 5-2 Alphabetical list of alarm messages . . . . . . . . . . . . . 5-16 Pneumatic problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-18 1006-0802-000 07/01 5-1 . . . . . . . . . . . . . . . . . . . . . . . . . If an alarm is related to control settings. safeguard the patient first. the limits flash and a box appears around the parameter. Others prevent mechanical ventilation (Minimum shutdown). if the audible circuit leak alarm is Off. For example. 3. the screen shows “Cir- cuit leak audio Off” as a low priority alarm. the lower priority alarms cycle every two seconds.Aestiva/5 7100 About alarms w WARNING If an alarm occurs. before performing troubleshooting or doing repair procedures. Some malfunctions cause reduced function (for example no PEEP). repeat be silenced Screen shows elapsed time Medium 3 tones. These are high and low limit settings that you adjust. repeat Low Single tone Tone does not repeat ---- Alarm messages have three general causes: 1. 25 sec 120 Seconds ---- pause. Informational. 2. 5-2 07/01 1006-0802-000 . The area at the top of the display (1) shows alarm information. Priority Alarm tone Alarm silence Note High 2 bursts of 5 tones. Control settings or system conditions can change operation. Malfunctions. pause. If there are more than four alarms at the same time. Patient monitoring. 1 AB43065 Alarm priority depends on the level of risk to the patient. 120 Seconds or cannot Reverse video. High priority alarms require immediate attention. Settings menu). Ventilate manually. Apnea Alarm Low Normal condition after Monitoring resumes after first . mechanical). Contact a qualified Analog Out-of.. This table does not include operator instructions. move the Not necessary. service Range (High) representative. acknowledge) A/D Converter Min. Apnea Alarm Off Low The cardiac bypass Apnea alarms are turned off --- option is selected (Alarm when this option is selected. Contact a qualified Failure (High) Monitoring is not reliable. shutdown Ventilator malfunction. There are two special types of alarms: • Minimum monitoring alarms stop mechanical ventilation. Contact a qualified Test Out of Shutdown Monitoring is not reliable. power-up. --- Open after completely closed. Ventilate manually. shutdown Ventilator malfunction. +12V / -12V Min. Standby to On. Absorber Panel Medium (low The top panel is not Close the panel. power-up self test. • After the case to repair a problem. than 12 hours without a system switch from On to Informational.- Standby End Case. or breath (mechanical) or 2 ACGO change from On to breaths within 30 sec (non- Off. 5 Alarms and Troubleshooting Alphabetical list of alarm messages The instructions in this section tell you what you can do: • During a case to protect the patient. Message Priority Cause Action/Concerns Repair 12 Hour Test w Low System in use for more At end of case. 12V -10VA Min. • Minimum shutdown alarms stop mechanical ventilation and monitoring. service representative. Ventilate manually. 1006-0802-000 07/01 5-3 . (High) Monitoring is not reliable. Ventilator malfunction. service Range representative. Battery Low Battery is not fully Leave the system plugged in --- Charging w charged. Cal Flow Low The last flow sensor Calibrate the flow sensors.. calibrating or Replace sensor if necessary. the total backup time will be < 30 minutes. representative. Battery Failure Low Battery voltage > 8 V for System is operational. may fail on battery if mains service power is lost. but Contact a qualified High for 10 seconds. power is lost. representative. If power fails. Battery Charger Low The current in the battery System is operational. Battery Failure Low The battery voltage is too System is operational. acknowledge) auxiliary common gas For mechanical ventilation or outlet. ventilation with monitoring. replacing the sensor does not correct problem. but If the battery does Low low (< 2 V) to supply the will fail on battery if mains not charge in 24 system if power fails.Aestiva/5 7100 Message Priority Cause Action/Concerns Repair Aux Gas Medium (low The outlet selection Connect the patient circuit to . hours. but Contact a qualified Fail charging circuit is too may fail on battery if mains service high. 5-4 07/01 1006-0802-000 . may fail on battery if mains service power is lost. select the common gas outlet. contact a Leave the system plugged in qualified service to charge the battery. representative. Contact a qualified Sensors calibration failed.- Outlet On after switch is set to the the auxiliary outlet. to charge the battery. but Contact a qualified High 10 seconds. Battery Current Low Battery current > 6 amps System is operational. Look for water in the flow service sensor tubes and dry if representative if necessary. representative. power is lost. causing a large release position. Apnea alarms are off. Circuit Leak Low Control setting on the This message tells you that --- Audio Off Alarm limit menu. CPU Failure Min.) sensor tubes? Replace flow sensor Is a flow sensor tube cracked module with the or broken? spare. Canister Open Medium (low The canister release is Close the canister release. Ventilate manually. Replace sensor if necessary. A switch checks the after open. Ventilate manually. Contact a qualified (High) Monitoring is not reliable. Check the Improper check valve condition of the flow operation? sensor and its tubing. acknowledge) leak. Contact a qualified service representative if calibrating or replacing the sensor does not correct problem. shutdown Ventilator malfunction. Water build up in the flow circuit module. the audio alarm for circuit leaks was turned off. Check Flw Medium (low System has detected an Are the flow sensors correctly Inspect one way Sensors w after improper flow pattern in installed? valves (breathing acknowledge) the breathing circuit. Cardiac Bypass Low The alarm limit settings Use the alarm limits menu to --- are set for a patient on change this setting. 5 Alarms and Troubleshooting Message Priority Cause Action/Concerns Repair Calibrate O2 Low Calibration failure or O2% Does the sensor measure Calibrate the O2 Sensor > 110% 21% O2 in room air? sensor. Contact a qualified O2 Sensor connected to the cable. CPU Internal Min. cardiac bypass. service representative. service representative. Contact a qualified Error (High) Monitoring is not reliable. shutdown Ventilator malfunction. 1006-0802-000 07/01 5-5 . service representative to replace the cable. Connect Low The O2 sensor is not Connect the sensor. High O2 Medium O2% > alarm high limit Is the limit set correctly? Calibrate O2 sensor. 5-6 07/01 1006-0802-000 . Are Plimit and other controls --- The ventilator cycles to set correctly? Look for expiration. breaths after you change the ventilator settings. Ventilator malfunction. service (Medium) representative. High V TE Medium V TE is greater than high Check patient for --- alarm limit. Contact a qualified (Voltage) monitoring Monitoring is still available.Aestiva/5 7100 Message Priority Cause Action/Concerns Repair Exp Flow Sensor Low System cannot read the Operation continues with Replace the flow Fail calibration data stored in reduced accuracy. Heliox Mode Low Control setting on When heliox is used. sensor. Ventilator malfunction. alarm is suspended for 9 Adjust control settings. What is the O2 flow? Replace O2 sensor. Check patient connection. High VE Medium The minute volume is > Check patient for --- the set high limit. Is a flow sensor tube cracked Check the flow or broken? sensor condition. suspended for 9 breaths Check ventilator and alarm after you change the settings. Flow Valve Min. ventilator settings. Did you just push Flush? Does the sensor see 21% O2 in room air? High Paw High Paw is greater than Plimit. Ventilate manually. Flow Valve Min. in a row). Replace the Flow after expiratory sensor during Water build up in the flow expiratory check acknowledge) inspiration (for 6 breaths sensor tubes? valve. This alarm is spontaneous breathing. the sensor. Exp Reverse Medium (low Flow through the Look at the check valves. blockages. This spontaneous breathing. the --- is On ventilation setup menu. setting. service Failure (Medium) representative. ventilator must adjust volume calculations. Ventilate manually. Contact a qualified (DAC) Failure monitoring Monitoring is still available. connected and Is the mains indicator light circuit breakers are on? closed. Contact a qualified Ventilator Clock (High) Monitoring is not reliable. Ventilate manually. service representative if problem continues. patient to save power.65V while Manually ventilate the Make sure power is Voltage using battery power. in a row). Inspiration High Drive gas safety switch Adjust controls. Check systems for blockages. shutdown Ventilator malfunction. sensor. Look for broken ID tabs or Normally the system uses tape on the tabs. installed. Internal Min. Invalid Circuit Low The system does not Make sure the breathing Contact a qualified Module recognize the type of circuit module is correctly service breathing circuit module installed. appropriate gas supplies (O2 or air) are connected and pressurized. 5 Alarms and Troubleshooting Message Priority Cause Action/Concerns Repair Insp Flow Low The system cannot read Operation continues with Replace the flow Sensor Fail the calibration data reduced accuracy. Contact a qualified Ventilator Clock (High) Monitoring is not reliable. representative. Low Drive Medium The ventilator does not Manually ventilate the Make sure that the Gas Press detect supply pressure. 1006-0802-000 07/01 5-7 . Is a flow sensor tube cracked Check the flow or broken? sensor condition. stored in the sensor. patient. Replace the Flow after inspiratory sensor during Water build up in the flow inspiratory check acknowledge) expiration (for 6 breaths sensor tubes? valve. service Too Slow representative. Low Battery Medium Voltage is <5. Internal Min. Ventilate manually. the ID tabs to identify circuits. shutdown Ventilator malfunction. Contact a qualified Stopped w activated (high pressure). Insp Reverse Medium (low Flow through the Look at the check valves. service Too Fast representative. shutdown Ventilator malfunction. the last 20 sec. Contact a qualified Failure (High) Monitoring is not reliable. 5-8 07/01 1006-0802-000 . Contact a qualified (EEPROM) Fail access some stored Ventilation is still possible service values. --- low limit alarm setting. shutdown Ventilator malfunction. Does the sensor see 21% O2 As sensors wear out. Ventilate manually. Memory (Flash) Min. Low V TE Medium Exhaled tidal volume < Check patient condition. Memory (RAM) Min. --- low limit alarm setting. Ventilate manually. Contact a qualified Failure (High) Monitoring is not reliable. absorber. Is the O2 flow sufficient? Replace O2 sensor. shutdown Ventilator malfunction.Aestiva/5 7100 Message Priority Cause Action/Concerns Repair Low O2 High O2% < alarm low limit Is the limit set correctly? Calibrate O2 sensor. representative. Check tubing connections. service representative. for 9 breaths after the ventilator settings are changed. This alarm is suspended Check alarm settings. Contact a qualified Failure (High) Monitoring is not reliable. Memory (Video) Min. Check tubing connections. Memory Min. Low VE Medium Exhaled minute volume < Check patient condition. service representative. Ventilate manually. but service is necessary. 9 breaths after you change the ventilator settings. service representative. This alarm suspended for Check alarm settings. shutdown Ventilator malfunction. service Storage) Failure representative. Low Paw High Paw does not rise at least Are circuit connections Ok? Look for circuit 4 cm above Pmin during Look at the Paw gauge on the disconnection. Contact a qualified (Redundant (High) Monitoring is still available. in room air? the measured % O2 decreases. setting. Ventilate manually. Memory Low The system cannot Default settings are used. Other alarms may also If the alarm clears. 1006-0802-000 07/01 5-9 . 5 Alarms and Troubleshooting Message Priority Cause Action/Concerns Repair Monitoring Only Medium A severe malfunction Ventilate manually. use. qualified service representative. module is on all the way. contact a (≥30 sec). module. mechanical ventilation. No Insp Flow Medium (low Electrical signals show Connect the flow sensors. service ventilation. for tabs on the back of the module. Connect a pipeline supply or install an O2 cylinder. O2 Sensor Low O2 sensor not installed Install a breathing circuit --- Out Circ in breathing circuit module and an O2 sensor. No Exp Flow Medium (low Electrical signals show Connect the flow sensors. --- Sensor after the flow sensor is not Make sure the flow sensor acknowledge) connected . restart occur. --- be silenced) Ventilate manually if necessary. User setting. --- Sensor after the flow sensor is not Make sure the flow sensor acknowledge) connected . Air flow will continue. On). Refer to the Setup section. Optical sensors look Module module is not installed. No O2 pressure High (cannot The O2 supply has failed. module is on all the way. O2 Flush Failure Low The pressure switch that This alarm occurs if you hold If the alarm occurs detects flush flow has down the Flush button for when flush is not in seen a very long flush more than 30 seconds. --- Absorption after Close the canister release to acknowledged) remove CO2 from exhaled gas. representative. Cycle Contact a qualified prevents mechanical system power (On. Is the module assembled? Are sensors dirty? No CO2 Medium (low CO2 bypass selected. No Circuit Low The patient circuit Install a module.Standby. Sensor not measuring gas in breathing circuit. representative. the battery circuit breakers are battery power. representative. Contact a qualified (Voltage) Failure Monitoring the PEEP Valve or the service (Medium) connections to the PEEP representative. off in the alarm (mechanical ventilation).Aestiva/5 7100 Message Priority Cause Action/Concerns Repair O2 Mon Medium An Oxygen cell has been The Oxygen Monitoring Contact a qualified Disabled connected to a non. 5-10 07/01 1006-0802-000 . spontaneous activity? sensors. representative. Calibrate the flow cmH2O pressure (<-10 cm H2O). Medium (low The mains supply is not Ventilate manually to save Make sure power is Power OK? after connected or has failed power. With active Increase fresh gas flow. Look scavenging. PEEP Safety Minimum Indicates that the power Ventilate manually. PEEP Valve Minimum Indicates a failure of the Ventilate manually. Paw < -10 High Subatmospheric Check patient condition. check for high flow through gas the negative relief scavenging. Valve. settings menu. PEEP Valve Minimum Indicates a problem with Ventilate manually. 30 min of closed. representative for monitoring feature. Patient Circuit Medium Exhaled volume <50% of Check breathing circuit Patient circuit leak Leak? inspired volume for at and flow sensor audio can be turned least 30 seconds connections. activation of this feature if provided or contact a Datex- Ohmeda sales representative for purchase of this feature. mechanical ventilation. connected and acknowledge) and the system is using At full charge. Contact a qualified Valve Failure Monitoring up test of the PEEP Safety service (Medium) Valve Failed. permits approx. Contact a qualified Valve (Drive) Monitoring allowed current draw on service Failure (Medium) the PEEP Safety Valve. On Battery . feature is not active on this service active ventilator system. PEEP Safety Minimum Indicates a higher than Ventilate manually. Contact a qualified (DAC) Failure Monitoring control circuit for the service (Medium) PEEP Valve. valve on the receiver. Ventilate manually. Check settings. service Failure representative. Software Error Min. Service Low Internal calibrations are The system is operational. Does the sensor read 21% O2 in room air? Use different monitor. will not ventilate a ACGO is On. Contact a qualified Watchdog (High) Monitoring is not reliable. 1006-0802-000 07/01 5-11 . breaths in patient circuit during the last 30 seconds. O2 Sensor O2. Indicates that a software Ventilate manually. to the aux. Contact a qualified Monitoring Monitoring patient airway service Channel Failure (Medium) overpressure monitor. Contact a qualified Calibration w necessary for maximum service accuracy. representative. representative. but flow patient at the aux. shutdown Ventilator malfunction. Contact a qualified Shutdown error has occurred. connections or that the ventilator is unable to deliver requested pressure to the patient. gas outlet or acknowledge) set the switch to the common gas outlet for normal operation. patient circuit to the ACGO. Select Medium Fresh gas may not flow to Turn ACGO off or connect the Note: the bag arm Gas Outlet the patient. Replace O2 sensor. service (High) representative. Software Min. Replace Low O2% < 5% Make sure patient receives Calibrate O2 sensor. 5 Alarms and Troubleshooting Message Priority Cause Action/Concerns Repair Pinsp Not Low Indicates a problem with Check breathing circuit --- Achieved breathing circuit connections. gas outlet. Monitoring is not reliable. Pres/Vol Mon Medium (low Outlet selection switch is Connect the patient circuit --- Inactive after set to aux. sensors have seen 3 outlet. Pressure Minimum Indicates a problem with Ventilate manually. Ventilate manually. Sustained Paw High Paw > sustained pressure Check tubing for kinks. Volume Apnea Medium No mechanical breaths Check patient. select Calibrate the flow ventilator and patient Heliox on the ventilator setup sensors. from the breathing Problem with flow sensors? system and inspect for leaks (repair). If the patient is on a heart lung machine. seconds. Calibrate the flow Airway Pressure (High) sec. representative. Contact a qualified Manually prevents mechanical Use a stand-alone monitor. service (High) representative. Contact a qualified Out-of-Range (High) Monitoring is not reliable. Ventilate High A severe malfunction Ventilate manually. Look for leaks in the Drain water buildup absorber system. monitoring. mechanical ventilation.Aestiva/5 7100 Message Priority Cause Action/Concerns Repair Sustained Min. sensors. blockages. System Leak? Low Leak detected between If you are using Heliox. service ventilation and Cycle system power (On. sensors. Contact a qualified Out of Range Shutdown Monitoring is not reliable. 5-12 07/01 1006-0802-000 . circuit. Inspect or replace flow sensors. Other alarms may also If the alarm clears. --- Limit alarm is off (Alarm Settings menu). Ventilator malfunction. Ventilate manually. disconnects. VBUS Fail Min. shutdown Paw > 100 cm H2O for 10 Check tubing for kinks. disconnects. --- or spontaneous breaths Bag as needed. 1 blockages. Calibrate the flow limit for 15 seconds. shutdown Ventilator malfunction. restart occur. Verify Low VE Low The circuit leak audio Set the low VE alarm. Standby-On). service representative. Vext_ref Min. > 5 mL in last 30 Check for disconnects. menu. but the low VE alarm is not consistent with the ventilator settings. select Cardiac Bypass on the alarm menu. > 5 mL in last 120 Check for disconnects. If the patient is on a heart lung machine. activation of this feature if provided or contact a Datex- Ohmeda sales representative for purchase of this feature. select Cardiac Bypass on the alarm menu. Compensation ventilation Select Volume Mode or mode has cleared. --- > 2 min or spontaneous breaths Bag as needed. representative for monitoring feature. 1006-0802-000 07/01 5-13 . feature is not active on this service active ventilator system. representative for monitoring feature. In pressure mode set Pinspir. Indicates a return to --- prevented V T normal operation. If the problem stops. Delivery feature is not active service acknowledge) active ventilator on this system. V T Comp Off Medium The system supplies the Adjust V T manually and Replace the flow (low after set breath but cannot continue without sensor module and acknowledge) adjust ventilation for compensation. seconds. inspect the two flow sensors. activation of this feature if provided or contact a Datex- Ohmeda sales representative for purchase of this feature. etc. or change to select the mode compliance and the pressure mode. V T Comp Avail Low A condition which None. again. 5 Alarms and Troubleshooting Message Priority Cause Action/Concerns Repair Vol Apnea High No mechanical breaths Check patient. V T Comp Medium (low A flow sensor has been The Volume Compensated Contact a qualified Disabled after connected to a non. resistance losses. Volume Mon Medium A flow sensor has been The Volume Monitoring Contact a qualified Disabled connected to a non. Pressure Mode (if available). Ventilation On . leaks. flow. Achieved by inspiratory flow sensor adequate tidal volumes. Modify settings or < set value 6 breaths in a Check I:E. the sustained pressure limit is 6 cm H2O plus Set PEEP. Alphabetical list note: 1 The sustained pressure threshold is calculated from the pressure limit setting. Mechanical • For Plimits ≤ 60 cm H2O. Water build-up in the flow Replace flow sensor sensor tubes? module with the Is a flow sensor tube cracked spare. • For Plimits between 30 and 60 cm H2O. whichever is greater. the sustained pressure limit is 20% of the Plimit plus Set Volume Mode PEEP. 5-14 07/01 1006-0802-000 . • For Plimits ≥ 60 cm H2O. the sustained pressure limit is 12 cm H2O plus Set PEEP. plus Set PEEP Ventilation On – and Pressure Mode • Pmax – Pmin must be less than 50% of set Pinsp or 4 cm H2O. of mechanical ventilation. The sustained limit is calculated as follows: Mechanical • For Plimits ≤ 30 cm H2O. Ventilation Off • For Plimits > 60 cm H2O.Aestiva/5 7100 Message Priority Cause Action/Concerns Repair V T Not Low Tidal volume measured Adjust controls to supply Possible leak. Inspect one way inspired volume for 6 Are the flow sensors correctly valves (breathing breaths with a circle installed? circuit module. and volume check for system row after the first minute settings. V TE > Insp V T Low Expired volume > Check patient condition. Check the or broken? condition of the flow Improper check valve sensor and its tubing. Plimit. the sustained pressure limit is 30 cm H2O. the sustained pressure limit is 50% of the Plimit setting. operation? V T Delivery Too Low V T > 20% of set value for Reduce fresh gas High six consecutive breaths.) module. Mechanical • Sustained pressure limit is 50% of set Pinsp or 4 cm H 2O. whichever is greater. The bellows fills when the Bag/ Leak through Bag/Vent Refer to “Manifold Vent switch is set to Bag or the switch. diaphragm . APL valve does not operate APL valve problem. gently clean. maintenance: APL Valve and bag fills when the switch is set to Bag/Vent” in Section Vent. manifold have been Reseat the absorber seals. is damage. use a different module. Remove the module and look identify the breathing circuit at the tabs on the rear. The ventilator does not read the Ventilator or absorber Ventilate manually. Refer to systems have a flow “Receiver maintenance” in indicator to show this.Refer to “Manifold maintenance” in section 4/User Maintenance. Section 4/Maintenance. 4/Maintenance. 1006-0802-000 07/01 5-15 . Contact a position of the Bag/Vent switch. Ventilator malfunction. Suction supply problem. Active Replace the filter. Contact a qualified service representative to repair the system. Use a different suction supply. Squeeze dry the prep before you use it. The ventilator does not correctly Broken identification tabs. Dirt on the optical sensors Use a DAMP alcohol prep to that read the tabs. malfunction. Filter blockage. qualified service representative to repair the system. unseated. 5 Alarms and Troubleshooting Breathing system problems (no alarm) Symptom Problem Solution(s) Gas scavenging flow is too low. bottom of the main system test plug. If there module. Replace APL Valve seal and correctly. Large breathing system leak not Oval absorber seals in the Disconnect patient wye from quickly located. The electrical power Connect the power cable. The electrical socket the Use a different electrical power cable connects to socket. have power. do not use the system.) wWARNING If a circuit breaker opens frequently.Aestiva/5 7100 Electrical problems (power failure. have power but the mains breaker is open. the circuit breaker rating. has a short. Connect power cable. 5-16 07/01 1006-0802-000 . A circuit breaker opens Equipment connected to Use a different power frequently. The electrical outlets do not The main outlet circuit Close the circuit breaker. the outlet(s) (figure 5-1) supply for some of the uses more current than equipment. Have a qualified service representative repair the system. One electrical outlet does not The outlet circuit breaker Close the circuit breaker. etc. The equipment Do not use the equipment connected to the outlet until it is repaired. An internal fuse is open. The inlet circuit breaker Close the circuit breaker (toggle switch) is open. Not plugged into outlet. (Figure 5-1). cable is not connected. Have a qualified service representative repair the system Symptom Problem Solution Mains indicator is not ON. damaged. The power cable is Replace the power cable. is open. has no power. indicator is ON. Tec 6 has no power. and the mains indicator 1006-0802-000 07/01 5-17 .117 2.108 (No Power) (OK) (No Power) (OK) AA. fuse. Inlet Circuit Breaker* 4.109 2 3 4 1 4 1.96. Master Circuit Breaker-All Outlets* 3. 5 Alarms and Troubleshooting Out In Open Closed AA. Outlet CIrcuit Breakers* AB43062 AA. Mains Indicator * Labels shows ratings.96. Figure 5-1 • Circuit breakers.96. connection. Standby and the auxiliary flowmeter to OFF. A port valve malfunction (the leak Have an qualified service continues if you use a different vaporizer person repair the vaporizer in the same manifold position). Vaporizer port o-rings (external) are Install new o-rings. Correctly install the fails with a vaporizer on. Equipment connected to pneumatic Disconnect the outlet. vaporizer).) Low-pressure leak test The vaporizer is not correctly installed. Incorrect cylinder connection (cylinder Make sure that there is yokes). Contact a qualified service a vaporizer OFF. The vaporizer filler is loose (fill port type Tighten the filler. vaporizer. only one cylinder gasket. representative. wCAUTION: No repair should ever be attempted by anyone not having experience in the repair of devices of this nature. A vaporizer malfunction (the leak stops if Send the vaporizer to a you use a different vaporizer in the same Datex-Ohmeda Service position). Incorrect cylinder connection (DIN Make sure the nut is tight. damaged or not installed. Low-pressure leak with Anesthesia machine problem. Center for repair. and the T handle is tight. equipment. 5-18 07/01 1006-0802-000 . manifold. Set the system switch to fails. the gasket is in good condition.Aestiva/5 7100 Pneumatic problems Symptom Problem Solution High-pressure leak test Controls are not set correctly. . . . . Breathing system parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-9 1006-0802-000 07/01 6-1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2 Main manifold . . . 6-8 Test tools and system parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6 Breathing Circuit Modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Illustrated Parts In this section This section lists user-replaceable parts only. . . . . . . . . . . . . . . . . . . . . . . . . 6-7 Bellows . . . . . . . . . . . . . . . . . . . 6-4 Exhalation valve . For other components refer to the service manual. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30 mm threaded outlet 1406-8214-000 Passive gas scavenging assembly. 30 mm ISO taper 1406-8215-000 Passive gas scavenging assembly. metal 1406-3500-000 14 Upper seal 1406-3414-000 15 Tube. 25 mm taper (DEU) 1406-8217-000 12 AGSS nylon filter. cell 6050-0004-110 O2 port plug 1006-4462-000 O-ring 1406-3466-000 4 Breathing circuit module. circle 1406-8102-000 Breathing circuit module. Not European Community) 1406-8216-000 Active gas scavenging assembly. Cover only 1406-3401-000 b. Holder only 1406-3400-000 2 Flow sensor (plastic) 1503-3220-000 Flow sensor (metal . corrugated.35 mm (ANSI. mapleson/Bain 1406-8211-000 5 Bag arm long (adjustable) 1406-3350-000 Bag arm short (not adjustable) 1406-3380-000 6 Main manifold assembly 1406-8209-000 7 Autoclavable Bellows assembly 1406-8105-000 8 Blank.Aestiva/5 7100 Breathing system parts Item Description (Figure 6-1) Stock Number 1 Flow sensor module (cover & holder) 1406-8208-000 a.autoclavable) 1503-3244-000 Flow port adapter 1503-3847-000 3 O2 sensor. 210 mm 1400-3009-000 16 Elbow transfer tube 1406-3576-000 17 Transfer tube 1406-3575-000 18 Drain dish assembly 1406-8218-000 6-2 07/01 1006-0802-000 . 225 µ (active scavenging systems only) 1406-3521-000 13 Upper dish (2 options) ------ Upper dish absorber plastic (seal not included) 1406-3413-000 CO2 bypass assembly. 6. pop-off valve 1406-3240-000 9 Exhalation valve assembly 1503-8114-000 10 Rear column cover 1406-3455-000 11 Gas scavenging (4 options) -------- Active gas scavenging assembly. Canisters (20) Not Shown: Barb cap. sight glass-drain 0229-2080-100 O-ring. Upper seal (14).095 1b 12* 11 10 20 19 18 17 16 15 14 13 * Active scavenging systems only Figure 6-1 • Breathing system top assemblies 1006-0802-000 07/01 6-3 . plug 0210-0594-300 20 Canister 0229-3015-800 Kit. 6 Illustrated Parts Item Description (Figure 6-1) Stock Number 19 Plug. tube (15). includes items: Upper dish (13). passive scavenging 1406-3524-000 6 7 5 8 9 4 3 2 1a AB. 1406-8207-000 Drain dish (18).23. Corr. 23.071 Top Bottom * Active scavenging systems only Figure 6-2 • Main manifold 6-4 07/01 1006-0802-000 .096 6 4 2 5 2 1 AB.23.Aestiva/5 7100 Main manifold 11 2c 2b 2a 10 9 3 8 7 AB. c) a. Absorber 1406-3315-000 6 Body. main 1406-3301-000 7 APL diaphragm assembly 1406-3331-000 8 Manifold APL/AGSS 1406-3302-000 9 Split ring (Qty 3) 1406-3319-000 10 M6x43 mm Thumbscrew 1406-3304-000 11 M6 x28. Main Manifold. Poppet (Qty 2) 1406-3279-000 c. O-ring (Qty 2) 1406-3278-000 b. bottom main manifold 1406-3303-000 5 Seal. b. 6 Illustrated Parts Item Description (Figure 6-2) Stock Number 1 Seal. Seal only 1406-3314-000 3 Plate. main manifold 1406-3300-000 4 Plate. manifold. APL/AGSS 1406-3316-000 2 Seal. bag/vent assy (a.5 mm Thumbscrew (Qty 2) 1406-3305-000 1006-0802-000 07/01 6-5 . main manifold. top. exhalation valve 1503-3584-000 4 Diaphragm assembly 1503-8121-000 5 Base.021 5 Item Description Stock Number 1 Exhalation valve cover 1503-3583-000 2 Lock ring 1503-3588-000 3 Seat.Aestiva/5 7100 Exhalation valve 6 1 7 2 3 4 AB.23. exhalation valve 1503-3585-000 6 Blank. pop-off valve 1406-3240-000 7 Tube. exhalation valve 1406-3566-000 Figure 6-3 • Exhalation valve 6-6 07/01 1006-0802-000 . Bain-Mapleson D Module 1406-3462-000 Figure 6-4 • Breathing circuit modules 1006-0802-000 07/01 6-7 .102.23. 6 Illustrated Parts Breathing Circuit Modules Circle 12 Module 11 10 9 Bain- Mapleson D 3 8 Module 12 1 2 AB.183 7 3 1 4 6 2 5 7 4 6 5 Item Description Circle Module Bain-Mapleson D Module 1 Assembly (items 2-12) 1406-8102-000 -------- Assembly (items 2-7. AB.23. 1406-8211-000 2 M6-1x23 Thumbscrew 1406-3304-000 1406-3304-000 3 Seal lens. 12) -------. fresh gas. flow sensor 1406-3395-000 1406-3395-000 4 Module body 1406-3390-000 1406-3467-000 5 Seal module cover 1406-3317-000 1406-3463-000 6 Split ring 1406-3319-000 1406-3319-000 7 Back plate 1406-3391-000 1406-3391-000 8 O-Ring check valve 1406-3397-000 Not Used 9 Valve seat check valve 1406-3396-000 Not Used 10 Disc check valve 0210-5297-100 Not Used 11 Retainer check valve 1400-3017-000 Not Used 12 Lens 1406-3394-000 1406-3461-000 Not shown: Tube. bellows 1406-3313-000 Base & latch bellows kit: 1406-8106-000 Figure 6-5 • Bellows parts 6-8 07/01 1006-0802-000 .23.Aestiva/5 7100 Bellows 1 2 3 4 5 6 AB. base 1500-3359-000 7 Base assembly.055 7 Item Description Stock Number 1 Bellows housing 1500-3117-000 2 Bellows 1500-3378-000 3 Rim 1500-3351-000 4 Pressure relief valve assembly 1500-3377-000 5 Latch. rim 1500-3352-000 6 Seal. 6 Illustrated Parts Test tools and system parts Description Stock Number Vaporizer port o-rings. Neutral Gray N9 (Light). Neutral Gray N7 (Medium Dark). 18 ml 1006-4198-000 Touch-up paint. external (Qty 6) 1102-3016-000 Cylinder gasket (pin indexed cylinders only) 0210-5022-300 Ring. 18 ml 1006-4200-000 Communication port cable 1006-3900-000 1006-0802-000 07/01 6-9 . sealing gasket (for DIN 477 and O2 high-pressure hose) 1001-3812-000 Ring. sealing gasket (for N2O high-pressure hose) 1202-3812-000 Handle for yoke tee 0219-3372-600 Yoke plug 0206-3040-542 DIN O2 plug (cylinder connection) 1202-7146-000 Positive low pressure leak test device (BSI) 1001-8975-000 Positive low pressure leak test device (ISO) 1001-8976-000 Negative low pressure leak test device 0309-1319-800 Test lung 0219-7210-300 Test plug 2900-0001-000 Cylinder wrench 0219-3415-800 Cylinder wrench (if cylinder cover installed) 0219-3412-800 Cylinder wrench (DIN 477 and high-pressure hose) 1202-3651-000 Krytox 1001-3854-000 Touch-up paint. 18 ml 1006-4199-000 Touch-up paint. Neutral Gray N8 (Medium). Aestiva/5 7100 6-10 07/01 1006-0802-000 . . . . . . . . . . . . . 7 External Communications In this section This section describes how to communicate between the ventilator and a data collection system or an external monitor. . . . 7-2 Protocol description (Datex-Ohmeda Com 1) . . . . . . . . . . . . . . . . . . . . . . . It also tells you what data can be sent and received. . . . . . External communications . . . . . . . . . . . . . . . . . . 7-3 1006-0802-000 07/01 7-1 . . . . 15-pin Female D connector .Aestiva/5 7100 External communications The ventilator has an RS-232C electrical interface.Signal ground 7-2 07/01 1006-0802-000 .Data Communications Equipment configuration (DCE): • Pin 6 . The RS-232 connector allows serial input/output of commands and data. Electrical Interface RS-232C signal standards.Transmit data • Pin 5 .Receive data • Pin 13 . 7 External Communications Protocol description (Datex-Ohmeda Com 1) Serial Baud: 19.2K Communication Byte format: Start bit + 7 data bits + parity bit + stop bit Parameters Parity: ODD Software Interface Datex-Ohmeda Com 1 Waveform Communication Protocol. Command Headers: <ESC>VTD DISABLE CHECKSUM <ESC>VTE ENABLE CHECKSUM <ESC>VTQ ENABLE COMPRESSED MODE <ESC>VTS SLAVE MODE (RESETS AUTO MODE) <ESC>VTX AUTO MODE <ESC>VT$ SEND SETUP DATA <ESC>VT? SEND ALL DATA <ESC>VTW ENABLE WAVEFORM DATA Response Headers: :VTD MEASURED DATA RESPONSE :VTM SETUP DATA RESPONSE :VTN NACK (negative acknowledge) :VTQ STATUS DATA RESPONSE :VTW WAVEFORM DATA RESPONSE :VTR ALARM SILENCE SWITCH PRESSED RESPONSE :VTY ACK (positive acknowledge) 1006-0802-000 07/01 7-3 Aestiva/5 7100 DEVICE COMMANDS Sent Data Transmit Mode Select Commands By External <ESC>VTXc<CR>Auto Mode Device <ESC>VTSc<CR>Slave Mode Data Format Mode Select Commands <ESC>VTQc<CR>Compressed Format Data Request Command <ESC>VT?c<CR>Send All Data (Valid in Slave Mode only) <ESC>VT$c<CR>Send Setup Data Enable Waveform Data Mode <ESC>VTWabc<CR>Send Waveform Data <ESC>VTW<CR>=header a&b = any one of the following: (Note 1: order determines order of data in response packet) (Note 2: b valid only if a!= 0) 0 Turn waveform data OFF P Include Pressure Data F Include Flow Data V Include Volume c Checksum <CR> Terminator Checksum Control Commands <ESC>VTEc<CR> Enable Checksum Mode <ESC>VTDc<CR> Disable Checksum Mode (checksum byte ignored in this command, but cannot be <CR>) 7-4 07/01 1006-0802-000 7 External Communications DEVICE ACK Response RESPONSES Sent Back By :VTYc<CR>Positive Acknowledge Response. Ventilator NAK Response :VTNc<CR>Negative Acknowledge Response. Alarm Silence Switch Pressed Response :VTRc<CR>Alarm Silence Switch Press Response (if no alarms are on or all displayed alarms are silenced). Compressed-Data Measured Data Response In auto mode, the Measured Data Response will be transmitted at the end of a breath or 10 seconds from the last transmission, whichever occurs first. :VTDaaaabbbbdddeeefffggghhhiiijc<CR> Compressed Data Measured Data Response (each entry is zero filled and right justified--i.e. aaaa = 0095) “?” means bad data due to any technical problem(s); “-” means data not available due to system state. aaaa Measured tidal volume mL, ?, - bbbb Measured minute volume L*100,?, - ; example: 1000 equals 10.00 L ddd Measured respiratory rate /min,?, - eee Measured oxygen level % O2,?, - fff Measured max positive pressure cm H2O,? ggg Measured inspiratory plateau pres cm H2O,? hhh Measured mean pressure cm H2O,? iii Minimum pressure cm H2O,? j Measured data status 0100000x (bit 0=1=new breath data; bit 0=0=10 second data) c Checksum 1006-0802-000 07/01 7-5 0 L) kkk Low minute volume alarm limit L*10 (e.g. 650 = 65. 090 =900 ml) nnn High oxygen alarm limit % O2 ooo Low oxygen alarm limit % O2 q Ventilation mode: “v”=volume Mode.’p’=pressure mode.’b’=Vt Compensation off. :VTQaaaabbbddddeeffggghhiijjjkkklllmmmnnnoooqrrrrrrrrrrrrc<CR> Compressed Data Status Data Response aaaa Set tidal volume mL bbb Set tidal volume /min dddd Set I:E ratio 1:eee.g. 150 =1500 ml) mmm Low Vte limit mL/10 (e.g.Aestiva/5 7100 Compressed-Data The Status Data Response will be transmitted every 1 second (if a change Status Data Response occurs in the status data) or a minimum of once every 10 seconds. Note that the status byte bit is set=1 for an active condition and 0 for an inactive condition.g.e ee Inspiratory pause % Pause ff Set PEEP cm H2O ggg Set peak pressure limit cm H2O hh Set inspired pressure cm H2O ii Set sustained pressure alarm limit cm H2O jjj High minute volume alarm limit L*10 (e. 050 = 5.’-’=bag mode only rrrrrrrrrrrr Status bytes (see bitmaps below) c Checksum 7-6 07/01 1006-0802-000 .0 L) lll High Vte limit mL/10 (e. No Fresh Gas Flow 1006-0802-000 07/01 7-7 .1 D4 .High O2 D0 .Flow Valve Failure D2 . Message Vent. Each Byte has eight bits of data from D7 (MSB) to D0 (LSB).No Pressure Mode/PEEP D2 .Bellows Empty D1 .High Vte Failure D4 . Message Vent.Low Paw D2 .Inspiratory Overshoot D4 .PEEP Not Achieved D1 .O2 Calibration Error D6 .Sustained Paw D4 .Manifold Pressure Sensor D3 .Low O2 D1 .No Insp Flow Sensor D0 . Message D0 .Check Flow Sensors D4 .Exp Reverse Flow D3 .Gas Inlet Valve Failure D3 . 7 External Communications Status Bytes Bitmaps The status bytes are a string of 12 bytes.High Pressure Limit Reached D6 .Volume Apnea > 2 min (min sys) Bit byte 5 Bit byte 6 Vent. Message D0 .1 Pressure Mode Available D6 .1 D2 .1 D3 .Volume Apnea D6 .1 Bit byte 3 Bit byte 4 Vent.No Exp Flow Sensor D1 . Message Vent.Insp Reverse Flow D2 . Bit byte 1 Bit byte 2 Vent.12 Hour Test D4 .Check O2 Sensor D5 .Vt Not Achieved D5 .Vdel Mismatch D6 .High VE D2 . starting from the left (Byte 1) to the right (Byte 12).“Bootup GIV Failure” D5 .Insp Vt/Vte Mismatch D5 .Sustained Paw (shutdown) D3 .Pinspired Not Achieved D0 .Low Vte D3 . Message D0 .High Paw D1 .Inspiration Stopped D5 .No O2 Pressure D6 .Sub-Atmos Paw D5 .Low VE D1 . CPU Failure D2 .+15V Manifold Out-of-Range D3 .No Battery D2 .Bootup Memory Failure Bit byte 9 Bit byte 10 Vent.Aestiva/5 7100 Bit byte7 Bit byte 8 Vent. Volt.1 D1 .Sensor(s) Cal Due 7-8 07/01 1006-0802-000 . Out Of Range Bit byte 11 Bit byte 12 Vent.Vaux_ref Out-of-Range D5 .Memory (RAM) Failure D5 .Mechanical Ventilation On D6 . Message D0 .Volume Mode Active D1 . Message Vent.Volume Compensation Off D5 .Internal Vent.Low Battery Charge D3 .Low Battery D4 .Very Low VE Limit Failed D4 .Memory (EEPROM) Failure D3 .Hardware Watchdog Failure D1 . Clock Too Slow D3 .Memory (redundant storage) D5 ."No confirmation of changed D2 .Internal Vent.Software Watchdog Failure D0 .Apnea Detect ON D2 .1 D6 . Message Vent. Out Of Range Fail D6 .Apnea Alarm Silenced setting" D3 .Heliox Mode is ON D4 . Curr.Control Settings Input Has D3 .Alarms Silenced D5 .Flow Sensor Cal Data Corrupt D6 .Display Voltage Out-of-Range D4 .Low Battery (shutdown) D5 .Batt.+15V SIB Out-of-Range D2 .Auxiliary Breathing Circuit D0 .On Battery D1 .Memory (flash) Failure D4 . Message D0 .Fail Batt.+Vanalog Failure D0 .CPU Internal Error D4 .-Vanalog Failure D1 .Vext_ref Out-of-Range D6 .A/D Converter Failure D1 . Message Vent. Message D0 . Clock Too Fast D2 .Memory (video) Failure D6 . of 2] waveform signal selected aaa 1st 16 ms waveform sample (0 -> “000”. 4095 -> “FFF”) bbb 2nd 16 ms waveform sample . Each data value is a 3-digit. 5 = 7850. D=Russian. Up to 2 blocks of fifteen (15) data samples taken every 16 ms. nnn 14th 16 ms waveform sample ooo 15th 16 ms waveform sample 1006-0802-000 07/01 7-9 ... 1=Spanish. 9=Polish.nnnooo]c<CR> Waveform Data Response for each [max.01-99.99 b Language (see Note 1) 0-9 & A-F dd Display contrast setting 1-64 e Alarm volume setting 1-5 ff Altitude setting -4 to 36. A=Chinese. 8 = 7100 c Checksum Note 1: Language: 0=English. will be sent with each message.. 5=Swedish. 6=French.. :VTW[aaabbb. 7=Italian. C=Hungarian.nnnooo][aaabbb. 2=German. 1 = 7810. 7 External Communications Setup Data :VTMaaaabddeffghc<CR>Setup Data Response Response aaaa Software revision number 0001-9999 = 0. 8=Danish.. E=Portuguese. zero filled. Waveform Data If Waveform Data Mode is enabled. right justified ASCII Hex representation of a 12 bit binary value. B=Czech. 512 -> “200”. 6 = 7900/Aestiva.. F=other. A = air h Ventilator model number 0 = 7800. 4=Dutch. 3=Kanji. a Waveform Data Response will be Response transmitted every 240 ms. in 100’s of meters g Drive gas O = oxygen. 2 L Scale: Raw Scaled Xmit.Aestiva/5 7100 Waveform Data shall be scaled as follows: Pressure Range: -20 . -100 512 “200” 0 2048 “800” +100 3584 “E00” Volume Range: 0 .120 cmH2O Scale: Raw Scaled Xmit. 0 512 “200” 2 3584 “E00” 7-10 07/01 1006-0802-000 .+100 L/m Scale: Raw Scaled Xmit. -20 0 “000” 0 512 “200” 120 3584 “E00” Flow Range: -100 . 1006-0802-000 07/01 7-11 . The volume will be reset to 0 at the start of a inspiration. expiratory flow will reduce the volume. The flow may be sampled from the inspiratory flow sensor and/or the expiratory flow sensor. Inspiratory flow will increase volume. Flow For each breathing system the samples will be every 16 ms. 7 External Communications The Waveform Data will be sampled as follows: Pressure Samples will be every taken from the airway pressure sensor every 16 ms. depending on the circuit type and the phase of the breath. The inspiratory flow sample will always be positive and the expiratory flow sample will always be negative. Circle Breathing System If inspiratory flow > threshold (start of inspiration) If inspiratory flow is increasing Use inspiratory flow 16 ms sample Else if inspiratory flow is decreasing and expiratory flow > threshold Use expiratory flow 16 ms sample Else Use inspiratory flow 16 ms sample Else If expiratory flow > threshold (start of expiration) Use expiratory flow 16 ms sample Else Use the 16 ms flow sample from the previously used flow sensor Volume Circle Breathing System Samples will be every 16 ms and will be based on integration of flow values. Aestiva/5 7100 7-12 07/01 1006-0802-000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-13 Ventilator Theory . . . . . 8-5 Electrical power . . . . . . . . . . . . . . . . . . . 8-11 Environmental requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-16 Ventilation operating specifications . . . . . . . . . . . . . . . . . . 8-21 Suction regulators (optional) . . . . . . . . 8-19 Ventilator accuracy data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-12 Breathing system specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-2 System specifications . . . . . . . . . . . . . . . . . . . . 8-23 1006-0802-000 07/01 8-1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Specifications and Theory of Operation In this section System pneumatic circuits . . . . . 8-8 Physical specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-23 O2 flowmeter (optional) . . Aestiva/5 7100 System pneumatic circuits Pneumatic circuit diagram O2 4 O2/He or O2 Air N 2O CO2 5 4 4 4 4 3 5 5 5 5 28 27 1 2 3 1 2 3 1 2 3 3 6 6 6 6 14b 7 7 13 14a 8 21 15 10 18 9 11 LINK-25 19 16 LINK-25 22 12 20 17 23 24 24 26 Key to Symbols: Pneumatic connection: Filter: Direction of flow: 25 AA.131 Check valve: 8-2 07/01 1006-0802-000 .96. N2O flow control valve 17. Pipeline connection 3. System switch 9. Cylinder connection 5.Switch (sensed by the ventilator) 15. 38 kPa (5. Pressure relief. Pressure relief. O2 Flush a. Air flow tube 21. Pipeline pressure gauge circuit diagram 2. CO2/Heliox gas balance regulator 22. Supply connections for the ventilator 8. Common gas outlet 1006-0802-000 07/01 8-3 . Vaporizer 26.Flush valve b. opens at approx. 918 kPa (133 psi) 7. Air flow control valve 20. N2O balance regulator 16. opens at approx. CO2/Heliox gas flow control valve 23. flowmeter (optional) 28.5 psi) 27. O2 flow tube 13. Cylinder pressure gauge 4. N2O flow tube 18. Alarm for low O2 supply pressure 10. Secondary O2 regulator 11. CO2/Heliox gas flow tube 24. Cylinder pressure regulator 6. Vaporizer port valve 25. 8 Specifications and Theory of Operation Numbered items in Pneumatic 1. Flush regulator 14. Air secondary regulator 19. Aux. O2 flow control valve 12. Air. O2 flows to the rest of the system and there is a minimum flow of 25 to 75 mL/min through the O2 flowmeter. A secondary regulator supplies a constant O2 pressure to the flowmeter. Changes in O2 pressure do not affect Air.Aestiva/5 7100 Gas supplies Gas goes into the system through a pipeline or cylinder connection. and into the breathing system. and Balance regulators control the flow of N2O and the optional gas (CO2. air flows to the rest of the system. Gauges show the cylinder and pipeline pressures. The flush valve supplies high flows of O2 to the fresh gas outlet when you push the flush button. Pipeline or regulated cylinder pressure directly supply Air to the ventilator (Air Ventilators). If the pressure is too low. When the system switch is ON. O2 flow Pipeline or regulated cylinder pressure supplies O2 directly to the pneumatic (Items 7-14. A secondary regulator supplies the air flow control valve. 8. A pressure relief valve helps protect the system from high pressures. leaving an insufficient reserve supply in case of pipeline failure. This stops flow during an O2 supply failure and ensures that (Items 7. Heliox) third gas flow to the flow control valves. (Item 26) to the fresh gas outlet. Cylinder supplies could be depleted. A regulator decreases the cylinder pressures to the appropriate system pressure. w WARNING Do not leave gas cylinder valves open if the pipeline supply is in use. 8-4 07/01 1006-0802-000 . When the system switch is ON. Because there is no balance regulator. A chain linkage on the N2O and O2 flow controls helps keep the O2 concentration higher than approximately 20% at the fresh gas outlet. air flow continues at the set rate during an O 2 supply failure. An electrical switch monitors the O2 supply pressure. N2O. Mixed gas The mixed gas goes from the flowmeter outlet through the vaporizer that is ON. and 15-23) the hypoxic gas pressures decrease with the O2 supply pressure. keep the cylinder valve closed. an alarm appears on the ventilator. 27) outlet and the ventilator (O2 Ventilator). A second regulator decreases the pressure for the flush valve and the auxiliary flowmeter. A message on the ventilator tells you when Flush is ON. A pressure relief valve sets the maximum outlet pressure. filters. Oxygen pressure at a control port adjusts the output of the regulator. • When a pipeline supply is connected. The flush switch uses pressure changes to monitor the position of the flush valve. All (Items 1-6) connections have indexed fittings. To help prevent problems with the gas supplies: • Install yoke plugs on all empty cylinder connections. and check (one-way) valves. Heliox. large cylinder kit available for O2 and N2O. N2O. DIN 13252. N2O. • Without a rear panel label: Approximately 587-690 kPa (85-100 psi). 5 cylinders total on the 3-vaporizer model. Air. S90-116 (French Air Liquide). BSPP 3/8 (Scandinavian) or NIST (ISO 5359). Pressure relief valve: Approximately 951 kPa (138 psi). Pressure displays: Color coded gauges. Air (maximum: 2 cylinders of each gas. All fittings available for O2. N2O. 4 cylinders total on 2-vaporizer model. Cylinder gases: CO2. 1006-0802-000 07/01 8-5 . Pipeline connections (filtered): DISS-Male. 414 kPa (60 psi). 2 cylinders max. Pipeline inlet pressure: 280-600 kPa (41-87 psi). Air). Primary regulator output pressure: • With a rear panel label: The primary The primary < 345 kPa regulator is set to < 414 kPa regulator is set to pressure less than pressure less than 345 kPa (50 psi). on pendant model. and N2O. Cylinder connections: Pin indexed (all gases). Air. Pneumatic outlet pressure: O 2 source Outlet pressure Pipeline pressure Pipeline pressure Pin indexed cylinders 240-345 kPa (35-50 psi) DIN 477 cylinders and 240-415 kPa (35-60 psi) high-pressure hoses Max. nut and gland DIN 477 (O2. O2. 8 Specifications and Theory of Operation System specifications Pneumatic Gas supplies: Pipeline gases: O2. AS 4059 (Australian). DISS-Female. pipeline draw: Not to exceed 60 L/min at 280 kPa (41 psi) under steady flow or averaged over 10 seconds. 05-0.05 -0.5 207 241.5 kPa (35 psi) 100 90 Flow 80 (L/min) 70 60 50 40 30 20 10 0 34.5 276 Outlet Pressure (kPa) Outlet pressure (kPa) (psi = kPa ÷ 6. outlet relief: Flow Flow rates: Minimum O2 flow: 25 to 75 mL/min.5 69. 8-6 07/01 1006-0802-000 .Aestiva/5 7100 Pneumatic outlet: O2 DISS or Mini schraeder connections: 130 DISS Power Outlet Flow (L/min) ISO Power Outlet 120 Inlet pressure constant 110 at 241.5 L/min 0.05 -0. Scale Scale Gas Flow Range (One Flow Tube) (Two Flow Tubes) O2 0.8947) Common gas Valve limits fresh gas pressure to 138 kPa (20 psi) at the flush flow.95 L/min 1-15 L/min N2O 0-10 L/min1 -----------------.5 138 172.5 L/min ------------------ 1. 0. The link system sets the nominal O 2 flow of the total O2 and N2O flow.05-15 L/min -----------------.95 L/min 1-15 L/min He 0-15 L/min 0.0 103.5-15 L/min ------------------ CO2 0-0.95 L/min 1-10 L/min Air 0-15 L/min 1-15 L/min 0. 0.05 -0. Accuracy Class 2. these changes can be larger than the tolerances.3 kPa (absolute) (14. O2 supply failure alarm and shutoff: O2 Pressure O2 supply failure alarm: 193 to 221 kPa (28 to 32 psi) N2O shutoff: 3. Flush flow: 35-50 L/min.5 kPa (0. 8 Specifications and Theory of Operation Accuracy: At 20 °C with gas supply pressures at 345 kPa (50 psi) and an outlet pressure of 101. Different breathing circuit pressures.5 or better. barometric pressures or temperatures change the accuracy. With some conditions.5 psi) CO2 or Heliox shutoff: 69 kPa (10 psi) 1006-0802-000 07/01 8-7 .7 psi) flowmeter accuracy agrees with VDE 3513 Part 3. 2 Ω 8-8 07/01 1006-0802-000 . System leakage current limit . Electrical circuit breakers: Inlet Circuit Breakers: 100-120 Vac 220-240 Vac No outlets: 5A 3A With outlets: 10A 6A Master Circuit Breaker (all outlets): 100-120 Vac 220-240 Vac 5A 3A Outlet Circuit Breakers: 100-120 Vac 220-240 Vac (3) 2A (3) 1A (1) 3A (1) 2A Japan: (2) 2A (1) 4A Circuit Breaker for Tec 6 Outlet: Limited by master circuit breaker. Note: Products connected to the electrical outlets may increase the leakage current above these limits. • IEC rated systems (Not USA and Canada): <500 µamps for the system and all systems connected to electrical outlets. Resistance to ground: <0.do not exceed: • UL and CSA rated systems (USA and Canada): <300 µamps for the system and all systems connected to electrical outlets.Aestiva/5 7100 Electrical power Supply voltage: 100-120 or 220-240 Vac ± 10% at 50 or 60 Hz. Class B. Battery information The Aestiva/5 7100 is not a portable unit. Immunity Levels: The Aestiva complies with the requirements of EN 60601-1- 2 (Electromagnetic Compatibility . IEC 801-3. The following basic EMC standards were applied to verify conformance. Thus.3L/250V. 3. Emissions: CISPR 11 Group 1 (EN 55011). 2 kV line to earth. Electro-magnetic Environment: Suitable for use in the EM environment described in compatibility EN 60601-1-2. 8 Specifications and Theory of Operation Power cord Length: 5 meters. Batteries meet the following: 1. 3 V/m. Only Datex-Ohmeda service representatives are to replace the battery. 2 kV power line. IEC 801-5. 1006-0802-000 07/01 8-9 . Unit functions to specifications through the transition to battery power. 5. the battery is in a float charge state most of the time. Long float charge life. Batteries must be disposed of in accordance with applicable regulatory requirements in effect at the time and place of disposal. IEC 801-4. Battery terminals and connecting wires are protected against short circuits. Battery pack has an auto-resettable thermal fuse. Voltage rating: 90 to 264 Vac. 4. 3 kV contact. Current capacity: 10 A. 1 kV line to line.Requirements and tests). 2. Capacity to operate for 30 minutes. Batteries are used as backup power in case of a power failure. Immunity: IEC 801-2. 8 kV air. Type: Three conductor medical grade power supply cord. a sealed lead acid battery supplies battery backup. Electrical fuses T6. Aestiva/5 7100 Electrical block diagram Inside Machine Control Module AC Inlet Universal Power Supply Power Fan Cord 2 fuses System Breaker 5mm x 20mm Line Filter +6V @ 5A max T2L/250 V +9V @ 0.5A max 6 Volt Battery Serial Adapter Board RS232 Control Board MCF5206e ColdFire Processor Memory and I/O Decoding Flash, SRAM & EEPROM SCR Circuitry A/D – D/A Converter Watchdog System Inspiratory Valve Control (10 VA limited) Machine/Ventilator O2 Flush PEEP Valve Control (10 VA limited) On/Standby switch 12 VDC Supply (10VA limited) DC Supply Monitoring O2 Supply Battery Management PEB Monitoring Interface Speaker Airway Assembly Pressure Inspiratory Flow Transducers Transducer Expiratory Flow Transducers LCD Backlight Pneumatic Engine O2 Sensor Flow Sensors LCD Display Keyboard Inspiratory Valve Bag/Vent Switch 320 x 240 Membrane PEEP Valve ACGO Switch Control Panel Switch Switches Supply Pressure Switch Canister Release Switch PEEP Safety Valve Absorber Bypass Switch Rotary Breathing Circuit ID Encoder Switch Breathing System 8-10 07/01 1006-0802-000 8 Specifications and Theory of Operation Physical specifications All specifications are approximate values and can change without notice. AB.43.074 System 2-vaporizer 3-vaporizer Height: 135.8 cm (53.4 in) 135.8 cm (53.4 in) Width: 75 cm (29.5) 93 cm (36.6 in) Depth: 83 cm (32.7 in) 83 cm (32.7 in) Weight: 136 kg (300 lbs) 154 kg (340 lbs) Top shelf options: Option: 87.5 cm W x 41 cm D 87.5 W x 41 cm D (34.4 in W x 16.1 in D) (34.4 in W x 16.1 in D) Option: 67.5 cm W x 41 cm D 67.5 W x 41 cm D (26.6 in W x 16.1 in D) (26.6 in W x 16.1 in D) Option: 47.5 cm W x 41 cm D — (18.7 in W x 16.1 in D) Shelf weight limit: 46 kg (100 lbs) per shelf 46 kg (100 lbs) per shelf Casters 12.5 cm (5 in) with brakes on the front casters. Drawers Standard: 10.5 cm H x 38.5 cm W x 26 cm D (4.1 in H x 15.2 in W x 10.2 in D). Optional: 14.5 cm H x 38.5 cm W x 26 cm D (5.7 in H x 15.2 in W x 10.2 in D). Ventilator display 120 x 90 mm c VGA. 1006-0802-000 07/01 8-11 Aestiva/5 7100 Environmental requirements Temperature Operation: 10 to 40 °C, (Oxygen cell operates to specifications at 10 to 40 °C). Storage: -5 to 50 °C, Oxygen cell storage is -5 to 50 °C, 10 to 95% Rh, 500 to 800 mm Hg. Humidity Operation: 15 to 95% Rh, non-condensing. Storage: 10 to 95% Rh, non-condensing per IEC 68-3. Altitude Operation: 500 to 800 mm Hg (3565 to -440 meters). Storage: 375 to 800 mm Hg (5860 to -440 meters). Compensation range: 525 to 795 mmHg (3,000 to -100 meters). 8-12 07/01 1006-0802-000 Remove the breathing circuit module to access. System leakage: These values are for continuous pressure and are higher than those expected during mechanical ventilation. To maintain compliance with standards.02 psi) expiratory resistance at 1 L/sec. System compliance: Volume of gas lost due to internal compliance (bag mode only) 5. ≤ 300mL/min total at 3 kPa (0. choose a patient circuit with low resistance to flow.15 mL/cm H2O. Common gas outlet: ISO 5356 type connector (Standard 22 mm OD or 15 mm ID conical friction fit connectors). 8 Specifications and Theory of Operation Breathing system specifications General Volume: 5500 ml Absorbent: 1. 6 Breathing System Resistance Inspiratory Pressure 5 Expiratory drop (cm H2O) 4 3 2 1 0 0 10 20 30 40 50 60 Flow rate (L/min) 1006-0802-000 07/01 8-13 .15 kPa (. ≤75 ml/min for all connectors and two part tubes and ≤ 225 mL/min for all other breathing system assemblies.35 kg (400 ml) / canister Connections: Breathing circuit modules: Circle or Bain modules.4 psi). Auxiliary outlet: ISO 5356 type connector on the front of the breathing system (Standard 22 mm OD or 15 mm ID conical friction fit connectors) (Optional). Breathing system resistance: Y piece or T piece adds ≤0. • Negative pressure relief: 0. hose barb 30 .77 30 0.163 APL Valve: Approximately 0 to 70 cm H2O Pressure Flow Data (APL Valve Completely Open): Flow (L/min) Flow (L/sec) APL Pressure cm H2O 3 0. Active scavenging: Particle filter at the outlet has a pore size of 225 microns.17 1.Aestiva/5 7100 Pressure required to open inspiratory and expiratory valves: Dry: 0.80 SLPM 1.68 3. All flow data uses a new filter.14 50 0.34 1.100 SLPM n/a * Other market-specific connectors may be available.83 4.0 5.14 System volume: 5500 mL Gas scavenging Passive scavenging: • Positive pressure relief: 10 cm H2O.26 20 0.51 2.106 90 cm H2O N/A 0.089 0. • Outlet connector: 30 mm male taper swivel. 8-14 07/01 1006-0802-000 .6 kPa at 75 SLPM Venturi/Ejector ain.5 cm H2O.063 60 cm H2O N/A 0.12 10 0. high vacuum DISS EVAC 36 ±4 SLPM @ 300 300 mmHg mmHg (12 inHg) Minimum vacuum High flow BS6834 50 .49 cm H2O Moist: 0.91 cm H2O Breathing system leakage (average during use): Pressure Bag Mode (ml/min) Vent Mode (ml/min) 30 cm H2O 0. Disposal System Type Outlet Swivel Connector * Flow Range Pressure Low flow.06 60 1.05 1.36 40 0. 05 / 40.) CONTROL OPTION MODULE SENSE SERVICE INLET TEST FILTER PORT AIRWAY 2 MICRON PRESSURE XDUCER (PEB) PNEUMATIC PRESSURE ENGINE AIRWAY ATMOSPHERE BOARD PRESSURE REGULATOR PROPORTIONAL GAUGE 25 PSIG INSPIRATORY DRAIN CANISTER BULKHEAD MECHANICAL CONNECTOR VALVE PRESSURE RELIEF VALVE RELEASE 241/689 241/689 172 4.) SENSE 10 L/min MAIN ABS OUT SENSOR SUPPLY MANIFOLD PRESSURE SWITCH (N.) 17 psi PEB P CABLE TO INSPIRATORY PEEP SAFETY VALVE VENTILATOR GAS IN GAS OUT O2 % MONITOR BYPASS 4.4 cm H2O BYPASS -10 TO 110 cm OPTION H2O ATMOSPHERE SENSE VENTILATOR PNEUMATIC ENGINE ACGO INSPIRATORY ELASTOMER POSITION 02% EXP INSP RECEIVER (2L) SENSE FLOW FLOW ELBOW XDUCER XDUCER +10 CM H20 RELIEF VALVE 80 L/min CONTINUOUS TO BAG/ @ 241/689 kPa BYPASS CANISTER VENTILATOR CONTROL BREATHING DISPOSAL OPTION RELEASE POSITION PANEL CIRCUIT SYSTEM DRAIN CANISTER SENSE LEVER SENSE OPEN/CLOSED ID (MIA) 30 mm -2 CM H20 RELEASE OR DRAGER RELIEF VALVE LEVER MONITORING INTERFACE ASSEMBLY OR 12 mm ABSORBER/CO2 BYPASS HOSE BARB PASSIVE GAS SCAVENGING INTERFACE OPTIONAL RECEIVER (2L) MIA SUPPLY HIGH OR LOW CABLE TO TO FLOW RESTRICTOR FILTER VENTILATOR AIR/OXYGEN ANESTHESIA DISPOSAL MACHINE CONTROL SYSTEM MODULE DISS EVAC OR BSI ROOM OR BOC FLOW AIR INDICATOR (OPTIONAL) ACTIVE GAS SCAVENGING INTERFACE 8-15 AB.1006-0802-000 BELLOWS ASSEMBLY BELLOWS CANISTER BAG/VENTILATOR FRESH GAS FLOW 07/01 BELLOWS POSITION 0.C.5 VDC (N.5 VDC (N.43.C.) 110 cm H2O LEVER kPa kPa kPa 70 L/min ABSORBER STANDARD TRANSFER FREE BREATHING TUBE CHECK VALVE 1.5 VDC (N.0 SENSE L/min O2 FLUSH 50 L/min EXPIRATORY CHECK VALVE ADJUSTABLE PRESSURE EXPIRATORY LIMITING RELIEF VALVE FLOW SENSOR Breathing system schematic (APL) BAG/VENTILATOR BAG/VENTILATOR SWITCH DRAIN EXHALATION VALVE BAG ABS IN MANIFOLD FLOW PATIENT AGSS BREATHING SENSOR LUNG 200 mL CIRCUIT MODULE VOLUME Y MANIFOLD AGSS MODULE (FSM) TRANSFER SYSTEM INSPIRATORY (CIRCLE) CHECK VALVE CALIBRATED BLEED ORIFICE PROPORTIONAL AIRWAY PEEP VALVE INSPIRATORY ATMOSPHERE FGF PRESSURE FLOW 4.073 8 Specifications and Theory of Operation .C.O. 8-16 07/01 1006-0802-000 .5 to 1. while a second. If necessary. Volume measurements come from two flow sensors in the flow sensor module. based on the user’s Plimit setting. A precision flow valve controls the amount of flow provided to the patient. At high airway pressures this can cause a slight hiss during inspiration. volume monitoring uses the right (expiratory) flow sensor. When using the optional Bain circuit module. while the left hand sensor measures the bidirectional flow to the patient. the right hand flow sensor measures fresh gas flow. With circle circuit modules. If it is blocked.Aestiva/5 7100 Ventilator theory General The Aestiva/5 7100 ventilator pneumatics are located in the rear of the breathing system. small leaks and gas compression within the breathing circuit. Note that the 7100 provides a small amount of bleed flow through a pneumatic resistor to control the exhalation valve. The average volume changes due to compression in the breathing circuit is small (0. Electronic control of PEEP is achieved by maintaining a small amount of sealing pressure on the exhalation valve. The ventilator uses the other flow sensor to adjust the tidal volumes being delivered into the patient circuit tubing for changes in fresh gas flow. add the compression loss to the tidal volume setting (Volume Control mode). which changes with the flow. During exhalation. the sealing pressure on the exhalation valve is released and flow exhausts from the breathing circuit. Breaths are provided to the patient by sealing the exhalation valve and providing flow to the breathing circuit. wWARNING Do not try to silence the pneumatic resistor. A third pressure transducer measures the airway pressure at the outlet of the breathing circuit.25 ml/cm H2O). The transducer measures the pressure change across each sensor. the ventilator can malfunction and cause patient injury. There is no adjustment for compression in the patient circuit. Two tubes from each sensor connect to a transducer. smaller flow valve is used to control the exhalation valve sealing pressure. Safety redundant valves are used to limit the airway pressure. 8 Specifications and Theory of Operation Modes The system has two modes of mechanical ventilation: • Volume control • Pressure mode Use the main menu to set the mode. This is called tidal volume compensation. 1006-0802-000 07/01 8-17 .43. “V T Comp Off” continues to supply the set tidal volume but cannot compensate for the effects of fresh gas flow and breathing circuit compliance losses. The ventilator calculates a set flow and the length of the inspiratory period from the I:E and frequency settings. If tidal volume compensation is not enabled.068 Figure 8-2 • Volume control diagram Typically a volume control waveform shows increasing pressure during inspiration. the ventilator adjusts gas flow to the bellows based on measured inspiratory volumes. To make sure that the set flow is actually delivered. An optional inspiratory pause can be set to improve gas distribution in the lungs.” and Volume Control is replaced by Volume Mode in the main menu. Volume control supplies the set tidal volume during inspiration.122 Figure 8-1 • Ventilation Setup Menu Volume control AB.29. AB. VT Comp Off During some malfunctions in volume control mode. the ventilator functions the same as “V T Comp Off. the ventilator automatically turns volume compensation off. • No mechanical ventilation. I:E. Minimum Monitoring • Shows data. PEEP is available and is additive. • The software goes to minimum monitoring when a non-recoverable error occurs in boot-up or normal operations. Rate and I:E settings are used to create the breathing profile. the user adjusts Pinspired until Pmax is reached. • “Ventilate Manually” and specific failure message. While observing the monitored Pmax value.Aestiva/5 7100 Volume control settings • VT (tidal volume). Plimit (pressure limit). Pinsp. Limited inspired pressure mode Selecting Pressure mode generates a pressure plateau linked to the P inspired setting. Pressure Pinsp (control pressure). • I:E. • Rate. 8-18 07/01 1006-0802-000 . • PEEP. mode settings Rate. The ventilator delivers a constant 30 L/m (inspiration) during this mode. • Plimit. PEEP. • “Monitoring Only” alarm message. 8 Specifications and Theory of Operation Ventilation operating specifications Pneumatics Gas source: Anesthesia system. Gas composition: Medical Air or O2. Nominal supply pressure: 350 kPa (51 psi). Pressure range at inlet: 240 to 700 kPa (35 to 102 psi). Flow valve range: 2 to 70 L/min at 240 kPa (35 psi). Fresh gas compensation Flow compensation range: 200 mL/min to 15 L/min. Gas composition: O2, N2O, Air, Heliox, CO2 Anesthetic Agents. Pressure Patient airway pressure range: -20 to +120 cm H2O. Patient airway display range: -9 to 99 cm H2O. High pressure alarm set range: 12 to 99 cm H2O, 1 cm increment. Sustained pressure alarm range: 6 to 30 cm H2O, 1 cm increment. Volume Tidal volume display range: 5 to 9999 mL, 1 mL resolution. Setting range: 45 to 1500 mL. Minute volume: 0.0 to 99.9 liters. Breath rate: 4 to 65 bpm (breaths per minute), 1 bpm resolution. Volume sensor type: Variable flow orifice. 1006-0802-000 07/01 8-19 Aestiva/5 7100 Oxygen Display range: 5 to 110% O2. Display resolution: 1% increments. Sensor type: Galvanic fuel cell. Measurement range: 0 to 100% O2. Measurement accuracy: ± 3% of full scale. Cell response time: 35 seconds,1 10 - 90%. Low O2 alarm range: 18% to 100%. High O2 alarm setting: 21% to 100%. Note: Low O2 limit may not be set above the high O2 limit, nor may the high O2 limit be set below the low O2 limit. Expected cell life: Four months of shelf life (23 °C room air) and one year of normal operation. 1. Response time of cell and adapter as measured using the test method described in ISO 7767 (1988-12-15), clause 50.9. 8-20 07/01 1006-0802-000 8 Specifications and Theory of Operation Ventilator accuracy data The following accuracy data are based on patient conditions and settings described in ASTM F1101. The ventilator is assumed to be operating in volume mode (Heliox OFF), with 100 percent oxygen in the breathing system; or, it is connected to an anesthesia gas analyzer. If the ventilator is operating without being connected to an anesthesia gas analyzer, additional errors may occur which are described in the “Gas composition related errors (both modes)” chart which is provided on the next page. Delivery accuracy Volume control: > 200 mL tidal volume: Better than ±10% of set V T. (100% O2) 75 to 200 mL tidal volume: Better than ±20 mL. < 75 mL tidal volume: Better than ±15 mL. V T comp off: Better than: ±15% of set V T. Pressure mode: Inspiratory pressure: ±2.0 cm H20 (repeatability only). (100% O2) PEEP: ±2.0 cm H2O (repeatability only). Monitoring accuracy Volume control: > 200 mL tidal volume: Better than ±10% of V T. (100% O2) 75 to 200 mL tidal volume: Better than ±20 mL. < 75 mL tidal volume: Better than ±15 mL. Pressure mode: Better than: ±2.0 cm H20 or ±5% of reading whichever is greater. 1006-0802-000 07/01 8-21 Heliox Mode The effect on volume delivery and monitoring accuracies of Heliox mixtures is within +8% to -15% when operating the Aestiva ventilator in the Heliox Mode. positive errors can have the effect of nulling out negative errors.87.5%/% volume agent in Heliox mode. Note: Use of anesthetic agent could affect the errors by approximately -0.95%/% volume agent in normal mode and roughly -2.Aestiva/5 7100 Gas composition related errors (both modes) AA. 8-22 07/01 1006-0802-000 .091 Note: Gas composition errors may be in addition to the above normalized accuracy. When adding errors. Minimum flow:20 L/min. Flow rates:0-10 L/min. Continuous Suction Supply:External vacuum. Maximum flow:>20 L/min. Maximum Vacuum:457 mm Hg with pipeline drive gas at 345 kPa (50 psi). O2 flowmeter (optional) Supply:O2 from system gas supply. Accuracy:±5% of full scale. Accuracy:±5% of full scale. 1006-0802-000 07/01 8-23 . Accuracy:±5% of full scale. not pressure compensated. Regulator Vacuum levels:0-200 mm Hg and full line vacuum. Drive gas consumption:75 L/min. 8 Specifications and Theory of Operation Suction regulators (optional) Venturi Suction Regulator Supply:Air or O2 from system gas supply. Aestiva/5 7100 8-24 07/01 1006-0802-000 . N2O. 8-5 circuit O2 sensor 2-6 flow sensors 2-13 metal or plastic flow sensors 2-6 Cleaning agents 2-1 1006-0802-000 07/01 I-1 . and third gas flow 8-4 D Alarms Disassembly alphabetical list 5-3 patient path 2-7 priority 5-2 scavenging path 2-12 Altitude 3-17 disassembly 2-12 APL valve Display disassembly/maintenance 4-3 change language 3-17 ASTM F1101 8-21 Drain dish 3-8 Autoclave 2-5 E B Electrical block diagram 8-10 Bag arm 2-8 Electrical connections 3-9 install 3-2 Electrical outlets 3-9 Balance regulators 8-4 Electrical power 8-8 Battery backup 8-9 Environmental requirements 8-12 Bellows Exhalation valve 2-12 disassembly/maintenance 4-12 External connection test for correct assembly 4-15 compressed data format 7-6 Bellows assembly 2-9 device commands 7-4 cleaning 2-6 electrical interface 7-2 install 3-2 responses 7-5 Breathing circuit module 2-8 serial parameters 7-3 disassemply/maintenance 4-10 setup data 7-9 identify type 3-3 status bytes 7-7 install 3-3 waveform data 7-9 Breathing profile 8-18 External data Breathing system protocol choices 7-2 setup 3-2 protocol com 1 7-3 Breathing system schematic 8-15 External O2 flowmeter 3-11 Breathing system specifications 8-13 F C Flow sensor Canister calibration 4-25 disassembly 2-10 replace 4-9 setup 3-7 Flow sensor module 2-7 Cell life 8-20 install 3-4 Chain linkage 8-4 CIDEX sterilization 2-13 G Circuit breaker 5-17 Galvanic fuel cell 8-20 Clean and sterilize 2-4 Gas composition related errors 8-22 Cleaning Gas supplies 8-4.A CO2 Bypass assembly 2-15 Absorbent 3-7 Cylinder installation 3-12 Air. H Scavenging receiver disassembly/maintenance 4-8 Heliox mode 3-17 Scavenging reservoir 2-12 High pressure leak test Serial port 3-11 cylinder yoke 3-13 Suction regulator 3-11 DIN cylinder 3-14 Suction regulators (optional) 8-23 Hiss during inspiration 8-16 Symbols 1-3 How to use this manual 1-2 System pneumatic circuits 8-2 System specifications 8-5 L T Leakage current 3-9 Leakage current limit 8-8 Tec 6 power 3-9 Troubleshooting M electrical problems 5-16 Main manifold high pressure leaks 5-18 disassembly/maintenance 4-3 low pressure leaks 5-18 install 3-2 V remove 2-9 Mains indicator 5-17 Ventilation operating specifications 8-19 Mains inlet 3-9 Ventilator Maintenance summary and schedule 1-6 accuracy data 8-21 Mixed gas 8-4 drive gas 3-17 theory 8-16 O Volume control diagram 8-17 O2 flow 8-4 W O2 flowmeter (optional) 8-23 O2 sensor 2-7 WARNINGS and CAUTIONS 1-3 calibration 21% 4-18 Water build-up install 3-3 and alarms 4-26 Optional ventilation functions 1-2 causes for 4-26 solutions 4-26 P Patient circuit connections 3-4 Patient path 2-2 PEEP 8-16. 2-12 I-2 1006-0802-000 07/01 . 8-18 Physical specifications 8-11 Pipeline inlets 3-10 Pneumatic circuit diagram 8-2 Pneumatic outlet 3-10 Pneumatic resistor 8-16 Pressure plateau 8-18 R Repair policy 4-2 S Scavenging connection 3-10 Scavenging path 2-3. negligence. at Datex-Ohmeda’s option. other than its expendable parts. or altered by anyone oth- er than Datex-Ohmeda. this Product. not later than seven (7) days after the expiration date of the applicable war- ranty. is warranted against functional defects in materials and workmanship and to conform to the descrip- tion of the Product contained in this operation manual and accompanying la- bels and/or inserts. but in no event for a period of more than two years from the date of original delivery by Datex-Ohmeda to a Datex-Ohmeda Authorized Dealer. upon Datex-Ohme- da’s examination. Datex-Ohmeda shall not be otherwise liable for any damages including but not limited to incidental damages. There are no express or implied warranties which extend beyond the warranties hereinabove set forth. misuse. free of charge. or if the Product has been subject to abuse. Such warranties are extended only with respect to the pur- chase of this Product directly from Datex-Ohmeda or Datex-Ohmeda’s Authorized Dealers as new merchandise and are extended to the Buyer there- of. This same warranty is made for a period of thirty (30) days with respect to expendable parts. The foregoing warranties shall not apply if the Product has been repaired other than by Datex-Ohmeda or in accordance with written instructions provided by Datex-Ohmeda. Warranty This Product is sold by Datex-Ohmeda under the warranties set forth in the fol- lowing paragraphs. Datex-Ohmeda makes no warranty of merchantability or fitness for a particular purpose with respect to the product or parts thereof. other than for the purpose of resale. Datex-Ohmeda’s sole and exclusive obligation and Buyer’s sole and exclusive remedy under the above warranties is limited to repairing or replacing. which is telephonically reported to the nearest Datex-Ohmeda Field Service Support Center and which. is thereafter returned with a statement of the observed defi- ciency. consequential damages. For a period of twelve (12) months from the date of original delivery to Buyer or to Buyer’s order. is found not to conform with above warranties. provided that the same is properly operated under the conditions of normal use. W-1 . or accident. or special damages. if so advised by Datex-Ohmeda. and which. that regular periodic maintenance and service is per- formed and that replacements and repairs are made in accordance with the in- structions provided. transportation charges prepaid. to the Datex-Ohmeda Service and Distribution Center during normal business hours. a Product. Yoido-Dong. Ltd. Jln. Thailand Madison.A. Taiwan Fax 31 33 253 7223 Tel 1 800 345 2700 Republic of China Calcutta 700091 Spain Sales and Service India Tel 886-2 2515-0457 Datex-Ohmeda S. English USA 1006 0802 000 Tel 608 221 1551 07 01 D 01 21 03 Fax 608 222 9147 Printed in USA www. No. Ltd. P. Room 809. Ltd. Sam Sung Building Tel 33 1 30 68 60 00 36 . Tel 39 2 21693431 Datex-Ohmeda. Ltd. West Malaysia Fax 33 4 78 43 26 58 Datex-Ohmeda Pte. Fax 61 297 461796 Technical Center Miami.V. 3870 CA Hoevelaken Datex-Ohmeda. Tokyo 143-0006 P O Box 356 Canada Japan Homebush NSW 2140 Latin America. Tel 34 1 334 26 00 Equipment Service Center Wisma Danamon Aetna Life 19th Floor Ho Chi Minh City. K.A. USA Datex-Ohmeda Pte. WI 53707-7550. Spain Tel 1 800 345 2700 Indonesia Datex-Ohmeda Pte. Japan France Datex-Ohmeda S. 17 rue Jean-Elysée Dupuy No. 222 Huaihai Road (M) Fax 65 291 6618 Datex-Ohmeda S. Ltd. Netherlands PO Box 7550 Sec. Ra.datex-ohmeda. Ltd. 510095. Milan Fax 8621 5382 1691 12th Floor (Unit F) Grand Amarin Tower Italy Customer Service and Distribution Center 1550 New Petchburi Road. 85. Ltd. Ltd. Chien-Kuo North Road. Datex-Ohmeda K.R. Dist. Malaysia Datex-Ohmeda S.A. COFCO Plaza Tel 81 3 5763 6850 8 Jianguomennei Avenue Fax 81 3 5763 6852 Parc de Pissaloup. 45-46 Jakarta Tel 848 865 5875 United Kingdom 1315 West Century Drive 12930. Fax 39 2 26926226 PO Box 7550 Bangkok 10320. Singapore 189721 Fax 49 2065 691-236 Room 2509 Lippo Plaza Tel 65 391 8618 Italy No. Makasan. USA Block EP & GP. 403 Huan Shi Dong Road Datex-Ohmeda GmbH PO Box 900 Room 1602. BP 10 Datex-Ohmeda Beijing 100005. Unit 2 TRC Annex 9F Units 1 & 2 Tel 44 1707 263570 Mississauga. Truroll Plaza jathevi. Ohmeda Drive PO Box 7550 Aestiva/5 7100 Madison WI 53707-7550 O & M Manual. Ltd. Tel 60 3 754 7872 Germany Datex-Ohmeda Division Fax 60 3 757 6948 Instrumentarium Corp. USA TRC A Bldg. Tel 91 33 3574002 Fax 886-2 2501-9136 C/Manuel Tovar 26 PO Box 7550 Fax 91 33 3574001 Vietnam 28034 Madrid Madison. Inc. India Datex-Ohmeda (India) Pvt. P R China Datex-Ohmeda Pte. Ltd. FL 33176. WI 53707-7550. AE 4-8 Tel 1 305 273 9940 Asia/Pacific Fax 1 305 273 4382 6-1-1 Heiwajima Europe China Ohta-ku. Inc. Datex-Ohmeda Pte. Ltd. Room 1708. USA Tel 62 21 575 0864 Ohmeda House Tel 1 800 345 2755 Fax 62 21 575 0865 71 Great North Road Australia Canada Japan Hatfield Hertfordshire Datex-Ohmeda (Canada) Inc. Ltd. P O Box 5527 Fax 86 10 6526 0653 France Dubai. Sector V Plot XI-16. 2nd Floor. Datex-Ohmeda Pty. Tel 1 800 345 2700 Fax 66 2 207 1014 Fax 1 608 221 4384 Tel 86 27 8571 2536 Kantemarsweg 18 Fax 86 27 8571 2655 Taiwan and Philippines Post Box 22 Technical Support Datex-Ohmeda Pte. Middle East Room B416. Inc. United Arab Emirates 10th Floor. All rights reserved . WI 53707-7550. USA Wusheng Road Netherlands Wuhan 430033. 20090 Segrate. Ltd. PR China 8 Avenue Jean d’Alembert Middle East Operations Korea F-78191 Trappes-Cédex Tel 86 10 6526 9773 Datex-Ohmeda Pte.Corporate Office Datex-Ohmeda Pte. Ltd. 122 Luoguo Street Level 2 Bangunan O'Connor F-69410 Champagne Au Mont d'Or Chengdu 610017. China Thailand Via Cassanese 100 United States Tel 8621 5382 5657 Datex-Ohmeda Pte. Ontario 6-1-1 Heiwajima 149 Arthur Street Fax 44 1707 260191 L5T 1J6 Ohta-ku. 10 Vietnam Fax 34 1 358 12 84 Datex-Ohmeda. Tokyo 143-0006 Datex-Ohmeda Pte.L. D-47228 Duisburg Finland Tel 86 20 8732 2521 152 Beach Road Germany Tel 358 10 394 11 Fax 86 20 8732 2518 #12-05/07 Gateway East Tel 49 2065 691-0 Fax 358 9 146 3310 Datex-Ohmeda Pte. Indonesia Fax 848 862 5501 Datex-Ohmeda Ltd.S. AL9 5EN England 1093 Meyerside Drive. 2 Tel 31 33 253 5404 Madison. Salt Lake City Taipei. Louisville CO 80027-9560.S. Jend Sudirman Kav.1. PR China 13 Jalan 223 France Tel 86 28 661 4424 46100 Petaling Jaya Tel 33 1 30 68 60 00 Fax 86 28 676 2703 Selangor.p. Datex-Ohmeda. Inc. Yunlong Mansion Datex-Ohmeda Pte. North America Shanghai 200021.com ©Datex-Ohmeda. Korea Tel 82 2 786 7421 Fax 82 2 786 7420 Datex-Ohmeda. Caribbean Tel 1 800 268 1472 Tel 81 3 5763 6801 Tel 1 905 565 8572 Fax 81 3 5763 6838 Australia Datex-Ohmeda Latin America Fax 1 905 565 8592 Tel 61 132 229 10685 North Kendall Drive Datex-Ohmeda K. Inc. Youngdeungpo-Ku Tel 97 14 822653 Fax 33 1 30 68 60 01 Fax 97 14 822659 Seoul. P R China Tel 66 2 2071012/13 Datex-Ohmeda B. Inc. GIE Tower Singapore Dr-Alfred-Herrhausen-Allee 24 FIN-00031 Helsinki Guangzhou. 522G Nguyen Tri Phuong St. K.
Copyright © 2024 DOKUMEN.SITE Inc.